Optical cells, which are experimental instruments, are small, square tubes sealed on one side. A sample is placed in this tube, and a measurement is performed with a spectroscope. The materials used for optical cells generally include quartz glass or plastic, but expensive quartz glass is reused by removing substances, other than liquids, to be analyzed that adhere to the interior of the container. In such a case, the optical cells are washed with water or ethanol and dried. Then, the next sample is added and measured. Optical cells are dried naturally or with a manual hairdryer. However, drying takes time, which makes it one of the factors that increase the experiment time. In this study, the objective is to drastically reduce the drying time with a dedicated automatic dryer that can dry multiple optical cells at once. To realize this, a circuit was designed for a microcomputer, and the hardware using it was independently designed and manufactured.
Optical cells are used as laboratory instruments in a wide range of fields. In life science research, biomolecules such as nucleic acids and proteins are often utilized for experiments, and spectroscopic methods are widely used for quantitative methods. Accurately quantifying the sample of the experiment is indispensable for obtaining more accurate and reproducible results. The absorption spectrum obtained by a spectrophotometer has often been used for the quantification of biomolecules such as nucleic acids and proteins1,2,3,4. Research on oxidation-reduction characteristics caused by the change in the absorption spectrum and photoluminescence of a carbon nanotube (CNT) dispersed using DNA has also been conducted5,6,7,8,9,10. Optical cells are used for these measurements, but accurate measurements cannot be made unless they are thoroughly washed and dried.
When measuring absorption spectra or photoluminescence, it is impossible to measure precisely in dirty optical cells11,12,13,14,15. Economical disposable optical cells made of polystyrene and poly-methyl-methacrylate are also used to eliminate washing and contamination. However, when precise measurements are required, quartz glasses are often used, because they have extremely excellent optical properties such as light transmittance. In this case, the optical cells are washed after the measurement of the sample and repeatedly used. Usually, after washing optical cells with water or ethanol, they are dried naturally. When rapid drying is required, they are dried one by one by using hair dryers or similar equipment. Cleaning optical cells is one of the most unpleasant and time-consuming procedures in the experiment. As the number of samples increases, the drying time increases, which, in turn, increases the time required to conduct the experiment and research. In past studies, there have been no reports on peripheral devices of optical cells. This study aims to reduce the research time by drying multiple optical cells simultaneously.
We investigated whether other similar products exist. A box-type constant temperature dryer with a temperature control function and a timer function already exists; however, no commercial products with the same configuration can be found.
An outline of the production of this device is described. First, the box-type case is made using an acrylic plate. Nylon netting is attached to the top. A plastic grid is placed on it to fix the optical cell. The control circuit is stored inside the case, and the plastic plate is attached to protect the circuit from water droplets. The control circuit consists of a CPU and is controlled by software. Blowers are attached to the back of the case, and the wind supplied by the blowers enters the optical cells set upside down. The blowers are activated by a switch on the front, and they are automatically stopped by the timer. Depending on the number of optical cells to be dried, two or four blowers can be selected for operation. Water droplets dripping from the optical cells evaporate with the wind from the blowers. The quartz cells are washed with water or ethanol, and the drying time is compared with that of natural drying.
1. Design
2. Hardware Design Outline
3. Software Design Outline
4. Method of Operation
5. Method to Measure the Drying Time
As shown in Table 1, in the case of ethanol washing, the average drying time in natural drying was 426.4 s, and the average drying time in the optical-cell dryer was 106 s. In the case of water washing, the average drying time in natural drying was 1481.4 s, and the average drying time in the optical-cell dryer was 371.6 s. In both cases, the drying time was reduced to approximately one-fourth. The drying time distribution of the optical-cell dryer is shown in Figure 4. The average drying time at 30 locations was 106 s. The number in the upper row represents the position of the cell. The number in the lower row represents the average value of the drying time.
The air volume of the blower was 31 m3/h per one unit, and the total of the four units was 124 m3/h. The air temperature was room temperature, and the temperature control was not performed.
Figure 1: Development drawing. Attach the net on the top of the acrylic case and mount a plastic grid for fixing the optical cells that are attached on top of it. The size of the case is 210 mm in width x 60 mm in height x 104 mm in depth, and 30 optical cells of 12.5 x 12.5 mm can be mounted at the same time. Please click here to view a larger version of this figure.
Figure 2: External view. The material of the casing is acrylic, which is easy to process. The material of the net is nylon. It is fixed to the frame and attached to the upper part of the case. The material of the lattice for the optical cell installation is acrylic, and it is attached to the top of the net. Number description: 1 = the plastic lattice, 2 = the net, 3 = the blower selection button (right side), 4 = the blower selection button (left side), 5 = the blower-operating lamp (right side), 6 = the blower-operating lamp (left side), 7 = the blower start button, 8 = the blower-operating lamp, 9 = the timer, 10 = the power supply switch, 11 = the power supply lamp, 12 = the OLED display, and 13 = the blowers. Please click here to view a larger version of this figure.
Figure 3: Circuit diagram. The power supply is 12 V. The operating voltage of the blowers is 12 V. Control the rotation speed of the blower by pulse width modulation (PWM) operation of the output pin. Connect the blower operation time setting volume to the analog input pin to change the voltage according to the rotational position. Connect the organic light-emitting diode (OLED) for the operation time display to the two digital output pins with an I2C. Please click here to view a larger version of this figure.
Figure 4: Drying time distribution using ethanol. The drying time at 30 locations was measured three times using ethanol to obtain the distribution. The average drying time at 30 locations was 106 s. The number in the upper row represents the position of the cell. The number in the lower row represents the average value of the drying time. Please click here to view a larger version of this figure.
Table 1: Comparison of the drying time for optical cells. The average drying time of natural drying after washing with ethanol was 426.4 s, and the average drying time using the optical-cell dryer was 106 s. The average drying time of natural drying after washing with water was 1481.4 s, and the average drying time using the optical-cell dryer was 371.6 s. Please click here to download this file.
The optical cells can be dried simultaneously with the blowers, and the drying time can be considerably reduced. Even if the stop operation is not executed, it can be safely stopped by using the automatic stop function of the timer. From the measurement results of the drying time distribution, there was no significant difference in drying time because of the difference in the installation position of the optical cells.
A critical step of the protocol is the design of the casing. The challenge is how to make the casing compact. It is also important to figure out how to prevent the excess ethanol or water from dropping into the blower.
To reduce the drying time, the wind volume of the blowers can be increased, but there is a risk that the optical cells might jump out. To raise the blowers' capacity and reduce the drying time, it is necessary to devise measures that prevent this, such as attaching a fixture for fixing the optical cells or attaching a lid to the dryer to put it in a box. There is also a method of increasing the inlet air temperature of the blowers to reduce the drying time. To this end, it is necessary to add a temperature control function so as not to damage the optical cell. However, this is a future task because more complicated devices and control circuits are required.
Another way to reduce the drying time is to vibrate the water droplets to make them drop, but that is also a future research topic.
The authors have nothing to disclose.
The authors have no acknowledgments.
blower | ebm-papst | 422JN | Mulfingen, Germany |
Microcomputer | Atmel Corporation | ATmega 328 P | CA, USA |
Blower selection button | Sengoku Densyo Co., Ltd. | MS-358 (red) | Tokyo, Japan |
Blower operationg lamp | Akizuki Denshi Tsusho Co., Ltd. | DB-15-T-OR | Tokyo, Japan |
Blower start button | Sengoku Densyo Co., Ltd. | MS-350M (white) | Tokyo, Japan |
Timer | Akizuki Denshi Tsusho Co., Ltd. | SH16K4A105L20KC | Tokyo, Japan |
Power supply switch | Marutsuelec Co., Ltd. | 3010-P3C1T1G2C01B02BKBK-EI | Tokyo, Japan |
Power supply lamp | Akizuki Denshi Tsusho Co., Ltd. | DB-15-T-G | Tokyo, Japan |