This paper presents a protocol for the visualization of gaseous streams of an ambient ionization source using schlieren photography and mass spectrometry.
This manuscript outlines how to visualize mass spectrometry ambient ionization sources using schlieren photography. In order to properly optimize the mass spectrometer, it is necessary to characterize and understand the physical principles of the source. Most commercial ambient ionization sources utilize jets of nitrogen, helium, or atmospheric air to facilitate the ionization of the analyte. As a consequence, schlieren photography can be used to visualize the gas streams by exploiting the differences in refractive index between the streams and ambient air for visualization in real time. The basic setup requires a camera, mirror, flashlight, and razor blade. When properly configured, a real time image of the source is observed by watching its reflection. This allows for insight into the mechanism of action in the source, and pathways to its optimization can be elucidated. Light is shed on an otherwise invisible situation.
Mass Spectrometry, an analytical tool available for molecular mass identification, has become one of the most powerful analytical techniques to date. Over the last decade a whole host of new ambient ionization sources have become available for mass spectrometry detection. For the data collected in this manuscript, the Direct Sample Analysis (DSA) source was utilized. Although these sources are extremely versatile, a more detailed knowledge of the physical ionization process is needed for its optimization and extension of purpose. The aim of this experiment is to gain a better understanding of the ionization process within the ambient sources through visualization of the nitrogen stream on the device using a technique called schlieren photography.
Scientific study often initiates through observation, which is difficult if the object of study is transparent to the naked eye. Schlieren photography is a technique that allows the invisible to become visible through relying on changes in the refractive index within transparent media1. The inhomogeneity of the refractive indices causes a distortion of the light allowing for visualization. The schlieren technique has been routinely used in a variety of specialty fields including ballistics modeling, aerospace engineering, general gas detection and flow monitoring, and at times to visualize protein bands in gel electrophoresis2-5.
Most ambient ionization sources use a stream of gas in order to facilitate the ionization. A wide range of conditions can exist for source options, however the parameters of this experiment must involve the utilization of a gas with a refractive index that differs from the surrounding lab air. This specific study utilizes hot nitrogen. It should be noted that only a small difference in refractive index is observed between pure nitrogen from the gas stream and air at RT6, mainly because air is composed mostly of nitrogen. This issue is overcome in this instance due to the high temperatures of the pure nitrogen in the gas stream which produces a significant enough change in refractive index for the gas to be observed.
Other mass spectrometry sources such as a Desorption Atmospheric Chemical Ionization (DAPCI)7, Flowing Atmospheric Pressure Afterglow (FAPA)8-10, and Direct Analysis in Real Time (DART)11 ionization sources have used schlieren photography. The intention of this protocol is to discuss how to study ambient ionization using a basic schlieren photography configuration. This technique, however, is applicable to any number of different analytical techniques that involve gaseous streams.
1. Schlieren Photography
2. Example Test Object: Mass Spectrometry Ionization Source
3. Determination of Spray Half Angle from a Collected Image
A schematic of the schlieren setup including the mass spectrometry ionization source can be found in Figure 1. When all schlieren components are properly aligned, gases within the test region can be seen as contrasting dark and light regions. Figure 2 illustrates how this contrast can be used to observe how the shape of the nitrogen jet flow from the mass spectrometry source changes as nozzle size decreases.
A full, uncropped schlieren image of the source and gas flow can be found in Figure 3. This image illustrates the orientation of the test objects relative to the mirror. The image in Figure 3 also shows what should be expected when the proper amount, approximately 50%, of light is cut off by the razor blade. If the cutoff is either too high (Figure 4), or too low (Figure 5), poor images will result.
Once the setup is complete, one can adjust various mass spectrometer parameters while watching their effect on the video screen of the camera. This image, alongside the actual signal of the mass spectrometer, allows for optimized conditions to be reached quickly due to the new understanding of the gas stream.
These images can then be used to calculate the spray half angle of the nitrogen stream. The spray half angle tells the user the overall size of the nitrogen gas stream. This angle is effected by diameter of the nozzle, as well as the pressure and temperature of the gas. Figure 6 is a representation of the half angle measurements with constant nozzle size and variations in gas pressure. As expected, the half angle increases accordingly with an increase in pressure, signifying an overall size increase of the gas stream. Figure 7 is a representation of the half angle with constant pressure while changing the nozzle diameter. As expected, the half angle increased with increased nozzle diameter. This signifies an overall scaling increase in size of the nitrogen jet coming out of the source as the nozzle diameter is increased.
Figure 1. Schlieren Schematic (re-print with permission from reference 7). Schematic representation of the schlieren photography apparatus with the mass spectrometry ionization source. Please click here to view a larger version of this figure.
Figure 2. Visualization of Nitrogen Streams (re-print with permission from reference 7). Schlieren photographs of gas flow from the ionization source with different nozzle internal diameters of (A) 4.8 mm, (B) 3.2 mm, (C) 1.5 mm, (D) 0.5 mm. Please click here to view a larger version of this figure.
Figure 3. Visualization of Ambient Source. Wide angle schlieren photograph of the ionization source with proper positioning of the cutoff. Please click here to view a larger version of this figure.
Figure 4. Poor Visualization with Low Cutoff. Schlieren photograph with the cutoff positioned too low. Please click here to view a larger version of this figure.
Figure 5. Poor Visualization with High Cutoff. Schlieren photograph with the cutoff positioned too high. Please click here to view a larger version of this figure.
Figure 6. Half Angle vs. Gas Pressure. A graph depicting the change in the spray half angle with constant nozzle size with varying gas pressure. Please click here to view a larger version of this figure.
Figure 7. Half Angle vs Nozzle Size. A graph depicting the change in the spray half angle with constant pressure with varying nozzle size. Please click here to view a larger version of this figure.
There are several considerations which must be addressed prior to attempting this protocol. In addition to the space around the mass spectrometer for the source and mirror, enough open space must be available to accommodate the distance of twice the focal point of the mirror. Furthermore, the size of the mirror is ultimately decided by the size of the source that is under study. If the mirror is too small, the source will not be fully visualized. It is important to note that some, if not all, of the source covers must be removed to implement the schlieren photography imaging technique.
The most crucial steps of the actual setup are the alignment of each part of the schlieren apparatus. The mirror must be perpendicular to the floor and razor blade must be placed exactly at twice the focal length of the mirror. At this distance, the reflected light will be focused to a small spot. The amount of light blocked by the razor blade is also important. If poor images are produced, the first aspect to adjust would be with the placement of the razor blade. When the razor blade doesn't block enough of the light reaching the camera, no contrast is formed and thus the gas will not be seen. If too much of the light is blocked the images appear dark, making it difficult to distinguish the more subtle details in the nitrogen flow from the object under study.
A limitation of the technique is that there must be a large difference in terms of the refractive index of the background and the area of study. This will depend on temperature and humidity of the laboratory in question. RT nitrogen is ordinarily difficult to see as the background air is composed of approximately 78% nitrogen. This is overcome in the described setup because the temperature of the nitrogen varies from the source which results in changes in the refractive index.
Overall, the significant contribution of this protocol is the ability to understand the physical processes involved with the ionization within the source. This in turn will allow the user to better tune the instrument instead of blindly varying parameters, as well as provide reasoning for the optimized conditions. The advantage of this technique is the ability use all the information from both the physical and chemical processes to gain better sensitivity and selectivity with an ambient ionization source6. The user can utilize the schlieren images to determine physical properties of the source while the mass spectrometry data can be used to understand the chemical properties of the source.
Future applications would be to apply this technique to either various other ambient ionization sources available on the market, or a non-commercial apparatus. This can also be applied to any other instruments/machines which utilize gas streams.
The authors have nothing to disclose.
The authors would like to acknowledge Caitlin Kowalewski for aiding in the editing and formatting of this publication.
Flashlight | EAGTAC | D25A Ti | or equvilent |
Spherical Concave Mirror | Anchor Optics | 27633 | |
Rebel EOS T2i | Canon | 4462B001 | or equvilent |
300 mm telephoto lens | Canon | 6473A003 | or equvilent |
Direct Sample Analysis (DSA) Ionization Source | PerkinElmer | MZ300560 | or equvilent |
Sq 300 MS with SQ Driver Software | PerkinElmer | N2910801 | or equvilent |
Ring Stand | Fisher Scientific | 11-474-207 | or equvilent |
Laser Pointer | Apollo | MP1200 | or equvilent |
razor blade | Blue Hawk | 34112 | or equvilent |
small drill bit #73 | CML Supply | 503-273 | or equvilent |
Protractor | Sterling | 582 | or equvilent |
Hose Clamp | Trident | 720-6000L | or equvilent |