Here, we present a protocol to set up and run an in vitro potentiodynamic corrosion system to analyze pitting corrosion for small metallic medical devices.
Different metallic materials have different polarization characteristics as dictated by the open circuit potential, breakdown potential, and passivation potential of the material. The detection of these electrochemical parameters identifies the corrosion factors of a material. A reliable and well-functioning corrosion system is required to achieve this.
Corrosion of the samples was achieved via a potentiodynamic polarization technique employing a three-electrode configuration, consisting of reference, counter, and working electrodes. Prior to commencement a baseline potential is obtained. Following the stabilization of the corrosion potential (Ecorr), the applied potential is ramped at a slow rate in the positive direction relative to the reference electrode. The working electrode was a stainless steel screw. The reference electrode was a standard Ag/AgCl. The counter electrode used was a platinum mesh. Having a reliable and well-functioning in vitro corrosion system to test biomaterials provides an in-expensive technique that allows for the systematic characterization of the material by determining the breakdown potential, to further understand the material’s response to corrosion. The goal of the protocol is to set up and run an in vitro potentiodynamic corrosion system to analyze pitting corrosion for small metallic medical devices.
Electrochemical techniques provide a quick and relatively inexpensive method to obtain the electrochemical properties of a material. These techniques are based predominately on the ability to detect corrosion of a metal by observing the response of the charge-transfer process to a controlled electrochemical disturbance 1-5. Corrosion of metal implants within a body environment is critical due to the adverse implications on biocompatibility and material integrity 6. The main factor contributing to corrosion of implants within the body is the dissolution of the surface oxide leading to an increased release of metallic ions 7-11. This results in adverse biological reactions, which can be found locally, but with potentially systemic effects leading to the premature failure of the implant 10,12-28.
The corrosion characteristics of a test specimen are predicted from the polarization scan produced by a potentiostat. A polarization scan allows for the extrapolation of the kinetic and corrosion parameters of a metal substrate. During a scan, the oxidation or reduction of an electro-active species can be limited by charge transfer and the movement of reactants or products. These factors are all encapsulated by the polarization scan; therefore the importance of having a system that produces a reliable and repeatable polarization scan across multiple cycles is of great importance. The main focus of this manuscript is to provide a protocol identifying the rationale and steps taken to obtain a well-functioning potentiodynamic corrosion system.
1. Construction of the Sample Holder
2. Cleaning of Glassware
3. Setup of Apparatus
4. Running Corrosion Test
5. After the Completion of the Corrosion Run
Note: After the completion of the corrosion run the polarization scan is shown within the analysis view of the software. For each polarization run the presenter view lists the OCP, the plot for E vs. t and the CV staircase which is a plot of E vs. Log (i).
6. Removing the Sample from the Electrode Holder
At the conclusion of the procedure an in vitro corrosion system is setup to conduct corrosion studies. Specific procedures such as the cleaning of the corrosion vessel and the Faraday cage were introduced into the protocol to improve noise performance. The fundamental concept of a good polarization scan is to identify the electro-physical conditions of the material providing valuable information in order to understand the corrosion susceptibility of a metallic material. The procedure and protocol is critical to achieving reliable and reproducible results. Obtaining information on which optimizations can help certain problems that can occur during use, are identified and added into the current procedure. The pre-existing non-calibrated system produced a polarization scan (Figure 1). This scan shows a collection of scattered points, which do not directly allow the determination of the electro-physical conditions of the material under examination. The identification of the OCP was not practical and the corrosion or repassivation potentials are difficult to read. Faults found with the pre-existing setup included a polarization scan that did not reach the maximal potential and was cut off prematurely due to high noise levels. Secondly the scan during the live recording phase exhibited oscillations, which can potentially be due to instability or a high noise levels. Oscillations during the run were seen as a result of instability. Lastly consecutive scans would not deliver reproducible results making it impossible to determine the electro-physical properties of a specific material.
Figure 1. The polarization scan of a Nitanol sample after undergoing electrodynamic potentiostatic polarization. This figure shows a noisy plot that does not provide accurate interpretation of the corrosion parameters. Please click here to view the original version of this figure.
The improvement in noise performance is presented in (Figure 2). The scan shows the direction of the forward and reverse scans and specifies the point at which the protection potential (Epro) and pitting potential (Epit) are observed. The plot is clean with no noise or discrepancy across the whole sweep allowing the entire range to be observed cleanly. The hysteresis loop is reversed at the set potential and returns to intercept the anodic curve, identifying the protection potential. The polarization scan and Tafel plot are the outputs, which provide the fundamental parameters required. These parameters are determined from the polarization scan, therefore having a system that provides high quality scans which are reproducible and reliable is important before identifying the parameters that can be extrapolated.
Figure 2. A polarization scan of stainless steel 316. This is a clear plot resembles a polarization scan following corrosion where the parameters of interest can be easily detected. Please click here to view the original version of this figure.
A study was conducted to analyze changes in surface topography of metal screws following pitting corrosion. The mean Ecorr value obtained from the study was (-0.414 ± 0.05) V. The mean pitting potential for the samples was (0.49 ± 0.12) V, which was within the active region of the polarization curve. The mean protection potential of the samples was (-0.16 ± 0.02) V. Each screw formed localized pits along the surface confirming the findings from the macroscopic images showing detailed topography changes due to the formation of the pits and the changes within the pits (Figure 3). The quantification of the surface topography of the material shows that the roughness of the material has decreased as the overall surface roughness; of the screws Ra was (159.9 ± 7.3) µm (non-corroded) and (124.7 ± 18.3) µm (corroded). The Ra was significantly lower (p =0.02) for the corroded specimen compared to the non-corroded. The average maximum height Rz being (469.3 ± 16.5) µm (non-corroded) and (683.2 ± 85.8) µm (corroded) identifies a significant difference (p = 0.04) for the height variation between a corroded and non-corroded sample. The mean maximum height (Rt) was lowered for non-corroded at (502.61 ± 51.2) µm in comparison to corroded screws at (592.23 ± 119.7) µm.
Figure 3. Images from the optical profilometer that show localized corrosion on a threaded surface. Corrosion has been detected both in the valleys and troughs of the surface. The visual observation of pitting corrosion was successfully achieved by the corrosion setup. Please click here to view the original version of this figure.
Polarization scans produced from the stainless steel samples showed clean continuous plots correlating with scans seen in literature indicative of a well functioning corrosion system which is both reliable and reproducible 29. Poor reproducibility of potentiodynamic pitting potentials is identified with a spread of a few hundred millivolts, with pitting potential being characterized by a stochastic process 29. This is commonly due to the variables of temperature, halide content and potential (V); therefore the smaller variation obtained in the Ecorr from the practical setup is indicative of the protocol and aforementioned adjustments having improved the in vitro setup.
A critical step in the procedure was to establish a stable environment within the reaction vessel and reduce noise. Creating and following specific steps to clean the reaction vessel prior to each run improved the results and provided reproducible and reliable readings. Contaminants within the electrolyte can alter the corrosion environment and the response of the material to corrosion, causing discrepancies in the results. Minimizing this was found to be a critical step in the protocol. The cleaning procedures in place for the electrodes and corrosion vessel remove potential impurities, which could have been a contributing factor to the discrepancies seen previously.
A second critical step within the procedure was to provide an electro shield to the sample holders to eliminate any metal contact within the chamber. The significance of shielding the metal holders completely from any electrochemical conductivity was to prevent interference of external metals. Without isolating the metal specimen under test from any other form of metal species the corrosion analysis cannot provide accurate readings of the test specimen. If the holders are not coated properly they will corrode. If corrosion is seen on metal components that are not under examination the readings cannot be used for analysis and another run would be required.
Initially the scan during the live recording phase exhibited oscillations, which can potentially be due to instability or a high noise level. Oscillations during the run were seen as result of instability. This is due to the failing of the potentiostat to maintain control of the cells potential 30. Oscillations due to a high noise level can be from external sources, which require a degree of filtration. The key to troubleshooting was to connect ceramic disk capacitors between the electrolytes and counter electrode. Suppression capacitors are generally incorporated into alternating current line filters to suppress electromagnetic or radio frequency interference as well as reducing electrical switching noise which is commonly produced by electrical/electronic equipment. Four different magnitudes of ceramic capacitors were used to analyze their noise suppression efficacy on the polarization curve, ranging from 0.001 to 1 µF. The 0.1 µF capacitor smoothed the polarization curve significantly. All noise was suppressed; removing all spikes found in the original scans. The experimental results indicated that the inductance starts to lower the noise suppression efficacy of the 1 µF capacitor, whilst not affecting the 0.1 µF in the frequency range of the present noise.
Potentiodynamic corrosion will provide an in vitro corrosion testing system for materials in controlled environments. A material's corrosion capability can be assessed following any form of manipulation made to the material. The analysis of corrosion with the capability to control different parameters will provide further examination and analysis of corrosion changes in metallic materials. The proposed protocol has both limitations and benefits. The significance of this method in relation to other methods is the relatively low cost and quick process to perform a sophisticated analysis 1,4,5. The protocol will provide a reliable source of laboratory testing to be conducted. However a limitation of the protocol is the limited number of parallel samples that can be tested at one point. The setup only provides one sample per test, which will prolong the test time for a large number of samples.
The authors have nothing to disclose.
The authors had no funding provided for this study.
Potentiostat | Metrohm | PGSTAT101 | |
Ag/AgCl reference electrode, shielded | Metrohm | 6.0729.100 | |
Electrode shaft | Metrohm | 6.1241.060 | |
Polisher Forcipol 1v | Metkon | 3602 | |
Clindrical flask 700mL | SciLabware | FR700F | |
Reaction lid | SciLabware | MAF2/41 | |
Dichloromethane | Sigma-Aldrich | MKBR7629V | use under a fumehood. Wear protective clothing |
Thermo / HAAKE D Series Immersion Circulators | Haake |