This manuscript describes the design and operation of a microtensiometer/confocal microscope to do simultaneous measurements of interfacial tension and surface dilatational rheology while visualizing the interfacial morphology. This provides the real-time construction of structure-property relationships of interfaces important in technology and physiology.
Adsorption of surface-active molecules to fluid-fluid interfaces is ubiquitous in nature. Characterizing these interfaces requires measuring surfactant adsorption rates, evaluating equilibrium surface tensions as a function of bulk surfactant concentration, and relating how surface tension changes with changes in the interfacial area following equilibration. Simultaneous visualization of the interface using fluorescence imaging with a high-speed confocal microscope allows the direct evaluation of structure-function relationships. In the capillary pressure microtensiometer (CPM), a hemispherical air bubble is pinned at the end of the capillary in a 1 mL volume liquid reservoir. The capillary pressure across the bubble interface is controlled via a commercial microfluidic flow controller that allows for model-based pressure, bubble curvature, or bubble area control based on the Laplace equation. Compared to previous techniques such as the Langmuir trough and pendant drop, the measurement and control precision and response time are greatly enhanced; capillary pressure variations can be applied and controlled in milliseconds. The dynamic response of the bubble interface is visualized via a second optical lens as the bubble expands and contracts. The bubble contour is fit to a circular profile to determine the bubble curvature radius, R, as well as any deviations from circularity that would invalidate the results. The Laplace equation is used to determine the dynamic surface tension of the interface. Following equilibration, small pressure oscillations can be imposed by the computer-controlled microfluidic pump to oscillate the bubble radius (frequencies of 0.001-100 cycles/min) to determine the dilatational modulus The overall dimensions of the system are sufficiently small that the microtensiometer fits under the lens of a high-speed confocal microscope allowing fluorescently tagged chemical species to be quantitatively tracked with submicron lateral resolution.
Air-water interfaces covered by surfactant films are ubiquitous in daily life. Surfactant-water injections are used to enhance oil recovery from depleted fields and are used as hydraulic fracturing solutions for shale gas and oil. Gas-liquid foams and liquid-liquid emulsions are common to many industrial and scientific processes as lubricants and cleaning agents and are common in food. Surfactants and proteins at interfaces stabilize antibody conformations during packaging, storage, and administration1,2,3,4,5, tear film stability in the eye6,7,8, and pulmonary mechanics9,10,11,12,13,14,15.
The study of surface-active agents or surfactants adsorbing to interfaces and their properties has a long history with many different experimental techniques16,17,18,19,20,21,22,23,24,25,26,27. A recent development is the capillary pressure microtensiometer (CPM), which allows the examination of interfacial properties on highly curved interfaces, at much smaller length scales, while using significantly fewer materials than other common methods9,23,24,25. Confocal fluorescence microscopy (CFM) can be used to study the morphology of lipids and proteins at the air-water interfaces in the CPM22 or on Langmuir troughs20,26,27,28,29. Here a CPM and CFM have been combined to connect morphological phenomena to dynamic and equilibrium interfacial properties to develop structure-function relationships for biological and technological interfaces.
There are numerous parameters of importance in interfacial surfactant systems accessible to the CPM-CFM. In the CPM, a 30-200 µm diameter air bubble is pinned to the tip of a glass capillary tube. In earlier versions of the CPM, the capillary pressure difference between the inside and outside of the bubble was controlled via a water column and oscillatory syringe pump9,30 ; the new version described here replaces these with a higher precision, computer-controlled microfluidic pump. The surface tension (γ) is determined via the Laplace equation, ΔP = 2γ/R, from the pressure drop across the interface set by the pump, ΔP, and optical analysis of the radius of curvature of the bubble, R. The dynamic surface tension of the interface can be determined with 10 ms time resolution following the generation of a new bubble in contact with a bulk liquid containing a soluble surfactant. The surfactant adsorption dynamics can be described by the classic Ward-Tordai equation10,31 to determine essential properties of the surfactant, including the diffusivity, surface coverage, and the relationship between bulk concentration and equilibrium surface tension. Once an equilibrium surface tension is achieved, the interfacial area can be oscillated to measure the dilatational modulus, , by recording the changes in surface tension, induced by small changes in the bubble surface area, A32. For more complex interfaces that develop their own internal structures such as entangled polymers or proteins, the surface tension, , is replaced by a more general surface stress4,33, .
Lung stability during breathing may be directly tied to maintaining both a low surface tension and a high dilatational modulus at the alveolar air-liquid interface9,10. All internal lung surfaces are lined with a continuous, microns-thick film of epithelial lining fluid to maintain tissue hydration34. This epithelial lining fluid is primarily water, with salts and various other proteins, enzymes, sugars, and lung surfactant. As is the case for any curved liquid-vapor interface, a capillary pressure is induced with the pressure higher on the inside of the alveolus (or bubble). However, if the surface tension was constant everywhere within the lungs, the Laplace equation, ΔP = 2γ/R, shows that smaller alveoli would have a higher internal pressure relative to larger alveoli, forcing the gas contents of the smaller alveoli to flow to larger, lower pressure alveoli. This is known as "Laplace Instability"9,35. The net result is that the smallest alveoli would collapse and be filled with liquid and become difficult to re-inflate causing part of the lung to collapse, and other parts would over-inflate, both of which are typical symptoms of acute respiratory distress syndrome (ARDS). However, in a properly functioning lung, the surface tension changes dynamically as the air-epithelial fluid interface in the alveolus interfacial area expands and contracts during breathing. If , or , the Laplace pressure decreases with decreasing radius and increases with increasing radius so as to eliminate the Laplace instability, thereby stabilizing the lung9. Hence, , and how it depends on frequency, monolayer morphology and composition, and alveolar fluid composition may be essential for lung stability. The CPM-CFM has also provided the first demonstrations of the effects of interfacial curvature on surfactant adsorption25, monolayer morphology22 and dilatational modulus9. The small volume (~1 mL) of the reservoir in the CPM allows for the quick introduction, removal, or exchange of the liquid phase and minimizes the required quantity of expensive proteins or surfactants10.
Contrast in a CPM-CFM image is due to the distribution of small fractions of fluorescently tagged lipids or proteins at the interface16,27. Two-dimensional surfactant monolayers often exhibit lateral phase separation as a function of surface tension or surface pressure, π is the difference between the surface tension of a clean fluid-fluid interface, γ0, and a surfactant-covered interface, γ. π can be thought as the 2-D "pressure" caused by the interactions of surfactant molecules at the interface that acts to lower the pure fluid surface tension. At low surface pressures, lipid monolayers are in a liquid-like disorganized state; this is known as the liquid expanded (LE) phase. As the surface pressure increases and the area per lipid molecule decreases, the lipids orient with each other and can undergo a first order phase transition to the long-range ordered liquid condensed (LC) phase16,20,27. The LE and LC phases can coexist at various surface pressures and can be visualized as fluorescently tagged lipids are excluded from the LC phase and segregate to the LE phase. Thus, the LE phase is bright and the LC phase is dark when imaged with CFM16.
The goal of this manuscript is to describe the steps necessary to build and operate the combined confocal microscope microtensiometer. This will allow the reader to perform adsorption studies, measure surface tension, rheological behavior, and examine interfacial morphology simultaneously on a micron-scale air/water or oil/water interface. This includes a discussion of how to pull, cut and hydrophobize the required capillaries, instructions for using pressure, curvature, and surface area control modes, and interfacial transfer of insoluble surfactant to the microtensiometer curved interface.
1. Preparation of capillary tubes
2. Hydrophobization of capillaries
3. Sample preparation and storage
4. Setting up the microtensiometer
5. Adsorption study
6. Oscillation/relaxation study
7. Solvent exchange study
8. Insoluble surfactant adsorption
NOTE: If the surfactant to be adsorbed is not soluble in the reservoir liquid, this method can be used to transfer a monolayer from the air/water interface of the cell to the bubble surface. Many bilayer forming lipids are almost insoluble in saline solution and do not spontaneously absorb to the bubble when suspended in the reservoir solution.
9. Clean up
10. Cleaning the cell
11. Oscillation analysis
A major source of measurement error arises from the capillaries that have defects either from the cutting process (Figure 5A,B) or the coating process (Figure 5D). Both types of defects lead to errors in determining the bubble shape and size by the optical image analysis system, leading to inaccurate surface tension values. It is important to carefully examine each new capillary after it is pulled and coated under the optical microscope before inserting the capillary into the CPM. A miscut capillary must be discarded, but a poorly coated capillary can be acid-cleaned and re-coated to improve the bubble pinning at the end of the capillary (step 2 of the Protocol). Capillaries work best if the end cut is perfectly perpendicular to the capillary (Figure 5C) and the bubble pins directly at the end of the capillary (Figure 5E). The hydrophobic coating on the capillary will become less effective at pinning with use, requiring the capillary to be re-cleaned and re-coated.
Representative data for surfactant adsorption vs. time is presented in Figure 6. Previous experimental techniques such as a pendant or sessile drops used to measure surfactant adsorption did not have a mechanism to dynamically adjust the capillary pressure as the change in surface tension causes the bubble area to change during adsorption30,36,37. In fact, for larger bubbles and drops, changes in the bubble or drop shape (and hence surface area) are required to determine the surface tension from the analysis of the interface shape as the capillary pressure is not measured independently and the capillary pressure varies over the drop or bubble surface37. This also complicates the analysis of the adsorption because as surfactant adsorbs to the interface, the surface tension decreases, and to satisfy the Laplace equation the surface area of the bubble must increase, requiring additional surfactant to adsorb to reach equilibrium. In the CPM, a fixed capillary pressure requires that the initial bubble radius must be within a small range prior to surfactant adsorption to prevent the bubble from ejecting from the capillary if the surface tension decreases too much. Surfactant adsorption dynamics are often modeled by the classic Ward-Tordai equation31, which describes the adsorption of surfactant molecules to a clean interface of constant interfacial area. While the Ward-Tordai equation can be modified to account for the changing surface area, this introduces additional parameters and greatly complicates the analysis38,39.
To overcome these issues, a model-based feedback loop was developed using the Laplace equation that holds the curvature (and surface area) of the bubble constant throughout the adsorption process by dynamically adjusting the capillary pressure. There are significant differences in the rate of change of the surface tension because the area of the bubble is not constantly increasing. The changes in the bubble area during adsorption are not constant with time as the surface tension changes slowly at first, and then rapidly accelerates prior to equilibration. An additional complication is that the fractional change in the area depends on the initial bubble radius. An additional benefit of constant bubble radius is that imaging the interface is simplified as the bubble surface remains fixed, which simplifies focusing the CFM. During the adsorption process, as surfactant adsorbs to the interface (Video 1), the fluorescent signal from the interface increases. If the surfactant forms surface domains, these domains can be observed forming and growing22.
The changes in surface tension during area oscillations are shown in Figure 7. In previous versions of the CPM, oscillations were made in the bubble capillary pressure; however, generating a sine wave in capillary pressure does not translate directly into a sine wave in surface area as the two are related via the Laplace equation. By taking advantage of a model-based feedback loop using the Laplace equation, oscillations are created in the area rather than in capillary pressure, leading to data that is easier to analyze and collect over a larger range of amplitudes. As a result, the surface tension vs. area data collected from this method can be used to directly calculate the interfacial dilatational modulus of the surfactant layer: (Figure 8), where is the total stress of the system and τstress is non-isotropic deviatoric stress often absent in simple surfactant solutions4,33. Thus, for a simple surfactant system, . For interfaces where elastic networks can be formed, such as surface-active proteins, extra stresses are often present and thus must be accounted for when defining the dilatational modulus. Video 2 shows a CFM video of the motion of black LC domains in a continuous colored LE phase matrix in phospholipid monolayers. The distinct LC domains on the interface reorganize into a branching network that covers the interface when oscillations take place on the curved bubble22,40. The Other Area Oscillations tab can be used to create sawtooth, square, and triangular waves as seen in Supplemental Figure 3 and the Compression tab allows for constant rate area compression and expansion.
For solvent exchange studies, a surfactant is first allowed to adsorb to the interface, and then the reservoir liquid is exchanged to allow a second surface-active species to contact that interface. It is possible to examine the change in surface tension as the second surfactant competes with the original surfactant at the interface. The surface dilatational modulus is often a more sensitive probe of the surfactant exchange along with the surface morphology via CFM. Figure 9 shows the change in surface tension, surface dilatational modulus, and surface morphology as one such solvent exchange takes place. While the specifics of such an exchange can vary, a change to any of the three properties could indicate integration of the second component into the monolayer or solvation of the primary component into the bulk. A second fluorescent tag could be attached to the secondary species to observe its interaction with the interface from the CFM images.
Figure 1: Capillary Treatment. (A) Image showing the scoring of the capillary. The glass scoring ceramic is held in a clamp to hold it steady. (B) Acid-cleaning of the capillary. The acid cleaning solution is pulled into the capillary with the vacuum pump. (C) Hydrophobization of the capillary. Silane solution plug held inside the capillary Please click here to view a larger version of this figure.
Figure 2: Cell construction. (1) Large aluminum cell holder, (2) Fluoroelastomer gasket (four in total), (3) glass slide (two in total), (4) PEEK cell, and (5) small aluminum cell holder. When assembled, a fluoroelastomer gasket is placed on either side of each glass slide. The cell is held together with screws and bolts. The zoomed-in image of the PEEK cell shows the locations of the various ports: (6) capillary port, (7) solvent exchange inlet, (8) solvent exchange outlet, and (9,10) temperature control jacket inlet and outlet. A PEEK plug can be used to attach the tubing or capillary to the cell. Ports that are not being used can be completely closed by plugs without channels. Please click here to view a larger version of this figure.
Figure 3: Schematic of CPM/CFM, not to scale. (1) the CPM cell, (2) the capillary tube with a bubble at the tip, (3) confocal microscope objective, (4) microscope camera objective with filter, (5) CPM light source, (6) microfluidic pump, (7) safety valve, (8) fluid exchange inlet, (9) fluid exchange outlet, (10) peristaltic pump, (11) exchange fluid reservoir, (12) fluid exchange waste, (13) direct to cell syringe, (14) temperature control jacket inlet and outlet, and (15) temperature-controlled reservoir and pump. Please click here to view a larger version of this figure.
Figure 4: CPM virtual interface. (1) the file path where the data will be saved; (2) system parameters, comments, and the Save button. All fields in this area are saved to the final data file; (3) the CPM camera image; (4) settings controlling the image analysis, annulus measurement, bubble reset, and frames per second tracking; (5) the Bubble Reset button; (6) the Collect Data button, data recording rate control, and data collection indicators; (7) controls for all operating mode centerline values, oscillation amplitude, and oscillation frequency; (8) operating mode switch: clicking each tab changes to that control mode. Each mode shows the pressure signal being sent to the pump in the "Pressure Signal" graph as well as some additional controls; (9) live surface tension data; (10) live pressure data; (11) live radius of curvature data; (12) live surface area data; and (13) live surface tension and surface area data, which can be used to roughly determine the phase angle during an oscillation study. Please click here to view a larger version of this figure.
Figure 5: Capillary Defects. (A) and (B) Miscut capillaries; (C) correctly cut capillary, (D) capillary with bad pinning due to poor or degraded coating, and (E) properly pinned capillary. The red arrows in D and E indicate where the bubbles are pinned. For the best results, the bubble will pin at the capillary tip. Please click here to view a larger version of this figure.
Figure 6: Adsorption study microtensiometer results for both constant pressure (orange) and constant area (blue) adsorptions. The bubble surface area for the constant area adsorption increases significantly throughout the study and causes the adsorption to take longer to reach the same surface tension. Please click here to view a larger version of this figure.
Figure 7: Typical surface area control oscillation. (A) Pressure, (B) curvature, and (C) surface area data. The surface area data is a sinusoid while the pressure and curvature data are not, as evidenced by the centerline values not being at the midpoint of the oscillation. The mathematical relationship between the three values means only one can be a true sinusoid. Please click here to view a larger version of this figure.
Figure 8: Sample rheological results after analysis. Dilatational modulus of Lyso PC (1-palmitoyl-2-hydroxy-sn-glycero-3-phosphocholine) as a function of frequency for increasing concentrations of Lyso PC for ~45 µm radius bubbles. Concentrations >0.1 mM of Lyso PC that accompany inflammation decrease the dilatational modulus over the range of normal ventilation/ breathing rates (yellow) to make 2ε−γ < 0, which is the crossover value for inducing the Laplace instability (dotted red line). Low concentrations of Lyso PC ≤0.01 mM, that may occur in normal lungs do not induce instability. At frequencies >10 rad/sec, all Lyso PC concentrations are above the crossover, and would not be susceptible to the Laplace instability. Solid red lines are fits of theory to the data. Figure reproduced from reference9. Please click here to view a larger version of this figure.
Figure 9: CFM and CPM results for a solvent exchange study for lung surfactant exchanged with DI water and then Lyso PC. (A) shows how surface tension and surface dilatational modulus change throughout the study. The graph is separated into four regions: when the lung surfactant is adsorbed to the interface (blue), when the LS is exchanged with DI water (green), when the exchange solution is switched to a Lyso PC solution (red), and when the cell is filled with the Lyso PC solution (orange). The properties can be seen to change throughout the various exchanges indicating the interface is changing. (B) shows a confocal image of the lung surfactant adsorbed to the interface before exchange and (C) shows the same surface after the exchange with the Lyso PC solution is complete. In both cases, the white dashed circle indicates the inner edge of the capillary. The structure of the domains on the monolayer changes drastically after the solvent exchange, corroborating the CPM results. Please click here to view a larger version of this figure.
Video 1: Confocal video of constant pressure adsorption study for lung surfactant. The false-color shows the distance in the z-direction with the color bar on the left side of the video, with purple indicating the bubble near the capillary and green being the top of the bubble. The interface is initially dimly illuminated as only a little of the fluorescent surfactant is adsorbed. As more and more surfactant adsorbs, the bubble starts to grow as the color shifts more to green and the interface becomes populated by black LC domains that can move across the interface. Aggregates of surfactant in the solution can be seen floating in the solution as bright amorphous shapes and several settle onto the bubble interface, disintegrating and depositing their surfactant onto the interface. Please click here to download this Video.
Video 2: Confocal video of oscillation study for lung surfactant. The false-color shows the distance in the z-direction with the color bar on the left side of the video. The surface is subjected to several different oscillation frequencies and the dark LC domains on the interface can be seen to change throughout the oscillations. Please click here to download this Video.
Supplementary Figure 1: Example of an intermediate step in the code to determine the dilatational rheology. When this screen appears, the user should left-click at the left-most edge of the oscillation to analyze, and then left-click the right-most edge. Multiple oscillations can be analyzed so that the user could left-click on 1, 2, 3, and 4, and then right-click to analyze those two oscillations. The oscillations shown are of different amplitudes and frequencies. Please click here to download this File.
Supplementary Figure 2: Example of the graphical results produced by the dilatational rheology code. This shows the fits of sinusoids to the oscillations in pressure, radius, surface area, and surface tension as well as the Fourier transform of each oscillation. Ideally, the second harmonic in the Fourier transform should be less than 10% of the first harmonic for the surface area and surface tension. Please click here to download this File.
Supplementary Figure 3: Alternate operation modes. (A) Sine wave, (B) Sawtooth wave, (C) Square wave, (D) Triangular wave, (E) Constant rate expansion, and (F) Constant rate compression. The compression and expansive modes allow for Langmuir type isotherms to be created for insoluble surfactants. Please click here to download this File.
Supplemental Coding File 1: Microtensiometer Virtual Interface.vi. Please click here to download this File.
Supplemental Coding File 2: Dilatational_Rheology_Analysis.m. Please click here to download this File.
The combined CPM/CFM is a powerful tool for examining interfacial dynamics, equilibria, and morphology. This protocol describes the steps necessary for obtaining data with CPM/CFM.
Figure 2 shows the cell design with channels for the capillary, solvent, and heat exchange indicated. The inlet for solvent exchange should be at the bottom of the cell while the outlet should be at the top, allowing for the cell to not overflow during the exchange. In practice, the inlet and outlet flow rates may be slightly different for the same peristaltic pump. One common problem with this cell design is leaking from the cell. This is most often caused by a poor connection between the cell and one of the connections, but if all connections are dry and not leaking, this may be due to a crack in the glass slide of the cell due to over-tightening of the bolts surrounding the cell.
Figure 3 shows the connections between the various pumps and the cell as well as the alignment of the cell with the CFM and CPM objectives. The CPM camera (4) is used to image the bubble shape during operation. The CPM camera must be outfitted with an optical filter that prevents the CFM exciting laser light from entering the CPM camera. Otherwise, the CFM laser makes images in the CPM camera exceedingly noisy and difficult to fit using image analysis. A safety valve connects the capillary and the microfluidic pump (7) and allows changes to be made to the pump and air pressure source, without the risk of backflow from the cell reaching the pump. A second valve (13) provides access to a syringe to allow direct injection of fluid into and out of the reservoir. Fluid may need to be added to the reservoir in the case of a leak and may need to be removed for step 8 of the protocol (insoluble surfactant adsorption) or to remove bubbles purged from the capillary if they have attached to the confocal objective.
During each experiment, several key steps must be performed carefully. Most of the problems that occur once the instrument is running, involve the capillary itself. As such, careful cutting and coating can minimize difficulties. Cutting the capillary to the desired diameter is a difficult and low-yield process. Any chip or unevenness in the tip of the capillary will lead to poor readings of the bubble radius. Additionally, if the hydrophobic coating is not applied correctly, or if it degrades over time and use, the bubble will not pin properly at the tip of the capillary. This can be indicated by the bubble appearing to be pinned inside the capillary or sliding along the inside of the capillary during an oscillatory study. A capillary that is cut well but not pinning properly can be recleaned and hydrophobically treated.
Another key step and possible source of error is cleaning the cell reservoir, tubing, and capillary between different materials or different concentrations of the same material. There are many small crevices in the reservoir and the surfactant can adsorb and alter measurements taken at later times if not cleaned properly. Complete disassembly and soaking of the cell are often required to ensure the removal of any excess surface-active material. It is better to start by using the lowest concentration if a series of concentrations of the same surfactant is to be studied.
At times, lining up the capillary tube with the confocal objective can be difficult. The microtensiometer camera can be used to help align the confocal objective, but for a large working distance of the CFM objective, this may not be helpful. If the confocal microscope is focused beyond the tip of the capillary, the capillary cross-section, a region devoid of any fluorescent material, can also be used to help orient the objective. If the capillary bubble does not eject, there may be a problem with the pressure being supplied to the capillary (which is supposed to be 150 mbar under normal operation). This can be checked by entering pressure control mode and setting the pressure to a high value. If the pressure does not reach the set pressure, there is likely a leak in the tubing from the microfluidic pump or the pump is not receiving sufficient gas pressure. As with many studies involving surface science, it is important to ensure that no contaminating materials are introduced to the solutions at any point. If readings are not as expected (surface tension starting too low or decreasing too quickly), making a new sample, or using a well-studied sample or pure liquid is also a good early step in troubleshooting.
Several modifications can be made to the apparatus to achieve other experimental goals. Oil or water can be added into the capillary allowing for the study of oil-water instead of air-water interfaces39. This increases the risk of backflow into the pump so additional care must be taken, potentially even adding an oil trap to the tubing between the pump and capillary may be necessary.
There are several limitations to the CPM/CFM. The CPM has a limited working range of capillary size, 20-300 µm for the capillary OD for the pump and optics in the system. While it is possible to add insoluble surfactant to the interface using solvent exchange41 or the method described here, the surface concentration can only be inferred from doing surface tension vs. area isotherms and comparing to those obtained from a Langmuir trough. CFM can only detect fluorescent materials, so any non-fluorescent or non-fluorescently tagged materials cannot be visualized. Many surfactants are small molecules, and tagging them can potentially change their properties, though this should be less of a problem for larger surface-active molecules such as proteins or polymers26,27.
This method has several key advantages over previous CPM and CFM analyses of surfactant-laden interfaces. The most important is that the hybrid instrument allows for visualization of the interface while various dynamic and equilibrium surface properties are being measured. Changes in the morphology of the interface can be directly linked to the interfacial dynamics and rheological properties. Previous CFM of surfactant-laden interfaces was done using a flat Langmuir trough16,20,28,29,42,43,44,45,46,47, while the method described here can be performed on highly curved interfaces22. Additionally, the entire interface can be imaged at once, showing a real-time trackable change of specific domains whereas surface flows on the Langmuir trough led to domains flowing in and out of the confocal visual window. Surface compressions on this apparatus are also isotropic, while the barriers on Langmuir troughs have particular compression directions. The CPM allows for much faster area oscillations than would be possible on a Langmuir trough.
The new curvature and area-based control in this study has major advantages over previous versions of the CPM30. Typically, bubble size was controlled by setting a fixed capillary pressure; for dilatational moduli measurements, the capillary pressure was oscillated. When the capillary pressure is held constant, as surfactant adsorbs to the interface, the surface tension of the bubble decreases. To satisfy the Laplace equation, ΔP = 2γ/R, the radius of curvature must decrease as the surface tension decreases. For the hemispherical bubble in the CPM, decreasing the bubble radius of curvature increases the bubble area9,48:
in which Rc is the capillary radius and R is the bubble radius of curvature. The changing radius of the bubble changes the area of the interface during adsorption, which complicates the analysis of the adsorption using the Ward-Tordai equations10,38 Additionally, if the surface tension of the bubble is lowered enough, the bubble radius will become smaller than the capillary radius and the bubble will be ejected. The feedback loop in this newer CPM/CFM keeps the bubble area constant throughout adsorption, meaning that the original Ward-Tordai equation can be used, there is no risk of bubble ejection, and adsorption happens more quickly as the surface is not increasing in the area. For oscillatory studies, producing a sine wave in the pressure does not produce a sine wave in the surface area48. Previous CPM methods relied on keeping oscillations small in order for the area change caused by the pressure-driven oscillation to approximate a sine wave48. The described method directly controls the area of the bubble and can be used to create true sine wave oscillations in the interfacial area. It is possible to directly relate the stress (change in surface tension) to the interfacial strain (change in surface area) to calculate the dilatational modulus.
To help with the implementation of this protocol, a brief description of the code controlling the microtensiometer is described here. The code consists of three segments in a loop: one issuing commands to the microfluidic pump, one controlling the reset mechanism of the bubble, and one measuring the radius of the bubble and saving the calculated values. The pump controller has three main operating modes: pressure control, curvature control, and area control. In pressure control, the user directly inputs a set point for the pressure created by the pump. This mode is important because it does not require a feedback loop, and as such is the most stable of the modes. Curvature control uses the previously measured surface pressure and the Laplace equation to calculate what pressure is required to create an interface of a given curvature. The surface area control mode builds on this by calculating what curvature is required to create a given surface area based on the geometry of the spherical cap, which also requires a precise measurement of the capillary radius. These two modes are especially useful for adsorption and oscillation studies but require a steady stream of consistent surface pressure data. As such, the feed into these two controllers may need to be smoothed from the raw data for better function. When the solution is not clear enough, often due to a highly turbid sample, this mode will not work properly as getting a good image of the bubble interface is not possible. The controls for the oscillation are also included in this section of the code. The middle segment of the code allows for the bubble to be cleared from the capillary. Here, the set pressure of the capillary is set to a high value and held there for a set amount of time allowing the bubble to pop and a new interface to be created. The last section of the code uses vision acquisition software to track the edge of the bubble and measure its radius. This radius is then used with the Laplace equation to calculate the surface tension, which is then fed to the initial part of the loop.
This hybrid CPM/CFM technique has proven to be greatly beneficial to our studies of model and clinical lung surfactants at air-water interfaces. The bubble dimensions approximate those in the alveoli in the human lung and the effects of interfacial curvature on the morphology and dynamics of lung surfactant monolayers can be observed9,10,22. The hybrid instrument will also be important to studies of other surface-active materials that are ubiquitous with applications ranging from petrochemical to household chemicals, from tear films to antibody stabilization. The combined CPM/CFM allows us to probe dynamic interfacial properties on the scale of phase-separated domains and visualize the morphologies on the surface as external conditions change. This method is particularly useful in applications where expensive materials require using minimal size samples. The simultaneous observation of the interfacial dynamics and monolayer morphology is almost impossible with any other technique, making it broadly applicable to the field of interfacial science.
The authors have nothing to disclose.
All the confocal microscopy images were obtained using the Nikon A1RHD Multiphoton upright confocal microscope. We acknowledge the guidance and assistance of the support staff, especially Guillermo Marques, at the University Imaging Center at the University of Minnesota. This work was supported by NIH Grant HL51177. SI was supported by a Ruth L. Kirschstein NRSA Institutional Research Training Grant F32 HL151128.
1.5 O.D. Tygon tubing | Fischer Scientific | Tubing | |
A1RHD Multiphoton upright confocal microscope | Nikon | Confocal Microscope | |
Acid Cleaning Solution | Sulfuric acid and Alnochromix diluted with water 50% by volume, wait until clear befor diluting | ||
Alnochromix | Alconox | 2510 | Mixed with sulfuric acid to package instructionand diluted to make acid cleaning solution |
Ceramic glass cutter | Sutter Instruments | ||
Chloroform | Sigma-Aldrich | 650471 | HPLC Plus |
Curosurf | Chiesi | Lung Surfactant | |
Di Water | 18.5 MΩ – cm | ||
Ethanol | any | 200 proof used for hydrophobization, denatured used for cleaning | |
Fiber-Lite Model 190 fiber optic illuminator | Dolan-Jenner Industries Inc. | 281900100 | Light source; other light sources should work as well |
Flow EZ F69 mbar w/Link Module | Fluigent | LU-FEZ-0069 | Microfluidic Pump |
Fluigent SDK VIs | Fluigent | Required for CPM virtual Interface | |
Fluoroelastomer gaskets | Machined from 1 mm thick Viton sheet, See figure 3 | ||
Gas filter | Norgren | F07-100-A3TG | Put between microfluidic pump and pressure regulator |
Gas regulator | Norgren | 10R0400R | Steps down pressure from sorce to range of pump, connected to gas filter range 2-120 psi |
Glass Capilary | Sutter Instruments | B150-86-10 | Borosilicate glass O.D. 1.5 mm I.D. 0.86 mm |
Glass Slide | any | 75 mm x 25 mm | |
Glass Syringe | Hamilton | 84878 | 25 μL glass syringe |
Hydrophobizing Agent | Sigma-Aldrich | 667420 | 1H,1H,2H,2H-Perfluoro-octyltriethoxysilane 98%, other hydrophobic triethoxysilane can be substituted |
Insoluble surfactant | Avanti | 850355C-200mg | 16:0 DPPC in chloroform |
LabVIEW Software | National Instruments | 2017 | |
Longpass Filter | ThorLabs | FEL0650 | 650 nm Longpass filter, wavelength must remove excitation lazer frequence |
Lyso-PC | Avanti | 855675P | 16:0 Lyso PC 1-palmitoyl-2-hydroxy-sn-glycero-3-phosphocholine |
Masterflex L/S variable speed analog consol pump system w/ Easy-Load II pump head | Masterflex | HV-77916-20 | Peristaltic Pump |
MATLAB | Mathworks | R2019 | |
Micropipette Puller P-1000 | Sutter Instruments | Capillary Puller | |
Microtensiometer Cell and Holder | Cell machined from PEEK, holder machined from aluminum, See Figure 3 and 4 | ||
Microtensiometer Objective | Nikon | Fluor 20x/0.50W DIC M/N2 ∞/0 WD 2.0 mm | |
NI Vision Development Module | National Instruments | Required for CPM virtual Interface | |
PEEK finger tight fittings | IDEX | F-120x | 10-32 Coned Ports |
PEEK plug | IDEX | P-551 | 10-31 Coned Ports |
pippette tips | Eppendorf | 22492225 | 100 μL – 1000 μL, Autoclaved |
Plastic Forceps | Thermo Scientific | 6320-0010 | |
Plastic Syringe | Fischer Scientific | 14-955-459 | 10 mL |
Plumbing parts | Fischer Scientific | 3-way valves and other plumbing parts to connect tubing. | |
Research Plus 1-channel 100 μL–1000 μL | Eppendorf | 3123000063 | Micro pipetter |
Sulfuric Acid | any | Used for acid cleaning solution | |
T Plan SLWD 20x/0.30 OFN25 WD 30 mm | Nikon | Confocal Microscope Objective | |
Texas Red DHPE triethylammonim salt | Thermo Fischer Scientific | 1395MP | Fluorophore |
Vaccum Pump | Gast | DOA-P704-AA |