Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level

Published: June 06, 2018
doi:

概要

Here, we present the portable autoRAD platform for quick radiometric separation and determination of 90Sr, an important fission product highly pertinent to nuclear waste.

Abstract

A quick, automated and portable system for the separation and determination of radiostrontium in aqueous samples, using Sr-resin and multi sequential flow injection analysis, has been developed. The concentrations of radioactive strontium were determined by flow scintillation counting, allowing for on-line and also on-site determination. The proposed system can determine radioactive strontium at industrial relevant levels without further modification using overall analysis time of less than 10 min per aqueous sample. The limit of the detection is 320 fg·g-1 (1.7 Bq/g).

Introduction

About 150 commercial nuclear power plants (NPP) are undergoing decommissioning, but the total number of nuclear facilities is much greater if research and reprocessing facilities are taken into consideration1. Decommissioning of nuclear facilities is very costly and involves transport and off site measurement of contaminated material. Savings are possible by adopting on-site and flexible measurement techniques1. There is therefore an urgent need for rapid on site analytical methods supporting the decommissioning of old nuclear facilities. Gamma-emitters can be easily and selectively determined using gamma-spectroscopy, but there is a deficiency in on-site analytical methods for the hard-to-measure (HTM) radionuclides2. Among the HTM, 90Sr is of great interest due to its toxicity and high energy emission. Its determination is time consuming and requires the separation from interferences and quantification using LSC or spectroscopic methods3,4,5,6,7,8,9,10,11,12,13.

Standard radiochemical methods are time-consuming and often require repetitions to obtain an adequate yield of strontium. Therefore, there is an urgent need for faster and accurate methods. Besides the standard separation protocols, flow injection techniques are applied in the purification and/or pre-concentration of radioisotopes5,14,15. Lab-on-Valve (LOV) devices are a further development in flow techniques. They are programmable, flow-based platforms with different coupling modes and exhibit high versatility16. Such devices allow the automatic separation and pre-concentration of the analytes prior to the detection, thus increasing reproducibility and repeatability8,17,18,19. LOV systems coupled to multisyringe flow pumps have been widely used for the detection of radionuclides, due to the minimization of the reagent consumption and waste generation8,10,17,18,19,20. Nevertheless, reported studies regarding online detection at ultra-trace levels are sparse8,17.

On-site measurements of radioactive materials have many benefits and advantages, but there is no application of flow scintillation to 90Sr monitoring. Basically, quantification is achieved using an online detector that mixes the eluate coming from the LOV device with the scintillating cocktail11,21,22. The mixture is then pumped through the counting cell and measurement is made using paired photomultipliers. Due to the small volume of the cell, the measurement time is in the seconds scale.

The aim of this research is the development of a fully automated method for on-line strontium detection over a large working range, covering both environmentally relevant concentrations and also those found in nuclear industry waste streams. The platform is mobile and can be mounted in a vehicle to perform on-site analysis of aqueous samples.

Protocol

Note: Solutions were prepared from high-purity water (18.2 MΩ cm) using devices designed for ultratrace analysis. Nitric acid was purified using a quartz sub-boiling distillation unit. Both the water purification system and the sub-boiling distillation unit were operated in a clean room.

Caution: 90Sr is acutely toxic and carcinogenic. Appropriate safety practices are essential when performing the experiments, including both engineering control systems and personal protective equipment.

1. Experiment Preparation

Note: A detailed description regarding the software architecture and its functionalities can be found elsewhere23.

  1. Connect the Modular Valve Position exit port to the entry port of the radio flow detector.
  2. Connect the Trigger line to selected detector.
  3. Ensure that the LOV ports are properly connected. Ensure that adequate volumes of working solutions are available for the entire protocol and that liquid sampling tubing will remain submerged.
  4. Ensure that the Autosampler is on, start the autosampler software, and via the software interface, initialize the autosampler. Click the Initialize button to generate the communication between the apparatus and the PC.
  5. Ensure the AutoRAD software has been started, check the communication ports using the options tab and initialize the software via the user interface. Refer to Discussion section for more details regarding the software graphical interface.
  6. Ensure that the analysis sequence is programmed in the detector Software. Program vew sequences using the Method editor tab in the AutoRAD software, by entering the number of steps needed and the tasks and speed for each device.
    Note: A detailed description of the software operation can be found in previous publication23.
  7. Consult the local radioprotection officer and employ the radioprotection measurements needed for the complete procedure using 90Sr.

2. System Cleaning

  1. Load 10 mL of 18.2 MΩ cm water from the flask to the syringe at 90 mL·min-1. Ensure that the syringe valve position is set to the In position for this step. In this position, loading occurs directly to the syringe and not through the LOV.
  2. Drop the water to the waste via the holding coil at a flow rate of 90 mL·min-1.
  3. Load 3 mL of ethanol to the holding coil at a flow rate of 3 mL·min-1. Set the syringe valve position to Out.
  4. Drop the ethanol to the detector coil at a flow rate of 3 mL·min-1.

3. Load Resin to the LOV

  1. While stirring the resin suspension in water (12 mg·mL-1), pick up 3 mL via the LOV. Set the flow to 3 mL·min-1
  2. Drop the resin suspension into the column channel at a flow rate of 1.2 mL·min-1
  3. Clean the holding coil and drop the resin residues onto the column channel. For this, load 9 mL of 18.2 MΩ cm water from the flask to the syringe at a flow rate of 90 mL·min-1. Ensure that the syringe valve position is in the In position. Drop the water to the waste via the holding coil at a flow rate of 3 mL·min-1.

4. Analysis Sequence

  1. Column conditioning
    1. Load 2 mL of HNO3 (4 M) via the LOV to the holding coil. Set the flow to 6 mL·min-1.
    2. Drop the HNO3 onto the column at a flow rate of 1.2 mL·min-1.
  2. Sample loading and elimination of interferences
    1. Load sample (1.3 mL) from the autosampler to the holding coil at a flow rate of 6 mL·min-1.
    2. Drop the sample onto the column at a flow rate of 1.2 mL·min-1.
    3. Load 0.5 mL of HNO3 (4 M) to the holding coil at a flow rate of 6 mL·min-1.
    4. Rinse the column with 0.5 mL of HNO3 (4 M) to elute matrix interferences at a flow rate of 1.2 mL·min-1.
  3. Elution of the sample and measurement
    1. Load 5 mL of 18.2 MΩ cm water to the holding coil at a flow rate of 6 mL·min-1.
    2. Trigger the detector. Set the flow rate of the scintillation liquid to 2 mL·min-1 in the detector software. Set the dwell to 10 s. The sample coil has a volume of 2 mL.
    3. Rinse the column.
  4. rinsing of the sample probe and holding coil
    1. Load 0.6 mL of HNO3 (1%) to the holding coil at a flow rate of 6 mL·min-1.
    2. Load 0.6 mL of air to the holding coil at a flow rate of 6 mL·min-1.
    3. Discharge 1.2 mL of the mixture to the waste.
  5. Resin replacement
    1. Load 0.2 mL of ethanol to the holding coil at a flow rate of 3 mL·min-1.
    2. Rinse the column with 0.2 mL of ethanol at a flow rate of 1.2 mL·min-1.
    3. Flush the column channel with 0.5 mL of water at a flow rate of 0.45 mL·min-1.
    4. Discharge the used resin to the waste.

Representative Results

The fully automated AUTORAD platform operated by homemade LabVIEW-based software has been developed and implemented (Figure 1). The software allows sufficient flexibility for daily operation in a user-friendly environment (Figure 2). The versatility of the platform has been also demonstrated by coupling it to different detectors (Figure 3). The applicability of the method has been demonstrated in aqueous samples spiked with a 90Sr standard (Figure 4). The parameters linearity, linear range, limit of detection (LOD) and repeatability have been evaluated (Figure 5).

Figure 1
Figure 1. Schematic representation of the AutoRAD system showing the port configuration and reagents used. The central port is connected to the syringe pump via the holding coil (10 mL); the methacrylate LOV was fabricated in house. Please click here to view a larger version of this figure.

Figure 2
Figure 2. AutoRAD Software options menu. The bottom right corner area shows the assigned ports. Via the scroll down menu, the user is able to make changes in the assignment. Please click here to view a larger version of this figure.

Figure 3
Figure 3. 86Sr elution peaks using an Inductively coupled plasma mass spectrometry (ICP-MS) as a detector. Strontium is eluted quantitatively from the column during the first 100 s of the elution. Please click here to view a larger version of this figure.

Figure 4
Figure 4. 34.5 ± 1 Bq 90Sr (6.6 pg·g-1) elution peak using ß-RAM 5 as detector. The residence time in the coil was 40 s. The overall recovery rate of the proposed methodology for 90Sr is 70% ± 5%. Please click here to view a larger version of this figure.

Figure 5
Figure 5. Calibration curve for the count rate versus the concentration of radiostrontium. A good linearity was obtained (R2 = 0.997). The detection limit was calculated as 320 fg·g-1 ± 5 (1.7 Bq) by three point three times standard deviation of the blank. Please click here to view a larger version of this figure.

Discussion

Radiological characterization and protection are critical issues during all phases of a nuclear facility life cycle. The need for radionuclide determination during the decommissioning of nuclear facilities requires continuous improvement of analytical procedures. This entails the enhancement of selectivity and sensitivity, and shortening the analysis time . The AUTORAD prototype fulfils these requirements. In addition, the prototype is portable and allows on-site determination. The proposed automated on-line method was successfully applied for the determination of 90Sr activity in aqueous samples.

Figure 1 shows a schematic diagram of the AutoRAD system. Communication with the prototype was achieved using the commercial graphical programming language LabVIEW 2014. The Virtual Instrument Software Architecture (VISA) mode for configuring, programming and controlling interfaces was deployed. Figure 2 shows the graphical user interface screen where the port configuration can be verified prior to system initialization. The VISA mode is independent on the operating system and program environment, so it provides exceptional versatility. The basic structure and features of the developed software have been widely discussed in previous publication24. Improvements have been achieved by controlling the burettes by the PC using an RS232 interface, thus allowing enhanced control of both burettes simultaneously and the possibility to perform measurements in stop flow mode. In contrast, the RS232 connection reacts quickly when the system backpressure increases. This can lead to system error and measurement stop. Therefore, special attention has to be paid in the process of resin loading and ionic strength of the samples.

The optimization of the experimental conditions was achieved in a set of experiments using the strontium stable isotope 86Sr as a surrogate for the radioactive 90Sr, and by coupling the AUTORAD system to an ICP-MS rather than a beta detctcor system. Figure 3 shows the ICP-MS 86Sr elution profiles. The obtained 86Sr elution profiles are in full agreement with previously reported results for total strontium using low pressure separation devices25, and have been blank subtracted. The least squares linear regression method was used to generate a fit for the purpose curve which does not introduce an extra uncertainty component. The linearity obtained was 0.995, with p-values less than the significance level. The LOD of 2 pg·g-1was determined by repeated measurements of the blank according to Currie26. The reproducibility of the method, based on the relative standard deviation of the peak area calculated on the basis of three repeated runs, was always less than 4% in the range of 10 to 120 pg·g-1. The shoulder witnessed in the elution profiles is most likely an artefact due to non-optimal packing of the column in the automated AUTORAD configuration.

Figure 4 shows the 90Sr profile using the the radio flow detector. The AUTORAD system is able to effectively separate 90Sr in aqueous samples.

Figure 5 shows the dependence of the 90Sr concentration with the signal is linear in the range of interest. The derived detection limit is in the fg. g-1 range, enabling, without further modification, the determination of radiostrontium in nuclear decommissioning and waste characterization samples. Reproducibility of the method, using the flow scintillation detector and based on the relative standard deviation of the peak area, is around 30% in the studied concentration range. The current set-up of the prototype, however, limits its application to environmental samples, essentially due to the short time counting in the detector. Furthermore, complex matrices might saturate the resin in the mini column.

The stopped-flow technique, where both Hamilton and radio flow detector pump are stopped during the measurement, will be implemented. This feature will improve the sensitivity by extending the residence time in the largest part of the sample zone within the flow cell. Thus, a statistically meaningful signal is acculated before the sample exits the detector. This approach will improve both the counting statistics and the detection limits. In addition, a new application is being developed including an additional ion-exchange mini-column for the removal of matrix components that could affect the retention and separation of different radionuclides in complex matric samples.

開示

The authors have nothing to disclose.

Acknowledgements

The authors would like to aknowledge the colleagues and people involved in the different steps of the project. The colleagues in the JRC Karlsruhe design office, Messrs. Dietrich Knoche and Volkmar Ernest, in the JRC Karlsruhe Workshop Messrs. Christian Diebold and Joachim Küst for the design and manufacturing of the lab on valve (LOV) manifold and Ringwald GmbH for the software implementation of the stopped-flow detection mode technique.

Materials

microLAB 600 series Hamilton ML600EE6910 Two dual syringe pump system, equiped with two 10 mL glass syringes. The instruments are interconnected using the CAN port (daisy chain). 
FlowLogic U LabLogic SG-BXX-05 Liquid scintillator with high flash point
ß-RAM 5  LabLogic flow detector, 2000 μL coiled Teflon flow cell.Software  Laura 4.2.8 (LabLogic, England) run on desktop PC and connected to the detector via USB
SC-μ DX Autosampler Elemental Scientific Instruments (ESI)
Cheminert selector Valco Instruments Co. Inc.  in-house made Lab-on-Valve has been mounted on this selector
Modular Valve
Positioner (MVP)
Hamilton
mini magnetic stirrer IKA
Nitric Acid Suprapur 65% Merck 1.00441.1000 purified using quartz sub-boiling distillation unit
Sr-resin Eichchrom Tecnologies, Inc SR-B100-A particle size 100-150 µm
Water system Elix 3 in combination with Mili-Q Element A10 Millipore high-purity water (18.2 MΩ cm) 
Sr-90 standard Eckert & Ziegler 7090 Sr-90 concentration 1.915 kBq/g ± 3.0%, reference date 15-May-2016 12:00 PST
MLS quartz sub-boiling distillation unit MLS GmbH Subboiling unit for the purification of HCl and nitric acid

参考文献

  1. Zapata-García, D., Wershofen, H. Development of radiochemical analysis strategies for decommissioning activities. Applied Radiation and Isotopes. 126, 204-207 (2017).
  2. Hou, X. Radiochemical analysis of radionuclides difficult to measure for waste characterization in decommissioning of nuclear facilities. Journal of Radioanalytical and Nuclear Chemistry. 273 (1), 43-48 (2007).
  3. Chung, K. H., et al. Rapid determination of radiostrontium in milk using automated radionuclides separator and liquid scintillation counter. Journal of Radioanalytical and Nuclear Chemistry. 304 (1), 293-300 (2015).
  4. Desmartin, P., Kopajtic, Z., Haerdi, W. Radiostrontium-90 (90Sr) Ultra-Traces Measurements by Coupling Ionic Chromatography (HPIC) and on Line Liquid Scintillation Counting (OLLSC). Environmental Monitoring and Assessment. 44 (1), 413-423 (1997).
  5. Grate, J. W., Strebin, R., Janata, J., Egorov, O., Ruzicka, J. Automated Analysis of Radionuclides in Nuclear Waste: Rapid Determination of 90Sr by Sequential Injection Analysis. Analytical Chemistry. 68 (2), 333-340 (1996).
  6. Holmgren, S., Tovedal, A., Björnham, O., Ramebäck, H. Time optimization of 90Sr measurements: Sequential measurement of multiple samples during ingrowth of 90Y. Applied Radiation and Isotopes. 110, 150-154 (2016).
  7. Kavasi, N., et al. Measurement of 90Sr in soil samples affected by the Fukushima Daiichi Nuclear Power Plant accident. Journal of Radioanalytical and Nuclear Chemistry. 303 (3), 2565-2570 (2015).
  8. Kołacińska, K., et al. Automation of sample processing for ICP-MS determination of 90Sr radionuclide at ppq level for nuclear technology and environmental purposes. Talanta. 169, 216-226 (2017).
  9. Lazare, L., Crestey, C., Bleistein, C. Measurement of 90Sr in primary coolant of pressurized water reactor. Journal of Radioanalytical and Nuclear Chemistry. 279 (2), 633-638 (2009).
  10. Mola, M., et al. Determination of 90Sr and 210Pb in sludge samples using a LOV-MSFIA system and liquid scintillation counting. Applied Radiation and Isotopes. 86, 28-35 (2014).
  11. Plionis, A. A., Gonzales, E. R., Landsberger, S., Peterson, D. S. Evaluation of flow scintillation analysis for the determination of Sr-90 in bioassay samples. Applied Radiation and Isotopes. 67 (1), 14-20 (2009).
  12. Temba, E. S. C., Reis Júnior, A. S., Amaral, &. #. 1. 9. 4. ;. M., Monteiro, R. P. G. Separation and determination of 90Sr in low- and intermediate-level radioactive wastes using extraction chromatography and LSC. Journal of Radioanalytical and Nuclear Chemistry. 290 (3), 631-635 (2011).
  13. Tomita, J., Yamamoto, M., Nozaki, T., Tanimura, Y., Oishi, T. Determination of low-level radiostrontium, with emphasis on in situ pre-concentration of Sr from large volume of freshwater sample using Powdex resin. Journal of Environmental Radioactivity. 146, 88-93 (2015).
  14. Egorov, O., Grate, J. W., Ruzicka, J. Automation of radiochemical analysis by flow injection techniques: Am-Pu separation using TRU-resin™ sorbent extraction column. Journal of Radioanalytical and Nuclear Chemistry. 234 (1), 231-235 (1998).
  15. Rodríguez, R., Avivar, J., Leal, L. O., Cerdà, V., Ferrer, L. Strategies for automating solid-phase extraction and liquid-liquid extraction in radiochemical analysis. TrAC Trends in Analytical Chemistry. 76, 145-152 (2016).
  16. Miró, M., Oliveira, H. M., Segundo, M. A. Analytical potential of mesofluidic lab-on-a-valve as a front end to column-separation systems. TrAC Trends in Analytical Chemistry. 30 (1), 153-164 (2011).
  17. Kołacińska, K., et al. Automation of sample processing for ICP-MS determination of 90Sr radionuclide at ppq level for nuclear technology and environmental purposes. Talanta. , (2016).
  18. Rodríguez, R., Avivar, J., Ferrer, L., Leal, L. O., Cerdà, V. Automated total and radioactive strontium separation and preconcentration in samples of environmental interest exploiting a lab-on-valve system. Talanta. 96, 96-101 (2012).
  19. Rodríguez, R., et al. Automation of 99Tc extraction by LOV prior ICP-MS detection: Application to environmental samples. Talanta. 133, 88-93 (2015).
  20. Villar, M., et al. Automatic and Simple Method for 99Tc Determination Using a Selective Resin and Liquid Scintillation Detection Applied to Urine Samples. Analytical Chemistry. 85 (11), 5491-5498 (2013).
  21. L’Annunziata, M. F. . Handbook of Radioactivity Analysis. , 1117-1178 (2012).
  22. Roane, J. E., DeVol, T. A., Leyba, J. D., Fjeld, R. A. The use of extraction chromatography resins to concentrate actinides and strontium from soil for radiochromatographic analyses. Journal of Environmental Radioactivity. 66 (3), 227-245 (2003).
  23. Barbesi, D., et al. A LabVIEW®-based software for the control of the AUTORAD platform: a fully automated multisequential flow injection analysis Lab-on-Valve (MSFIA-LOV) system for radiochemical analysis. Journal of Radioanalytical and Nuclear Chemistry. 313 (1), 217-227 (2017).
  24. Barbesi, D., et al. A LabVIEW®-based software for the control of the AUTORAD platform: a fully automated multisequential flow injection analysis Lab-on-Valve (MSFIA-LOV) system for radiochemical analysis. Journal of Radioanalytical and Nuclear. , 1-11 (2017).
  25. Strahlenschutz, F. f. . Moderne Routine- und Schnellmethoden zur Bestimmung von SR-89 und SR-90 bei der Umweltüberwachung : Bericht einer Ad-hoc-Arbeitsgruppe des Arbeitskreises Umweltüberwachung (AKU). , (2008).
  26. Currie, L. A. Limits for qualitative detection and quantitative determination. Application to radiochemistry. Analytical Chemistry. 40 (3), 586-593 (1968).

Play Video

記事を引用
Vicente Vilas, V., Millet, S., Sandow, M., Aldave de las Heras, L. Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level. J. Vis. Exp. (136), e57722, doi:10.3791/57722 (2018).

View Video