We designed and developed an effective nanopowder aerosolization setup and operating protocol. The system generated nanoparticle aerosols with stable number concentrations and size distributions for long durations, requiring only small quantities of test material (min. 200 mg).
Nanoparticle aerosols released from nanopowders in workplaces are associated with human exposure and health risks. We developed a novel system, requiring minimal amounts of test materials (min. 200 mg), for studying powder aerosolization behavior and aerosol properties. The aerosolization procedure follows the concept of the fluidized-bed process, but occurs in the modified volume of a V-shaped aerosol generator. The airborne particle number concentration is adjustable by controlling the air flow rate. The system supplied stable aerosol generation rates and particle size distributions over long periods (0.5-2 hr and possibly longer), which are important, for example, to study aerosol behavior, but also for toxicological studies. Strict adherence to the operating procedures during the aerosolization experiments ensures the generation of reproducible test results. The critical steps in the standard protocol are the preparation of the material and setup, and the aerosolization operations themselves. The system can be used for experiments requiring stable aerosol concentrations and may also be an alternative method for testing dustiness. The controlled aerosolization made possible with this setup occurs using energy inputs (may be characterized by aerosolization air velocity) that are within the ranges commonly found in occupational environments where nanomaterial powders are handled. This setup and its operating protocol are thus helpful for human exposure and risk assessment.
Nanomaterial powders are widely used in different industrial sectors, as the raw materials for manufacturing new products or as additives for their functional applications 1-4. However, the potential for workers' exposure to nanopowder aerosols has been noted during various occupational handling activities 5-8, and the associated health risks have been investigated in in-vivo and in-vitro toxicological studies 9-12. In order to facilitate the development of effective strategies to protect workers dealing with nanomaterials, occupational health professionals require a better understanding of how nanoparticle aerosols are generated from powdered materials subjected to external energy inputs.
Different laboratory systems have been developed to simulate powder aerosolization behavior in realistic conditions. Among them, two standard procedures are the established reference methods 13 for testing powder dustiness, which is defined as the tendency of a powder, subjected to a given energy input, to release airborne particles. The first method uses a rotating drum as the medium for energy input and the aerosolization of powder particles 14,15. The second method drops a powder at a constant rate through a vertical cylinder and aerosolizes the powder particles by means of an ascending air flow 16. However, these methods require relatively large quantities of test materials (35 cm3 or 500 g), and this can be a problem with nanomaterial powders due to their high cost and the potential exposure risks. A scaled-down test system combining single-drop and rotating drum processes was developed, allowing the use of smaller amounts of test powders (6 g samples) 17. A recently developed aerosolization system based on vortex shakers has also been used for nanopowders, allowing tests down to 1 cm3 of raw materials 18.
Here, we present a novel aerosolization and deagglomeration system for nanopowder testing based on laboratory funnels. It provides a stable aerosol generation process using less than 1 g of test powder. Steady aerosolization can be maintained for sufficiently long durations for robust aerosol characterizations. The system's performance has been described in detail in two previous publications 19,20.
The test setup is composed of an aerosol generator, mixing and measurement compartments, and characterization instruments, as shown in Figure 1. Particle transport tubing and connectors link these different elements. A flow tuner and two flowmeters control and monitor the air flow conditions in the system. A manometer and a temperature and humidity sensor monitor the environment inside the measurement chamber. Dry compressed air is filtered using a hyperfilter before it enters the system. An elongated, V-shaped, glass aerosol generator is used for powder aerosolization. This geometry facilitates a robust aerosolization process and smooth particle transport into the subsequent compartment. The flow regime at the bottom of the funnel is turbulent due to the interaction with the powder particles, whereas it is laminar in the upper section (Re-number < 15). The thickness of the generator walls was specifically designed to resist the high pressures (up to 400 kPa ΔP) needed for deagglomeration testing using critical orifices. A high-precision flow tuner controls the flow rate in 0.01 L/min increments. Conductive tubing (6 mm outer diameter, 1 mm thickness) is used to avoid particle losses due to electrostatic deposition during transportation. The tube length is about 50 cm between the aerosol generator and the mixing chamber, 20 cm between the mixing chamber and the measurement chamber, and 100 cm for the sampling tubes. A 1 L metallic bottle is used as the mixing chamber, and a 12 L metallic drum is used as the measurement chamber. Particle samples are drawn from the top of the measurement chamber. An exit port directs the extra flow into a filtration system. The mixing and measurement chambers are electrically grounded to prevent electrostatic losses of particles. The measurement instruments include a scanning mobility particle sizer (SMPS) and an optical particle counter (OPC) for particle number concentration and size distribution, and a transmission electron microscopy (TEM) sampler (MPS) for particle morphology analysis.
The setup's aerosolization procedure resembles a fluidized-bed process. The air flow enters from the bottom opening into the funnel (2 mm diameter) and aerosolizes the powder. The powder particles move in a similar way to water in a fountain. The aerosol generated meets a dilution flow in the mixing chamber. The dilution air flow can be conditioned for different levels of humidity if the effect of this parameter requires analysis. The chamber's air also serves as a buffer volume to smoothly mix the aerosol with dry dilution air according to the sampling needs. The aerosol flow is then introduced into the measurement chamber via a normal tube outlet (for aerosolization testing) or a critical orifice (for deagglomeration testing). The orifice can provide different pressure-drop conditions, applying shear forces to the particles passing through it. This mechanism allows the study of their deagglomeration potential (mechanical stability).
Figure 1. Diagram of the aerosolization and deagglomeration system. By default, a tube connects the mixing chamber with the measurement chamber. The orifice shown is an optional add-on (not described in this protocol). Please click here to view a larger version of this figure.
1. System Preparation
2. Material Preparation
3. Aerosolization
4. Characterization
5. Post-sampling Operations and Clean-up
Figure 2 shows a typical example of total aerosol particle number concentration and size changes over time, using the above protocols in an aerosolization experiment with hydrophobic SiO2. Particle concentrations started to rise as soon as the aerosolization flow was introduced. The geometric mean size of particles gradually increased as well. After approximately 10 SMPS scans (3.5 min/scan), the aerosol started to enter a steady state, where particle concentration and mean diameter no longer varied by any significant amount. This state lasted more than 30 min, which was sufficient to complete ten 3-min SMPS scans. Figure 3 shows the change in particle concentration in the form of individual size distributions (based on the same data as in Figure 2). The peak rose slowly over time, and once the aerosol became stable, it remained within the same size range throughout the rest of the test.
The very small mean diameter shown at the beginning of the experiment was not due to unstable powder aerosolization. Rather, it was caused by the residual ambient air inside the funnel after the powder filling procedure. This volume of air was the first to flow into the measurement chamber and was sampled by the SMPS during its initial scans (Figure 4). This could be avoided by carrying out all the experiments in a clean room if this was required by the scientific question at hand. Indeed, the size distribution of the first scan was very similar to that of the ambient air. As the powder aerosol particles continued to flow in, the interference from the ambient particles diminished rapidly, and the effect had nearly disappeared after a few SMPS scans.
Figure 2. Change in total particle number concentration and mean diameter in an aerosolization experiment (241 mg hydrophobic SiO2; aerosolization flow 0.3 L/min). Please click here to view a larger version of this figure.
Figure 3. Change in particle size distribution in an aerosolization experiment. Please click here to view a larger version of this figure.
Figure 4. Aerosol particle size distributions at the beginning of the aerosolization test. Particle concentration is presented on a relative scale (normalized to the total number) in order to compare the spectrum from the first scan in a very low concentration to spectra from later scans in higher concentrations. Please click here to view a larger version of this figure.
The changes in particle concentration do not always follow the same patterns. Four possibilities can usually be seen in an aerosolization test. In Figure 5A, the concentration slowly increased to a "plateau" region, then remained nearly unchanged for the rest of the experiment. In Figure 5B, the concentration first rose to a maximum point, gradually decreased to a low level, and then remained stable for more than 1.5 hr. In Figure 5C, the concentration continued decreasing to zero. In Figure 5D, the concentration increased to a maximum level, remained there for a certain period, and then decreased again.
Scenario (a) is usually seen when the standard operating procedure is followed. The aerosolization air flow is slowly introduced and finally stabilized within the proper range. The amount of raw material is sufficient with respect to the aerosolization level, and a constant aerosol generation rate can be maintained for a long period of time. Scenario (b) is most likely due to an excessive aerosolization flow throughout the experiment, combined with an insufficient quantity of powder. The powder is rapidly consumed and is not able to sustain stable aerosol generation. Scenario (c) shows a similar drop in particle number concentration to Scenario (b) except that after a short time, the air flow rate was re-adjusted to a suitable range and kept constant throughout the rest of the test. This allowed the particle concentration to gradually reach a stable range. Scenario (d) appears when an insufficient amount of raw material is used. At the latter phase of the experiment, there is no longer enough test powder to generate aerosol particles at a constant rate, as was possible in the early phase of aerosolization. Consequently, the particle concentration in the system decreases.
Figure 5. Typical patterns for changing total particle concentrations during aerosolization experiments: (A) slowly increase until a plateau is reached; (B) gradually decrease to zero; (C) rapidly reach a peak and then decrease to a stable level; (D) increase to a steady state and maintain for a certain period of time, then decrease. Please click here to view a larger version of this figure.
Different aerosolization flow rates were tested in order to study their influence on aerosol generation. Flow rates from 0.3-1.1 L/min were used, and the resulting particle size distributions are shown in Figure 6. The peak of the spectrum rose as the flow increased. At the highest flow rate (1.1 L/min), micron-sized airborne particles started to enter the system (the secondary peak). The modal sizes of the aerosol particles stayed similar when under the same aerosolization flow, however, they decreased gradually when air flow increased through the range from 0.3-0.7 L/min (Figure 7). The increasing particle generation rate and diminishing mean particle diameter as flow rates increased suggest that the more dynamic aerosolization process (with significant particle movements and collisions) facilitated deagglomeration of powder particles, resulting in a modified size distribution of the aerosol particles generated.
Figure 6. Changing particle size distributions with increasing air flow rates (0.3-1.1 L/min). Please click here to view a larger version of this figure.
Figure 7. Comparison of particle size distributions under different flow rates. The spectra were turned into similar heights in relative scale (normalized to total particle number), which shows better the shift of the peak. Please click here to view a larger version of this figure.
The particle number concentrations of aerosols generated in replicate tests can vary by up to several folds, but usually well within one order of magnitude. The mean particle size, on the other hand, is highly reproducible. Figure 8 shows an example of the variation in particle size distribution from four replicate tests using the same material. The standard deviation was 39.7% for the total particle concentration and 6.6% for the geometric mean size. The variation of the number concentration could be due to several reasons: 1) different raw material status (e.g., agglomeration level); 2) human factors in powder filling process (influence the powder quantity deposited at the funnel bottom, thus the amount available for aerosolization); or 3) air flow adjustment at the beginning of aerosolization.
Figure 8. Variation of test results from replicate aerosolization experiments with hydrophobic SiO2. The error bars represent the standard deviation of particle number concentration in individual size channels. Please click here to view a larger version of this figure.
The funnel based aerosolization setup can effectively activate powder particles at a chosen energy input level (may be quantified by air flow velocity during aerosolization). Particle movements and collisions at the generation site reach an equilibrium state, breaking up powder agglomerates and emitting airborne particles of the same size distribution at a constant rate. A stable aerosolization can last from 30 min to up to 2 hr, which is sufficient time for even slow measurement instruments with high size resolutions, such as the SMPS, to produce statistically significant results. The setup only requires small quantities of test materials, which can be an advantage for testing precious materials such as nanoparticle powders.
However, the system's environment and process parameters can significantly influence the test results. To produce repeatable data, standard operating procedures must be strictly followed throughout the experiments. When conducting aerosolization tests using this system, the following aspects should be carefully considered.
Firstly, to obtain meaningful results, it is critical that the setup's internal parts provide a clean environment for the tests. Potential sources of contaminants are ambient particles and test materials from previous experiments. The effect of ambient particles usually disappeared relatively quickly, as soon as the aerosolization and dilution flows were introduced. However, interference from residual materials can persist throughout the experiment. As the generated aerosol particles flow through the system, they can deposit on the inner walls of the transport tubes, the bending points and the narrow channels of the connectors, and the inner surfaces of the mixing and the measurement chambers. If these parts are not properly cleaned prior to new experiments, previously deposited materials can be constantly re-suspended into the main stream of the aerosol flow, thus disturbing the test results.
Secondly, the powder filling process should be conducted very carefully. The most significant issue here is the quantity of the powder fed into the setup, especially when very small amounts of materials are used. At a given aerosolization flow rate, smaller amounts of powder generate lower aerosol concentrations, and possibly particles with smaller sizes, due to the higher energy input per unit weight of powder. Moreover, the storage conditions for test materials (e.g., relative humidity and temperature) have been shown to influence powder aerosolization behavior and levels of dustiness 22. Therefore, raw powders should always be kept in the same atmospheric conditions, where possible.
Thirdly, adjustments to the aerosolization flow at the beginning of the experiment greatly affect the test results. Sharp increases in the flow blow large powder particles up into the air and spread them all over the funnel surface, dramatically reducing the amount of material available for the rest of the experiment. The consequences could be a failed test due to insufficient powder.
Because the setup described here is not built using standardized laboratory equipment, when attempting to replicate the core parts of this system, the following aspects should be considered. Standard laboratory separatory funnels can be used as the aerosol generator (note that they should not be used under pressurized conditions). Separatory funnels of different geometries were tested in the experiments, and they provided similar functionality to the tailored funnel. A rubber sealing block with an embedded transport tube can be used as the funnel lid.
Mixing and measurement compartments of different geometries but similar volumes can be used. Note that compartments that are too large will significantly delay the time needed to reach stable aerosol conditions (concentration). The time required can be estimated by taking into account the total air flow rate and the volume of the compartment. Although the process can be accelerated by using a large dilution flow, it should be remembered that the final particle number concentration can be dramatically decreased due to dilution, and this may influence the aerosol size distribution as well as the performance of the measurement instruments (depending on their detection limits). Electrically conductive materials are recommended.
The length of the transport tubing may vary, depending on the general laboratory settings. However, the length should be kept as short as possible in order to avoid significant particle losses during their transport. The particle penetration efficiency can be calculated by taking into account particle diameter, air flow rate, tube diameter and length, and bearing in mind either gravitational deposition or diffusion loss, or both.
Different characterization methods may be employed. However, the air supply (dilution flow) should be adjusted to match the total sampling flow rate. Insufficient air supply will result in negative pressure in the measurement chamber, drawing in ambient particles thus leading to mistakes in the conclusions. Different air supply sources can be used, but ensure that they are particle-free or pre-treat the air with a high-efficiency filter.
One major limitation of this aerosolization method is that it requires good flowability of the test powders in order to maintain stable particle generation over a relatively long period. Sticky materials, such as hydrophilic powders with a high moisture content, often stop flowing at an early stage of the aerosolization process and produce very low particle concentrations. Potential ways of solving this issue could include a pre-treatment of the raw powder-such as drying-so as to improve its flowability. The storage condition of the raw materials after uses should be well maintained, e.g., kept in a dry environment and under suitable temperature. During the experiments, higher aerosolization flow rate (0.5-1 L/min) and larger amounts of raw material (e.g., 500 mg) could be used. Additionally, lowering the dilution flow rate can increase particle concentration in the measurement chamber.
Another limitation of this method is the reproducibility of airborne particle generation rate (thus particle number concentration in the measurement chamber). Certain level of variation still exists. Possible ways of improvement are a better defined feeding process to reduce material losses, and well-controlled aerosolization flow rate.
The system and protocols described here could be used for various applications. The use of relatively small amounts of test materials makes the method potentially valuable as an alternative tool for testing powder dustiness. The ranking of levels of airborne particles generated by our system for some common materials was similar to those observed in existing aerosolization systems 19, such as the rotating drum 15,17, continuous drop 23, and vortex shaker methods 24. Furthermore, the adjustable energy input (air flow rate) can also be used for studying the stability of nanoparticle powder agglomerates. Finally, stable aerosol generation can serve as a reliable source of airborne engineered nanoparticles for in vivo or in vitro toxicological studies. The controllable particle concentration would allow an analysis of dose-dependent biological responses. Compared to other aerosolization methods using liquid suspensions, the presented method avoids potential problems such as material suspendability and modification of physical-chemical properties of particles in suspension (e.g., agglomeration, surface properties).
The authors have nothing to disclose.
The authors are grateful for the financial support given to this study by EU FP7 project “Managing Risks of Nanomaterials“ (MARINA) (grant agreement no: 263215).
titanium dioxide nanopowder | JRC | NM-103/104 | Reference materials provided within EU FP7 MARINA project |
silicon dioxide nanopowder | AEROSIL | R974 | |
silicon dioxide nanopowder | JRC | NM-200 | Reference materials provided within EU FP7 MARINA project |
zinc oxide nanopowder | JRC | NM-110/111 | Reference materials provided within EU FP7 MARINA project |
cerium dioxide nanopowder | JRC | NM-211/212 | Reference materials provided within EU FP7 MARINA project |
The V-shaped aerosol generator | Souffleur de verre S.A. | Specially made based on conditions required in the experiments (e.g., geometry, thickness) | |
scanning mobility particler sizer (SMPS) | GRIMM | Model N° 5.403 | Size range: 11.1–1083.3 nm (impactor: d50=1082 nm); composed of a condensation particle counter (CPC) and a dynamic mobility analyzer (DMA); sampling flow ate: 0.3 L/min; sheath flow rate: 3.0 L/min; with standard multiple charge correction and diffusion loss correction; |
optical particle counter (OPC) | GRIMM | Model N° 5.403 | Size range: 0.25-32 µm |
mini-particle sampler (MPS) | ECOMESURE | ||
transport tubes | Milian S.A. | 8 mm-conductive | 6 mm inner diameter |