Summary

Light-Induced In Situ Transmission Electron Microscopy for Observation of the Liquid-Soft Matter Interaction

Published: July 26, 2022
doi:

Summary

The present protocol describes transmission electron microscope (TEM) modifications with a light illumination system, the fabrication of liquid cells, and in situ TEM observations of light-induced interactions between bacterial cells and a photosensitizer. The sample preparation methods, electron beam damage, and imaging are also discussed.

Abstract

The current protocol describes the modifications of the transmission electron microscope (TEM) setup for in situ light-induced observations. A glass optical fiber inserted into the electron column above the objective lens polepiece, and a laser, an adjustable light source, was used to fabricate the device. After the illuminator has been calibrated using an external measuring system, it allows one to adjust the intensity of the lighting to the needs of the observed process. This lighting system was utilized to image antimicrobial photodynamic therapy phenomena, which are currently the subject of intense research. The sample was prepared by spotting a suspension of bacteria on a carbon, graphene, or silicon nitride substrate, blotting the excess solution, spotting the photosensitizer solution, blotting the excess liquid again, and then assembling the liquid cell with a second substrate or graphene film. The process of the imaging experiment itself includes choosing the right place for observation with the use of low magnification and a minimum dose of electrons, and then cyclical activation of the light source to capture subsequent images at specified intervals with the minimum amount of electrons necessary. The electron dose of each exposure and the time and intensity of lighting used need to be carefully recorded due to the complexity of the observed phenomena as, at the same time, the process is both light- and electron-driven. After the actual experiment is performed, additional control observations must be made, in which the same doses of electrons are used but without additional light influence and smaller doses of electrons are used for higher doses of light. This makes it possible to distinguish light-induced microstructural effects from those caused by electrons in both the fields of life and materials science.

Introduction

Light-induced phenomena in high resolution are interesting in many fields such as nanoengineering1,2,3, catalysis4,5, and biophotonics6. Some original designs that allow such experiments can be found in the literature,including modifications of the sample holders1,4,7,8,9 and the optical fiber attached to the microscope10,11.

The combination of light illumination, a liquid environment, and transmission electron microscopy (TEM) gives a great opportunity for detailed, dynamical studies of photo-induced processes. However, the high-vacuum condition inside the microscope is rather unfavorable for many liquids, especially water solutions. Liquid encapsulation, which protects it from the environment, can be achieved using a few techniques based mainly on graphene12, silicon nitride13, or carbon14 substrates. In addition to research in materials science2, so-called liquid cells offer possibilities for conducting unconventional microscopic observations on biological specimens near their native conditions15. Such observations are extremely demanding, especially for living microorganisms such as bacterial cells. The electron beam as ionizing radiation causes irreversible damage to the hydrated specimens, so the electron dose must be specified16. This is necessary to minimize unfavorable effects, control the damage, and avoid confusing artifacts. The optimal maximum electron dose that allows observations of living cells is still a questionable topic16, but the dose of 30 e/nm2 appears to be the threshold value, at least for bacteria17.

Some of the subjects of interest for such microscopic studies are processes during antimicrobial photodynamic therapy (APDT)18. In short, the therapy proceeds as follows. The bacterial cells are surrounded by the photosensitive liquid called photosensitizer. When light illumination is given at a specific wavelength, the cytotoxic reactive oxygen species (ROS) are generated from energy or charge transfer from the excited photosensitizer molecules to the oxygen naturally present in the solution. Pathogens exposed to ROS are quickly inactivated with very high efficiency, with no side effects19. The response to the therapy varies for distinct microbes – for example, the impact of the same photosensitizer may be quite different for Gram-positive and Gram-negative bacteria20. In general, it has been established that the main target of ROS is the outer structures of cells, where damage results in functional disorders of the cell membrane and, consequently, lead to the death of bacteria21,22. However, damage to nucleic acids and proteinscan also be considered a cause of inactivation18, so it is still unknown which cell structures are the main targets during this process19. A deeper understanding of the damaging processes could help to improve this definitive therapy. Compared to the light microscopy methods used in APDT research23, TEM techniques give more possibilities to look at the APDT mechanism with higher resolution and magnification24. TEM has already been successfully used for cell observation during ongoing therapy, which allowed us to study Gram-positive bacteriadamage and describe the changesoccurring within the cell wall in detail6,25.

The current protocol presents a suitable experimental setup for high-resolution imaging of light-induced bacteria inactivation using TEM, which requires a proper light illumination system, the encapsulation of cells with a liquid, and strict electron dose control. The bacteria used for the observation was Staphylococcus aureus, and a methylene blue solution was used as a photosensitizer. The special light illumination setup comprises a tunable semiconductor laser connected directly to the microscope column using the light fiber. This design provides uniform irradiation throughout the sample because of the almost-parallel placement of the optical fiber to the microscope axis. Monochromatic light of high intensity generated by the laser can then be used to study various photochemical effects. The light used in the experiment had a wavelength equal to 660 nm because, in the visible region, methylene blue has absorption peaks at 613 nm and 664 nm26. The protocol for liquid encapsulation is based on carbon substrates, which makes the procedure quick and uncomplicated. Finally, a method for low-dose in situ TEM observation of cells in liquid is presented. The difficulties regarding sample preparation, the electron dose effects on the sensitive specimen, and reasonable image interpretation are discussed.

Protocol

1. Transmission electron microscope modification

  1. Modify the transmission electron microscope by connecting the optical fiber (see Table of Materials) to the microscope column at the top of the objective lens. Allow a few centimeters of space between the objective lens and the condenser lens, plus a free slot for accessories.
    NOTE: A dedicated sample holder4 or a holder for cathodoluminescent observations in reverse configuration1 can also be employed.
  2. Check the system for leaks using rough vacuum pumps. If the system does not show a vacuum leak, test the system under a high vacuum.
    1. Check that the actuation of the vacuum system does not deviate from the standard vented microscope start. In case of problems with the execution, assembly, or commissioning of the module, contact authorized service personnel or the device manufacturer.
      NOTE: The correctly made and mounted illuminator will not obscure the electron beam. If the beam shifts successively during the microscope's operation, the nonconductive optical fiber will not be sufficiently covered with a metal shield and will accumulate an electric charge. In this case, slide the optical fiber deep into the conducting jacket.
  3. Measure the light intensity using the photodiode power sensor (see Table of Materials). If the space between the pole pieces of the objective lens does not allow measurement inside the microscope, measure the geometry of the sample and the illuminator and reproduce these conditions in an external measurement setup.
  4. If the microscope is not equipped with an electron dose measurement system, perform the calibration by measuring the beam current with a Faraday cup (see Table of Materials) and numerically calculating the density of electrons hitting the specimen27.

2. Encapsulation of bacteria with the photosensitizer

  1. Place two untreated TEM grids covered with amorphous carbon between two crossed tweezers facing the carbon-coated side to the top. Hold the grids as close to the edge as possible.
    NOTE: For certain substrate, solution, and sample combinations, it may be beneficial to use glow discharge on one or both meshes to remove factory impurities and obtain a maximum hydrophilic surface28. Using the example of S. aureus bacteria and methylene blue solution, it was determined that the best and most reproducible results were obtained on new, non-treated carbon on copper grids without additional glow discharge or plasma cleaning6.
  2. Put 4 µL of bacteria solution on one of the grids. The optical density of the sample must be in the range of 5-10.
    NOTE: The bacteria used in the study were purified from the culture and dispersed in PBS buffer. It is highly recommended to check the concentration of bacteria in the fresh sample each time using it, e.g., using negative staining methods, following the established protocols available29,30,31.
  3. Wait for 60 s and carefully blot the liquid using filter paper. Try to spread the liquid throughout the entire grid surface during this process.
  4. Put 4 µL of photosensitizer on the same grid.
    NOTE: Methylene blue at a concentration of 2 µg/mL was used as the photosensitizer for the present study.
  5. After 5 s, blot out the liquid. Leave a very thin layer of liquid on the substrate.
  6. Quickly place the dry grid on top of the one covered with liquid.
  7. Gently move the upper grid to the edge of the second one and squeeze the substrates at the edges using tweezers.
    NOTE: Some liquid may flow out of the sample at this stage, which means that not enough was drained off. However, this does not necessarily mean that the formation of the liquid cell will be unsuccessful.
  8. Carefully repeat the squeeze.
    NOTE: It is easier to perform this step by rotating the tweezers by 90° so that they lie perpendicular to the table and both tweezers' tips remain at the same height, not depending on the open/close position.
  9. Leave the liquid cell between the tweezers for 1 min.
  10. Insert the sample into the TEM holder right after finishing the preparation.
    NOTE: The tension of the holder mounting plate/ring will help the substrates to adhere and keep the liquid inside.
    1. If possible, pump the specimen holder into an external pumping station to reduce the contamination of the microscope by evaporating liquid.

3. In situ TEM observations of bacterial cells in liquid-electron dose effect

  1. Insert the sample into the microscope.
  2. Start the observations in low magnification mode to find a suitable observation area.
    NOTE: Darker areas are most likely to be liquid because of electron beam scattering.
    1. During the observations, keep the electron dose as low as possible. Expand the exposure time (up to 2 s for an accelerating voltage of 150 kV and a magnification of 5,000x). The magnification of 5,000x for the bottom-mount camera (approximate field of view is 10 µm) is sufficient to image bacteria with a diameter of ~1 µm.
      NOTE: Sometimes, it is not easy to distinguish the liquid, so, at the beginning of the observations, it is helpful to focus the electron beam where it is likely to be liquid. The liquid should move upon a focused electron beam, and the bubbles will appear. After this verification, it is necessary to find another area for further observations.
  3. Find the cells surrounded by liquid at the appropriate magnification with extended exposure time and capture the first image immediately. Keep the electron dose constant and collect images every 30 s to observe the changes.
  4. Carefully examine the images and calculate the total electron dose27 that corresponds to the first visible changes in the cells.

4. In situ TEM observations of light-induced processes between bacterial cells and photosensitizer

  1. Before starting the experiment, set the laser light intensity by adjusting the current value. Choose the appropriate wavelength according to the absorption spectrum of the photosensitizer. According to the previous calibration plot (step 1.4.), choose the current value corresponding to the lowest light intensity.
    NOTE: The light wavelength used in this study was 660 nm.
  2. Insert the sample into the microscope and find the observation area following step 3.2. in the previous section.
  3. Capture the first image of the cell before starting illumination of the sample and use the beam blanker to turn off the electron beam to avoid the effects of electron irradiation.
  4. Turn on the laser. Start with a short illumination time (here 30 s) as the laser light has relatively high intensity. After that time, turn off the light source.
  5. Turn off the beam blanker and capture the image immediately. If possible, use an automatic algorithm to expose the sample only for the time needed for taking the image. Carefully measure the electron irradiation time to calculate the electron dose27 used for imaging.
  6. Repeat step 4.4. and step 4.5. to obtain the expected light illumination time.

Representative Results

The experimental setup was designed to observe the light-induced processes occurring in a liquid with high resolution. The competitive analysis of the images allowed us to distinguish the damage caused by the electron beam from changes related to the photochemical reaction. The effect of the electron beam on the sensitive specimen (in this case, the cells encapsulated with liquid) was visible all over the irradiated area as the electrons penetrated through it uniformly. Of course, the effect depends on the electron dose, so more significant changes in the sample morphology can be observed faster for a more focused beam. Compared to this severe damage, the actual destructive effects of photodynamic inactivation occurred in specific parts of cells only in areas surrounded by the illuminated photosensitizer.

In the example of bacteria imaging, and according to the literature, the dose of 30 e/nm2 seems to be the lethal threshold value for bacteria17, which was highly exceeded in the authors' previous studies6,25. However, the visible changes in the cell were not noticed before reaching the electron dose of 6000 e/nm2. The results showed a notable degradation of the outer layer of the cell caused by the reactive oxygen species generated by light irradiation of the photosensitizer after 1 min (Figure 1). Further light illumination made the changes more visible (Figure 1D,G).

The encapsulation of the cells between the carbon substrates was sufficient to perform constant observations for at least 1 h. Proper observation spots with enough liquid can be easily found at low magnification, as these areas appear darker and blurred (indicated by the frames in Figure 2A). The liquid-covered bacteria are less visible than the dry ones but can still be distinguished (Figure 2B). It is useful to look at the liquid boundary during the imaging, and its movement can be easily detected, which implies that the chosen area might be unsuitable and unstable (Figure 3). Another issue during observations is the evaporation of the water and precipitation of the solution, which can occur in random areas of the sample, including the areas surrounding the bacteria (Figure 3B,C).

Figure 1
Figure 1: Exemplary results of the photodynamic treatment of bacteria. (A) The scheme of the liquid cell containing bacterial cells and methylene blue. (B) The cell before light illumination. (C) The cell after illumination for 1 min. (D) The cell after illumination for 10 min. (EG) The magnified areas from (BD). The electron dose was equal to 1600 e/nm2, 2500 e/nm2, and 6000 e/nm2, respectively. The figure is reproduced from Żak et al.25. Please click here to view a larger version of this figure.

Figure 2
Figure 2: TEM images of S. aureus encapsulated with methylene blue between two carbon films. (A) Low magnification image presents both areas with a high amount of liquid (indicated by the yellow frames) and dry spots. A quick examination of the sample at this magnification gives information on whether the encapsulation was successful or not. (B) The liquid boundary is clearly visible in the magnified area, and the bacteria covered with the photosensitizer can be distinguished. Please click here to view a larger version of this figure.

Figure 3
Figure 3: TEM images of unfavorable effects that may occur during imaging. (A) In the beginning, the cell was uniformly surrounded by the liquid, which covered most of the observation area. (B) Based on the constant electron beam irradiation, the movement and evaporation of the liquid and the start of the precipitation from the solution can be noticed. (C) The continuation of the processes leads to irreversible movement, foaming, and disintegration of the photosensitizer, making further observations impossible. Please click here to view a larger version of this figure.

Figure 4
Figure 4: The unfavorable effects leading to wrong conclusions. (A) The crystallized liquid. (B) The contamination settled on the cell. (C) Foaming caused by the electron beam. Please click here to view a larger version of this figure.

Discussion

The installation and commissioning of the illuminator require basic service knowledge and can damage the microscope. The simplest way to introduce the light to the microscope is to connect the optical fiber from the top of the objective lens, where there is usually room for the TEM deflection coils and additional detectors. More free space can also be expected from older top-entry devices, where the same location houses the sample vacuum lock and the sample installation mechanism. This configuration is widely described in a previous report25, which contains an exemplary illuminator design. If the interior of the microscope does not allow the installation of the optical fiber above the objective lens, the illuminator may be installed inside the objective lens polepiece from the side32. Other solutions assume using a dedicated sample holder4 or the reverse use of the holder for cathodoluminescent observations1.

In general, the preparation of liquid samples for TEM is complicated and requires experience and manual skills. The critical step in liquid cell preparation is step 2.6. of the protocol. It is important to place the second grid quickly to avoid the evaporation of the liquid. The method presented here is simple and does not require uncommon materials; however, it might not suit all liquids. In some cases, it is beneficial to clean the substrate with plasma to increase its hydrophilicity. For liquid samples, especially water solutions, which do not contain any particles and/or cells that could serve as a spacer between the carbon films, the encapsulation might be unsuccessful as all the liquid may flow out from between the flat surfaces. Using a graphene liquid cell prepared using a holey substrate, which creates small “pockets” for liquid, might be more suitable33.

TEM observations performed in liquid can be disturbed by a few factors related to the sample preparation and the unfavorable effects caused by the electron beam. The first includes the detachment of substrates and the movement of the liquid, often followed by the breaking of the carbon film, which occurs for high liquid volumes in the area and/or upon electron irradiation. This can be avoided by minimizing the liquid volume in step 2.5. of the protocol.

The irradiation with highly energetic electrons leads to energy transfer, resulting in heating or radiolysis. After interaction with radiation, water decomposes into radicals (H· and OH·), H2, and solvated electrons. These species participate in further reactions, thus generating different products and causing damage to the sample34. Therefore, it is very important to keep the electron dose low during the observations (step 3.2.), especially for sensitive specimens like cells. Other effects caused by the electron beam are precipitation from the solution (Figure 4A) and foaming of the liquid (Figure 4C). Lowering the electron dose helps to avoid these two negative effects, but sometimes it is not possible to eliminate them. There may also be some radiation-sensitive contamination in the solution, which settles on the cell (Figure 4B) and changes upon electron beam influence. This is unfavorable as the impurity might be taken as part of the cell, which is damaged upon the light irradiation, leading to wrong conclusions. The best way to obtain reliable images is to find an uncontaminated area and keep the electron dose as low as possible.

Reliable in situ TEM studies of photoreactions require a good comparison between the processes induced by light with those caused by the electron beam. Except for lowering the electron dose described in step 3.2., it is helpful to turn off the electron source for the time of light illumination so that the undisturbed results can be observed, as emphasized in step 4.3. of the protocol.

The present protocol can be used to observe light-induced reactions (especially in liquid), for example, light on metal nanoclusters2, light-induced catalysis processes1, and the influence of light on biological specimens (e.g., cells with a photosensitizer, chloroplasts).

Declarações

The authors have nothing to disclose.

Acknowledgements

The research was supported by the Miniatura grant (2019/03/X/NZ3/02100, National Science Center, Poland).

Materials

Carbon film on 200 mesh copper grid Agar Scientific AGS160 The standard TEM grids for observations and liquid cell preparation
Crossover Tweezers Dumont N5 The tweezers are neecesarry for liquid cell preparation
Photodiode Power Sensor ThorLabs S130C The sensor used for light intensity measurement
Polyimide-Coated Multimode Fiber Thorlabs FG400UEP Must be built into the microscope using the on-site built adapter, according to the 10.1016/j.ultramic.2021.113388
Transmission Electron Microscope Hitachi H-800 Can be replaced with any side-entry microscope, available for modification
Tuneable Diode Laser CNI MRL-III-660D The light wavelength must be chosen basing on photosensitizer's absorption spectrum

Referências

  1. Vadai, M., Angell, D. K., Hayee, F., Sytwu, K., Dionne, J. A. In-situ observation of plasmon-controlled photocatalytic dehydrogenation of individual palladium nanoparticles. Nature Communications. 9 (1), 1-8 (2018).
  2. Weng, B., et al. Visualizing light-induced dynamic structural transformations of Au clusters-based photocatalyst via in situ TEM. Nano Research. 12 (1), 1-5 (2021).
  3. Cao, K., et al. In situ TEM investigation on ultrafast reversible lithiation and delithiation cycling of Sn@C yolk-shell nanoparticles as anodes for lithium ion batteries. Nano Energy. 40, 187-194 (2017).
  4. Cavalca, F., et al. In situ transmission electron microscopy of light-induced photocatalytic reactions. Nanotechnology. 23 (7), 075705 (2012).
  5. Tung, C. -. W., et al. Light-induced activation of adaptive junction for efficient solar-driven oxygen evolution: In situ unraveling the interfacial metal-silicon junction. Advanced Energy Materials. 9 (31), 1901308 (2019).
  6. Żak, A. M., Kaczmarczyk, O., Piksa, M., Matczyszyn, K. Light-induced in situ transmission electron microscopy: Novel approach for antimicrobial photodynamic therapy imaging. Photodiagnosis and Photodynamic Therapy. 35, 102463 (2021).
  7. Shindo, D., et al. Development of a multifunctional TEM specimen holder equipped with a piezodriving probe and a laser irradiation port. Journal of Electron Microscopy. 58 (4), 245-249 (2009).
  8. Zhang, C., et al. Photosensing performance of branched CdS/ZnO heterostructures as revealed by in situ TEM and photodetector tests. Nanoscale. 6 (14), 8084-8090 (2014).
  9. Zhang, C., et al. Statistically analyzed photoresponse of elastically bent CdS nanowires probed by light-compatible in situ high-resolution TEM. Nano Letters. 16 (10), 6008-6013 (2016).
  10. Yoshida, K., Yamasaki, J., Tanaka, N. In situ high-resolution transmission electron microscopy observation of photodecomposition process of poly-hydrocarbons on catalytic TiO2 films. Applied Physics Letters. 84 (14), 2542-2544 (2004).
  11. Yoshida, K., Nozaki, T., Hirayama, T., Tanaka, N. In situ high-resolution transmission electron microscopy of photocatalytic reactions by excited electrons in ionic liquid. Journal of Electron Microscopy. 56 (5), 177-180 (2007).
  12. Textor, M., De Jonge, N. Strategies for preparing graphene liquid cells for transmission electron microscopy. Nano Letters. 18 (6), 3313-3321 (2018).
  13. Ring, E. A., De Jonge, N. Microfluidic system for transmission electron microscopy. Microscopy and Microanalysis. 16 (5), 622-629 (2010).
  14. Inayoshi, Y., Minoda, H. A carbon sandwich environmental cell for wet specimens. Journal of Electron Microscopy. 62 (6), 623-628 (2013).
  15. Koo, K., Dae, K. S., Hahn, Y. K., Yuk, J. M. Live cell electron microscopy using graphene veils. Nano Letters. 20 (6), 4708-4713 (2020).
  16. De Jonge, N., Peckys, D. B. Live cell electron microscopy is probably impossible. ACS Nano. 10 (10), 9061-9063 (2016).
  17. Kennedy, E., Nelson, E. M., Damiano, J., Timp, G. Gene expression in electron-beam-irradiated bacteria in reply to "live cell electron microscopy is probably impossible.&#34. ACS Nano. 11 (1), 3-7 (2017).
  18. Alves, E., et al. An insight on bacterial cellular targets of photodynamic inactivation. Future Medicinal Chemistry. 6 (2), 141-164 (2014).
  19. Cieplik, F., et al. Antimicrobial photodynamic therapy-What we know and what we don’t. Critical Reviews in Microbiology. 44 (5), 571-589 (2018).
  20. Huang, L., et al. Type I and Type II mechanisms of antimicrobial photodynamic therapy: An in vitro study on gram-negative and gram-positive bacteria. Lasers in Surgery and Medicine. 44 (6), 490-499 (2012).
  21. Caminos, D. A., Spesia, M. B., Pons, P., Durantini, E. N. Mechanisms of Escherichia coli photodynamic inactivation by an amphiphilic tricationic porphyrin and 5,10,15,20-tetra(4-N,N,N-trimethylammoniumphenyl) porphyrin. Photochemical and Photobiological Sciences. 7 (9), 1071-1078 (2008).
  22. Chu, J. C. H., Chin, M. L., Wong, C. T. T., Hui, M., Lo, P. C., Ng, D. K. P. One-pot synthesis of a cyclic antimicrobial peptide-conjugated phthalocyanine for synergistic chemo-photodynamic killing of multidrug-resistant bacteria. Advanced Therapeutics. 4 (3), 1-10 (2021).
  23. Rogers, S., Honma, K., Mang, T. S. Confocal fluorescence imaging to evaluate the effect of antimicrobial photodynamic therapy depth on P. gingivalis and T. denticola biofilms. Photodiagnosis and Photodynamic Therapy. 23, 18-24 (2018).
  24. Maldonado-Carmona, N., et al. Porphyrin-loaded lignin nanoparticles against bacteria: A photodynamic antimicrobial chemotherapy application. Frontiers in Microbiology. 11, 606185 (2020).
  25. Żak, A. M., Kaczmarczyk, O., Piksa, M., Grzęda, J., Matczyszyn, K. Fiber-optic sample illuminator design for the observation of light induced phenomena with transmission electron microscopy in situ: Antimicrobial photodynamic therapy. Ultramicroscopy. 230, 113388 (2021).
  26. Dinh, V. P., et al. Insight into the adsorption mechanisms of methylene blue and chromium(III) from aqueous solution onto pomelo fruit peel. RSC Advances. 9 (44), 25847-25860 (2019).
  27. Żak, A. Guide to controlling the electron dose to improve low-dose imaging of sensitive samples. Micron. 145, 103058 (2021).
  28. Drulyte, I., et al. Approaches to altering particle distributions in cryo-electron microscopy sample preparation. Acta Crystallographica Section D. 74 (6), 560-571 (2018).
  29. Scarff, C. A., Fuller, M. J. G., Thompson, R. F., Iadaza, M. G. Variations on negative stain electron microscopy methods: Tools for tackling challenging systems. Journal of Visualized Experiments. (132), (2018).
  30. Booth, D. S., Avila-Sakar, A., Cheng, Y. Visualizing proteins and macromolecular complexes by negative stain EM: From grid preparation to image acquisition. Journal of Visualized Experiments. (58), (2011).
  31. Rames, M., Yu, Y., Ren, G. Optimized negative staining: A high-throughput protocol for examining small and asymmetric protein structure by electron microscopy. Journal of Visualized Experiments. (90), e51087 (2014).
  32. Miller, B. K., Crozier, P. A. System for in situ UV-visible illumination of environmental transmission electron microscopy samples. Microscopy and Microanalysis. 19 (2), 461-469 (2013).
  33. Yuk, J. M., et al. High-resolution EM of colloidal nanocrystal growth using graphene liquid cells. Science. 336 (6077), 61-64 (2012).
  34. Schneider, N. M., et al. Electron-Water interactions and implications for liquid cell electron microscopy. Journal of Physical Chemistry C. 118 (38), 22373-22382 (2014).

Play Video

Citar este artigo
Żak, A., Kaczmarczyk, O. Light-Induced In Situ Transmission Electron Microscopy for Observation of the Liquid-Soft Matter Interaction. J. Vis. Exp. (185), e63742, doi:10.3791/63742 (2022).

View Video