A protocol for the matrix-assisted laser desorption ionization time of flight mass spectrometry (MALDI-TOF MS) characterization of synthetic polymers is described including the optimization of sample preparation, spectral acquisition, and data analysis.
There are many techniques that can be employed in the characterization of synthetic homopolymers, but few provide as useful of information for end group analysis as matrix-assisted laser desorption ionization time of flight mass spectrometry (MALDI-TOF MS). This tutorial demonstrates methods for optimization of the sample preparation, spectral acquisition, and data analysis of synthetic polymers using MALDI-TOF MS. Critical parameters during sample preparation include the selection of the matrix, identification of an appropriate cationization salt, and tuning the relative proportions of the matrix, cation, and analyte. The acquisition parameters, such as mode (linear or reflector), polarization (positive or negative), acceleration voltage, and delay time, are also important. Given some knowledge of the chemistry involved to synthesize the polymer and optimizing both the data acquisition parameters and the sample preparation conditions, spectra should be obtained with sufficient resolution and mass accuracy to enable the unambiguous determination of the end groups of most homopolymers (masses below 10,000) in addition to the repeat unit mass and the overall molecular weight distribution. Though demonstrated on a limited set of polymers, these general techniques are applicable to a much wider range of synthetic polymers for determining mass distributions, though end group determination is only possible for homopolymers with narrow dispersity.
With improvements in living polymerization techniques, precision polymers with quantitatively functionalized end groups are increasingly available1. The concurrent development of azide-alkyne and thiolene click chemistries has enabled the nearly quantitative coupling of macromolecules to other moieties, providing access to a range of hybrid materials2,3,4. However, precise analytical techniques are required to characterize both the starting materials and products of these polymer conjugation reactions. Matrix-assisted laser desorption/ionization time of flight mass spectrometry (MALDI-TOF MS) is a valuable soft ionization analytical technique for characterizing polymers because it can generate polymer ions in a single charge state with minimal fragmentation5,6. MALDI-TOF MS has major advantages over other conventional methods of polymer characterization because it can provide mass spectra with resolution of the individual n-mers within the polymer mass distribution. As a consequence, such mass spectra can provide precise information about the average molecular weight, repeat unit mass, and molecular weight dispersity7, which can in turn elucidate competing polymerization mechanisms such as chain transfer8. However, MALDI-TOF MS is particularly powerful at providing information about polymer end groups9,10, which can be used to confirm end group modifications10,11 and other transformations12 such as polymer cyclizations11,13. Equally important, the relatively small amount of analyte (sub-micrograms) required for mass spectrometric analysis makes this technique useful for characterization when only trace quantities of material are available.
The MALDI-TOF MS analysis of polymers can be divided into four distinct steps: sample preparation, instrument calibration, spectral acquisition, and data analysis. Sample preparation is the most essential step for generating optimized MALDI-TOF mass spectra and occurs before the sample is introduced into the instrument14,15. The selection of an appropriate matrix with similar solubility parameters as the polymer analyte is critical to obtain high quality MALDI-TOF mass spectra and guidelines for matrix selection have been reported elsewhere14,15,16,17. A database of polymer MALDI "recipes" for sample preparation has also been published online18. For novel polymers, matrix selection can be approached by first understanding the solubility of the polymer and selecting a matrix with similar solubility parameters14,19. Polymers with high proton affinity can be protonated by most matrices14 (which frequently contain carboxylic acid groups), but for other polymers, a cationization agent is required14. Alkali ions adduct well with oxygen-containing species (e.g. polyesters and polyethers), whereas unsaturated hydrocarbons (e.g. polystyrene) adduct with transition metals such as silver and copper ions14,19. Because the polymer samples in this experiment contained oxygen atoms in the backbone, sodium or potassium trifluoroacetate (TFA) were used as the cationization agent. Once the matrix and cationization agents have been selected, the relative proportions of analyte, cation agent, and matrix must be carefully optimized to ensure high signal to noise. In this procedure, the parameters for sample preparation have already been optimized, however an empirical sample optimization procedure (step 1.4.1., Figure 1) that systematically varies the concentrations of the three components (analyte, matrix and cation) is effective for rapidly determining their optimal ratios.
Data acquisition also requires the optimization of a number of parameters. The most important parameters include the positive or negative ion mode of the spectrometer, the instrument operation mode (linear versus reflector), the acceleration voltage, and the extraction delay time. Another way that resolution can be increased is through the utilization of "reflectron" mode20,21,22,23. Reflectron mode essentially doubles the flight path of the ions to the detector by reflecting the ions at the end of the flight tube back towards a detector near the source while refocusing ions with different momentums, and therefore increasing the resolution though decreasing signal strength. In addition, higher resolution spectra can be obtained by decreasing the laser power which minimizes the signal-to-noise ratio by decreasing the number and energy of collisions and therefore reducing the fragmentation and kinetic inhomogeneities24. By tuning all of these parameters, the ions can be focused to minimize the effect of any inhomogeneity in initial position or velocity that occurs during the laser desorption process. When the acquisition parameters are optimized, isotopic resolution can often be achieved for ions with masses in excess of 10,000 Da, though this is also dependent upon the length of the flight tube and the instrument design. Most organic compounds that contain at least one heteroatom are prone to complexing with alkali cations such as lithium, sodium, and potassium. Many of the alkali metals are monoisotopes or of limited isotopes and therefore do not broaden the distribution.
While the instrument parameters can be tuned to optimize data precision, data accuracy is only achieved with an appropriate calibration11. Proteins and peptides were originally used as calibrants due to their monodispersity and availability, but suffer from variable stability and the prevalence of impurities25. More cost-effective and stable alternatives have included inorganic clusters and polydisperse polymers26,27,28,29. Unfortunately, these alternatives feature disperse masses, which complicate mass assignments, as well as smaller masses overall, making them useful only for calibrations below 10,000 Da. To combat these issues, Grayson et al.25 developed a dendrimer-based, polyester MS calibration system that is monodisperse, and boasts both broad matrix and solvent compatibility, shelf-life stability (> 8 years), and lower production cost. Based upon the strengths of this system, it was selected as the calibrant for these experiments.
There are two main types of calibration: internal and external30. When calibrating externally, a standard with masses that bracket that of the analyte are placed on the MALDI target plate in a different sample position than the analyte to generate a separate mass spectrum from which a calibration file can be generated. On the other hand, increased accuracy can often be achieved with an internal calibration, which involves mixing the calibrant with the analyte to obtain a hybrid spectrum with both calibrant and analyte signals. In the procedure described below, an external calibration was implemented. After proper calibration of the mass scale, accurate analyte mass data can be acquired. To ensure the most accurate calibration, it is important that the data acquisition occurs soon after the calibration.
Finally, once the optimized, calibrated data sets were acquired, and the data were analyzed to provide structural information about the polymer samples. The spacing between n-mers within the polymer distribution can provide accurate measurement of the repeating unit mass. The number average molecular weight (Mn) and other mass distribution calculations (e.g., Mw (weight average molecular weight) and Đ (dispersity)) can also be determined from the signal distribution in the mass spectra (step 4.2 for calculations). Perhaps most uniquely, in the case of homopolymers, the sum of the end group masses can be confirmed by determining the offset of the polymer distribution with respect to the mass of the repeating units alone. The information-rich MALDI-TOF mass spectra provide valuable characterization data that are complementary to more traditional polymer characterization techniques such as size exclusion chromatography, Fourier-transform infrared spectroscopy, and nuclear magnetic resonance.
Caution: All reactions were run in a fume hood. Please read all Material Safety Data Sheets (MSDS) for any chemical used and take appropriate precautions.
1. Sample Preparation
2. Data Acquisition Optimization
3. MALDI Calibration
4. Data Analysis and Interpretation
Sample 1: A sample of poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn = 5000) (Figure 3) was analyzed using potassium trifluoroacetate as a cationization agent with HCCA as the matrix. The spectrum exhibited the expected K+ adducts as well as those observed from residual Na+.
MALDI-TOF MS confirms the narrow distribution (Figure 3) of poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn = 5000). Because the monoisotopic peak (comprising exclusively the most abundant elemental isotopes, namely 12C, 1H 16O, and 14N) is not sufficiently resolved to enable its identification, a pick peaking protocol is used that determines the average mass across the entire isotopic distribution for each n-mer peak. Likewise, all theoretical calculations are determined using average, rather than monoisotopic, masses for each element. Using the equations from step 4, analysis software was used to calculate the following characteristics of the polymer mass distribution: Mn: 4700, Mw: 4710, Đ: 1.00.
In order to confirm the identity of the end groups, an individual n-mer (104) was selected for further analysis (Figure 4). As with the mass distribution calculations, because the monoisotopic peak could not be resolved, average mass values were used for subsequent calculations. The theoretical mass value of the 104-mer of poly(ethylene glycol) 2-aminoethyl ether acetic acid is comprised of the mass of the repeat units (44.0530 × 104) plus the mass of the α-amine end group (+ 16.02300) and the mass of the ω-carboxyl end group (+ 59.0440) plus the mass of the potassium cation (+ 39.09775) which yields a total 104-mer mass of 4695.67675. The observed mass value for the 104-mer + K+ is 4695.5 which matches the theoretical value, given the precision of average mass calculations. The series of smaller, offset peaks in the spectrum corresponds to the polymer ionizing with sodium where the theoretical mass value of the 104-mer is comprised of the mass of the repeat units (44.0530 × 104) plus the mass of the α-amine end group (+ 16.02300) plus the mass of the ω-carboxyl end group (+ 59.0440) plus the mass of the sodium cation (+ 22.98922) giving a total 104-mer mass of 4679.56822. The observed mass value for the 104-mer + Na+ is 4679.4 which is only 0.2 Da different from the theoretical value. More accurate determinations of end group mass can be determined by measuring the average across multiple peaks, and has been discussed elsewhere11.
The poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn = 5000) sample maintained its narrow distribution when selectively functionalized by reaction (Figure 5) with 2,4-dinitrofluorobenzene (DNFB) (Figure 6). The spectrum exhibited sodium adducts and used HCCA as the matrix.
MALDI-TOF MS confirms the narrow distribution (Figure 6) of poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn = 5000) when modified with DNFB. Using the equations from step 4, analysis software was used to calculate the following characteristics of the polymer mass distribution: Mn: 4940, Mw: 4950 Đ: 1.00.
In order to determine if complete functionalization of the poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn = 5000) had occurred with DNFB, an individual n-mer of the distribution was selected for analysis (Figure 7). The theoretical mass of the functionalized 104-mer of poly(ethylene glycol) 2-aminoethyl ether acetic acid reacted with 2,4-dinitrofluorobenzene is comprised of 44.0530 × 104 (the mass of the repeat units) + 182.115 (mass of the α-amine group reacted with 2,4-dinitrofluorobenzene) + 59.044 (mass of the carboxyl group) + 22.98922 (mass of the sodium cation) = 4845.66022. The observed mass value for n = 104 is 4845.8 which is -0.1 Da different from the theoretical value. This close agreement between the theoretical and observed values is indicative of a complete modification of the starting material to product, but more significantly, the lack of signals associated with the starting material, 4811.72722 and 4855.78022 for this mass range, or any additional byproducts confirms the quantitative selective functionalization of the amine. A second peak is observed at 4823.8 which matches the 103-mer of the functionalized polymer, but with the loss of the proton on the carboxylic acid end group that complexes with another sodium ion with a theoretical mass of 4823.58899 which has difference of -0.2 Da.
Sample 2: A sample of polyoxyethylene bis(azide) (Mn = 2000) (Figure 8) was analyzed using sodium trifluoroacetate as a cationization agent and HCCA as the matrix and only exhibited the expected Na+ adducts.
Because of the resolution achieved in this lower mass range, the monoisotopic peaks for each of the n-mers could easily be resolved, and so a monoisotopic peak picking protocol was selected (averaging only the mass signal of the first peak in the isotopic distribution) and all corresponding calculations utilized the monoisotopic masses of each element. MALDI-TOF MS confirms the narrow distribution (Figure 8) of polyoxyethylene bis(azide) (Mn = 2000). Using the equations from step 4, analysis software was used to calculate the following characteristics of the polymer mass distribution: Mn:1940, Mw: 1950, Đ: 1.01.
In order to confirm end group functionalization, an individual n-mer (42) was selected (Figure 9). As with the mass distributions determined above, monoisotopic masses were used because the monoisotopic peaks were well-resolved in each n-mer's isotopic distribution. The theoretical mass value of the 42-mer of polyoxyethylene bis(azide) corresponds to 44.02621 × 42 (the mass of the repeat units) + 42.00922 (mass of the azido end group) + 70.04052 (mass of the azidoethyl end group) + 22.98922 (mass of the sodium cation) = 1984.13978. The observed mass value for n = 42 is 1983.95 which is 0.19 Da different from the theoretical value. It should be noted that especially at higher laser powers, the azide functionality can exhibit metastable fragments; however, this was not observed in this specific case31.
The polyoxyethylene bis(azide) (Mn = 2000) sample maintained its narrow distribution when selectively functionalized by a copper catalyzed azide-alkyne cycloaddition reaction (Figure 10) with 1-ethynyl-4-fluorobenzene(EFB) (Figure 11) to yield a 4-fluorophenyltriazolyl (FPT) group. The spectra exhibited the expected Na+ adducts from using sodium trifluoroacetate as a cationization agent and HCCA as the matrix.
MALDI-TOF MS confirms the narrow distribution (Figure 11) of polyoxyethylene bis(azide) (Mn = 2000) after functionalization with EFB. Using the equations from step 4, analysis software was used to calculate the following polymer characteristics: Mn: 2240, Mw: 2250, Đ: 1.00.
To confirm complete functionalization of the sample, monoisotopic masses were used to analyze a selected individual n-mer (42) (Figure 12). The theoretical mass value of the 42-mer of polyoxyethylene bis(azide) reacted with 1-ethynyl-4-fluorobenzene corresponds to 44.02621 × 42 (the mass of the repeat units) + 162.04675 (mass of the FPT end group) + 190.07805 (mass of the FPT ethyl end group with 1-ethynyl-4-fluorobenzene) + 22.98922 (mass of the sodium cation) = 2224.21484. The observed mass value for n = 42 is 2224.16 which is 0.05 Da different from the theoretical value.
Sample 3: A sample of poly(L-lactide), thiol terminated (Mn = 2500) (Figure 13) was analyzed using sodium trifluoroacetate as a cationization agent and only exhibited the expected Na+ adducts and DHB as the matrix.
MALDI-TOF MS confirms the narrow distribution of poly(L-lactide), thiol terminated (Mn = 2500) (Figure 13). Using the equations from step 4, the program analysis was used to calculate the following polymer characteristics: Mn: 2310, Mw: 2360, Đ: 1.02.
To confirm complete functionalization of the sample, monoisotopic masses were used to analyze a selected individual n-mer (26) (Figure 14). The theoretical mass value of the 26-mer of poly(L-lactide), thiol terminated (Mn = 2500) corresponds to 72.02113 × 26 (the mass of the repeat units) + 17.00274 (mass of the hydroxyl group) + 61.0112 (mass of ω-thiol end group) + 22.98922 (mass of the sodium cation) = 1973.55254. The observed mass value for n = 26 is 1973.62 which is -0.07 Da different from the theoretical value. A smaller signal is observed at 2045.74 which corresponds to 72.02113 × 27 (the mass of the repeat units) + 17.00274 (mass of the hydroxyl end group) + 61.0112 (mass of ω-thiol end group) + 22.98922 (mass of the sodium cation). The theoretical mass is 2045.57367 which is a 0.17 difference from the observed mass. This small intensity, odd-numbered repeat unit is indicative of transesterification during the ring opening polymerization of lactic acid. A third, very minor peak is observed at 2057.73. This is -0.14 Da different than the theoretical mass of a poly(L-lactide) with a carboxylic acid end group (rather than the thiol end group) with a theoretical mass of 72.02113 × 27 (the mass of the repeat units) + 17.00274 (mass of the hydroxyl end group) + 73.02895 (mass of carboxylic acid) + 22.98922 (mass of the sodium cation) = 2057.59142. This additional minor impurity is likely the consequence of initiation from water during the ring opening polymerization of the lactide monomer.
The poly(L-lactide), thiol terminated (Mn = 2500) sample maintained its narrow distribution when selectively functionalized by a thiol-ene reaction (Figure 15) with maleimide (Figure 16). The spectra exhibited the expected Na+ adducts from using sodium trifluoroacetate as a cationization agent and DHB as the matrix.
MALDI-TOF MS confirms the narrow distribution of the poly(L-lactide), thiol terminated (Mn = 2500) after a thiol-ene reaction with maleimide (Figure 16). Using the equations from step 4, analysis software was used to calculate the following polymer characteristics: Mn: 2310, Mw: 2340, Đ: 1.01. It should be noted that the decrease of Mn and Mw in comparison to the starting material is due to ionization bias (one of the shortcomings of MALDI-TOF MS). When the modification to the starting material is relatively small (~97 Da in this particular modification) and the dispersity decreases post-modification, MALDI-TOF MS calculations of average molecular weight can become less accurate.
To confirm complete functionalization of the poly(L-lactide), thiol terminated (Mn = 2500) with maleimide via a thiol-ene reaction, monoisotopic masses were used to analyze a selected individual n-mer (26) (Figure 17). The theoretical mass value of the 26-mer of poly(L-lactide) thiol terminated corresponds to 72.02113 × 26 (the mass of the repeat units) + 17.00274 (mass of the hydroxyl end group) + 158.02757 (mass of ω- thiol end group linked to maleimide) + 22.98922 (mass of the sodium cation) = 2070.56891. The observed mass value for n = 26 is 2070.54 which is 0.03 Da different from the theoretical value. The same species ionizing with potassium is also observed at 2086.49, which corresponds to a 0.05 Da difference form the theoretical mass. A very small peak is observed at 2167.58 which corresponds to 72.02113 × 28 (the mass of the repeat units) + 17.00274 (mass of the hydroxyl end group) + 72.02168 (mass of carboxylate anion) + 22.98922 (mass of the sodium cation) + 38.96371 (mass of potassium cation). The theoretical mass is 2167.56844 which is a -0.01 difference from the observed mass and is indicative of the same trace impurity from water initiation that was observed in the starting material. This polymer exhibits ionization with one equivalent of sodium, one of potassium, and loss of a proton. The loss of the carboxylic acid proton and complexation with two cations is a common mode of ionization for monocarboxylic acid functionalized polymers. It is important to note that the same shift in mass that is observed for the thiol-ene reaction products does not occur for this carboxylic acid-terminated compound which further indicates that it lacked the thiol end group to undergo the functionalization reaction.
Figure 1: 3 x 3 grid for sample ratio determination. Using a 3 x 3 grid of samples, the relative concentrations of cationization agent-analyte-matrix can be systematically varied to empirically determine an optimized sample preparation. This is typically done by holding one of the three variables constant (15 µL of analyte solution) while increasing the amount of the other two (cationization agent (y-axis) and matrix (x-axis)) components by a set multiple (3-fold in the example depicted). Please click here to view a larger version of this figure.
Figure 2: MALDI-TOF MS target plate. The MALDI-TOF MS target plate is a metal plate which holds the MALD-TOF MS samples in individual wells for analysis. Please click here to view a larger version of this figure.
Figure 3: MALDI-TOF mass spectrum of Sample 1. This full spectrum shows the overall distribution of poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn= 5000) ionized with both Na+ and K+. Please click here to view a larger version of this figure.
Figure 4: MALDI-TOF mass spectrum of an individual repeat unit of Sample 1. This spectrum shows an individual repeat unit of poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn = 5000) for end group analysis. Please click here to view a larger version of this figure.
Figure 5: Reaction scheme for Sample 1 modification. To confirm the end groups of the starting material, poly(ethylene glycol) 2-aminoethyl ether acetic acid was reacted with 2,4-dinitrofluorobenzene (also known as Sanger Reagent). Please click here to view a larger version of this figure.
Figure 6: MALDI-TOF mass spectrum of Sample 1 modification. This full spectrum shows the overall distribution of poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn = 5000) functionalized with 2,4-dinitrofluorobenzene. Please click here to view a larger version of this figure.
Figure 7: MALDI-TOF mass spectrum of an individual repeat unit of Sample 1modification. In order to confirm end group functionalization, this spectrum shows an individual repeat unit of poly(ethylene glycol) 2-aminoethyl ether acetic acid (Mn = 5000) after reaction with 2,4-dinitrofluorobenzene. Please click here to view a larger version of this figure.
Figure 8: MALDI-TOF mass spectrum of Sample 2. This full spectrum shows the overall distribution of polyoxyethylene bis(azide)(Mn = 2000) ionized with Na+ adducts. Please click here to view a larger version of this figure.
Figure 9: MALDI-TOF mass spectrum of an individual repeat unit of Sample 2. This spectrum shows a repeat unit of polyoxyethylene bis-azide (Mn = 2000) in order to confirm end groups Please click here to view a larger version of this figure.
Figure 10: Reaction scheme for Sample 2 modification. To confirm the end groups of the starting material, polyoxyethylene bis-azide (Mn = 2000) was reacted with 1-ethynyl-4-fluorobenzene via a copper-catalyzed azide-alkyne cycloaddition (CuAAC). Please click here to view a larger version of this figure.
Figure 11: MALDI-TOF mass spectrum of Sample 2 modification. This full spectrum shows the overall distribution of polyoxyethylene bis(azide) (Mn = 2000) functionalized with 1-ethynyl-4-fluorobenzene. Please click here to view a larger version of this figure.
Figure 12: MALDI-TOF mass spectrum of an individual repeat unit of Sample 2 modification. This spectrum shows an individual repeat unit of polyoxyethylene bis(azide) (Mn = 2000) reacted with 1-ethynyl-4-fluorobenzene via copper catalyzed azide-alkyne cycloaddition to confirm end group functionalization. Please click here to view a larger version of this figure.
Figure 13: MALDI-TOF mass spectrum of Sample 3. This full spectrum shows the overall distribution of poly(L-lactide), thiol terminated (Mn = 2500). Please click here to view a larger version of this figure.
Figure 14: MALDI-TOF mass spectrum of an individual repeat unit of Sample 3. The spectrum shows an individual repeat unit of poly(L-lactide), thiol terminated (Mn = 2500) to confirm end groups. Please click here to view a larger version of this figure.
Figure 15: Reaction scheme for Sample 3 modification. To confirm the end groups of the starting material, poly(L-lactide), thiol terminated (Mn = 2500) was reacted with maleimide via a thiol-ene coupling. Please click here to view a larger version of this figure.
Figure 16: MALDI-TOF mass spectrum of Sample 3 modification. This full spectrum shows the overall distribution of the product of the reaction between poly(L-lactide), thiol terminated (Mn = 2500) and maleimide. Please click here to view a larger version of this figure.
Figure 17: MALDI-TOF mass spectrum of an individual repeat unit of Sample 3 modification. In order to confirm end group functionalization, this spectrum shows an individual repeat unit of poly(L-lactide), thiol terminated (Mn = 2500) after the thiol-ene reaction with maleimide. Please click here to view a larger version of this figure.
MALDI-TOF mass spectrometry is an invaluable analytical tool for polymer characterization because of its ability to generate polymer ions in the singly charged state and with minimal fragmentation. This soft ionization technique utilizes short laser pulses to desorb solid-state samples of the polymer analyte embedded in a matrix compound to generate polymer ions in the gas phase. The macromolecules are typically ionized by complexation with cations that are added to the matrix to enable their analysis by mass spectrometry. These macromolecular ions are then accelerated by an extraction voltage to bring them into the field-free region of the flight tube which can enable their m/z to be determined based on their time-of-flight between the ion source and the detector5,32.
Compared to other polymer characterization techniques, MALDI-TOF MS spectra quality is heavily dependent on data acquisition parameters and sample preparation. Although there is no set formula for sample preparation, understanding the function of each component of the sample preparation allows for more rapid empirical optimization. The most important factor in MALDI sample preparation is selection of the matrix because compatibility of the matrix with the polymer analyte is critical for allowing excited matrix to generate single, desorbed macromolecules in an ionized state5,15,17,19. Once appropriate matrix and cationization agents have been selected, the correct ratio of analyte, matrix, and cationization agent must be determined. This can be achieved empirically by creating a two-dimensional grid of samples (Figure 1) on the MALDI-TOF MS target plate (Figure 2) with increasing matrix concentration on one axis and increasing cationization agent concentration on the other.
Similar to MALDI sample preparation, there is no set formula for determining data acquisition parameters; however, certain trends should be considered to expedite spectral optimization. Reflectron mode, which increases resolution but decreases overall signal, is typically chosen for lower mass ranges (in these examples, below 4,000 Da) where isotopic resolution can be achieved. In these cases, monoisotopic mass calculations and peak picking methods were used. For polymer samples with masses above 4,000 Da, linear mode was used with average mass calculations and peaking picking methods. To improve signal resolution, the ion source voltages should be adjusted in small increments with the general trend of larger mass polymers having a larger voltage differential (IS1 versus IS2).
While optimized sample preparation and acquisition parameters can provide precision, mass accuracy can only be achieved through effective calibration. The time-of-flight for a given mass can vary subtly with respect to the variable acquisition parameters and even plate positions, therefore a calibration should be carried out for each set of optimized acquisition parameters in order to yield accurate mass determinations5,30. Once the acquisition parameters and sample preparation have been optimized, the spectra should be calibrated using these exact same conditions.
Because of the exceptional resolution and mass accuracy observed in the optimized MALDI-TOF mass spectra of polymers, this technique has become a valuable complimentary tool for determining polymer mass distribution data. However, its ability to resolve individual repeat units within the polymer mass distribution provides a particular advantage for end group analysis relative to other polymer characterization techniques such as gel permeation chromatography (GPC) and nuclear magnetic resonance (NMR). This is particularly valuable for determining the fidelity of end group functionalization reactions and the quantitative nature of end group conjugations reactions. This manuscript has demonstrated the ability to resolve the mass of individual polymer repeat units with up to two decimals points of mass accuracy, enabling the confirmation of end group modifications with a high level of confidence. With the substantial advances that have been made recently in the field of precision polymer synthesis, MALDI-TOF MS is becoming an increasingly important tool for determine macromolecular structure and functionality.
The authors have nothing to disclose.
The authors acknowledge the Smart MAterials Design, Analysis, and Processing consortium (SMATDAP) funded by the National Science Foundation under cooperative agreement IIA-1430280 and the LA Board of Regents for a graduate fellowship (MEP). Polymer samples for these experiments were provided by MilliporeSigma (Sigma-Aldrich). Open Access publication of this article is sponsored by MilliporeSigma.
polyoxyethylene bis(azide) (Mn=2000) | MilliporeSigma (Aldrich) | 689696 | http://www.sigmaaldrich.com/catalog/product/aldrich/689696?lang=en®ion=US |
poly(ethylene glycol) 2-amino-ethyl ether acetic acid (Mn= 5000) | MilliporeSigma (Aldrich) | 757918 | http://www.sigmaaldrich.com/catalog/product/aldrich/757918?lang=en®ion=US |
poly(L-lactide), thiol terminated (Mn=2500) | MilliporeSigma (Aldrich) | 747386 | http://www.sigmaaldrich.com/catalog/product/aldrich/747386?lang=en®ion=US |
SpheriCal® peptide low | MilliporeSigma (Sigma-Aldrich) | PFS20 | http://www.sigmaaldrich.com/catalog/product/sial/pfs20?lang=en®ion=US |
SpheriCal® peptide medium | MilliporeSigma (Sigma-Aldrich) | PFS21 | http://www.sigmaaldrich.com/catalog/product/sial/pfs21?lang=en®ion=US |
SpheriCal® peptide high | MilliporeSigma (Sigma-Aldrich) | PFS22 | http://www.sigmaaldrich.com/catalog/product/sial/pfs22?lang=en®ion=US |
2,4 dinitrofluorobenzene | TCI | A5512 | |
maleimide | MilliporeSigma (Aldrich) | 129585 | http://www.sigmaaldrich.com/catalog/product/aldrich/129585?lang=en®ion=US |
1-ethynylfluorobenzene | Fisher Scientific | 766-98-3 | |
triethylamine | MilliporeSigma (Sigma-Aldrich) | 471283 | http://www.sigmaaldrich.com/catalog/product/sial/471283?lang=en®ion=US |
N,N,N',N",N"-pentamethyldiethylenetriamine | MilliporeSigma (Aldrich) | 369497 | http://www.sigmaaldrich.com/catalog/product/aldrich/369497?lang=en®ion=US |
Copper(I)Bromide | MilliporeSigma (Aldrich) | 254185 | http://www.sigmaaldrich.com/catalog/product/aldrich/254185?lang=en®ion=US |
glacial acetic acid | Fisher Scientific | A38212 | |
sodium metabisulfite | MilliporeSigma (Sigma-Aldrich) | 13459 | http://www.sigmaaldrich.com/catalog/product/sigald/13459?lang=en®ion=US |
potassium trifluoroacetate | MilliporeSigma (Aldrich) | 281883 | http://www.sigmaaldrich.com/catalog/product/aldrich/281883?lang=en®ion=US |
trans-2-[3-(tert-butylphenyl)-2-methyl-2-properylidene]malononitrile | MilliporeSigma (Aldrich) | 727881 | http://www.sigmaaldrich.com/catalog/product/aldrich/727881?lang=en®ion=US |
a-cyano-4-hydroxycinnamic acid | MilliporeSigma (Sigma) | C8982 | http://www.sigmaaldrich.com/catalog/product/sigma/c8982?lang=en®ion=US |
tetrahydrofuran | Fisher Scientific | T425-1 | |
dichloromethane | VWR Analytical | BDH1113-4LG |