We describe use of full-field optical coherence microscopy as a method for high quality assessment of corneal donor stroma. This protocol can be used to identify features indicative of health or disease, and is aimed at improving the screening and selection of donor tissues, and hence the results of keratoplasty.
The quality of donor corneal stroma, which makes up about 90% of total corneal thickness, is likely to be one of the main, if not the major, limiting factor(s) for success of deep anterior lamellar and penetrating keratoplasty. These are surgical procedures that involve replacing part or all of the diseased corneal layers, respectively, by donated tissue, the graft, taken from a recently deceased individual. However, means to evaluate stromal quality of corneal grafts in eye banks are limited and lack the capability of high-resolution quantitative assessment of disease indicators. Full-field optical coherence microscopy (FF-OCM), permitting high-resolution 3D imaging of fresh or fixed ex vivo biological tissue samples, is a non-invasive technique well suited for donor cornea assessment. Here we describe a method for the qualitative and quantitative analysis of corneal stroma using FF-OCM. The protocol has been successfully applied to normal donor corneas and pathological corneal buttons, and can be used to identify healthy and pathologic features on both the macroscopic and microscopic level, thereby facilitating the detection of stromal disorders that could compromise the outcome of keratoplasty. By improving the graft quality control, this protocol has the potential to result in better selection (and rejection) of donor tissues and hence decreased graft failure.
Corneal diseases are amongst the major causes of blindness worldwide1. Some diseases can only be treated surgically, often involving the replacement of part (i.e., lamellar keratoplasty) or the entire (i.e., penetrating keratoplasty) diseased cornea, by donated tissue, the graft, taken from a recently deceased individual. For corneal diseases that do not affect the endothelium (e.g., keratoconus, stromal scars after infectious keratitis, trauma, and stromal dystrophies), deep anterior lamellar keratoplasty (DALK) is currently considered as the surgical technique of choice2,3,4,5. This technique makes possible the preservation of the recipient's corneal endothelium, by replacing only the central corneal epithelium and stroma, which is associated with a lower incidence of graft rejection, absence of endothelial rejection, lower endothelial cell loss, and favorable cost-effectiveness ratio6,7,8,9,10,11. DALK further allows corneas with less than optimal endothelium quality to be used as grafts, as this compromised layer will not be transplanted12. Conversely, quality of donor corneal stroma is likely to be the major limiting factor for graft success and vision recovery because the stroma is the only donor corneal layer to remain, while the donor epithelium will be replaced by recipient epithelium. Unfortunately, means to assess donor corneal stroma in eye banks are limited. They usually include slit-lamp examination of the donor eyeball when tissue retrieval is made by enucleation and light microscope examination of the donor stroma13. Some eye banks have started supplementing such standard procedures using Fourier-domain optical coherence tomography (FD-OCT)14.
Ophthalmic optical coherence tomography (OCT), an optical analogue to ultrasound imaging15, uses interference of broadband or tunable light to generate optical sections of the retina16 and anterior segment17. In time-domain OCT, the basis of early clinical systems, the position of a reference mirror is altered, so that interference patterns show up whenever the reference beam has traveled almost the same amount of time as the beam reflected at the various tissue interfaces, with A-scans being generated as a function of time. In FD-OCT (also called spectral- or frequency-domain OCT), the basis of most modern clinical systems, the reference mirror is fixed at one position and an individual A-scan, with all interference patterns mixed together, is acquired at a time, and dissected apart via Fourier analysis.
While clinical (time or spectral-domain) OCT systems permit cross-sectional views of the cornea and the detection of stromal opacities at a higher axial resolution than slit-lamp biomicroscopy, their lateral resolution is limited. Confocal microscopy18 allows examination of the cornea at a lateral resolution approaching histological detail, but is limited axially.
Full-field optical coherence tomographic microscopy (FF-OCT or FF-OCM)19,20 combines elements of both confocal microscopy and OCT, achieving a lateral resolution comparable to the axial resolution of about 1 µm. More specifically, FF-OCM uses incoherent broadband light sources (e.g., a halogen lamp) and high numerical aperture optics to acquire en face 2D tomographic images without lateral scanning. By scanning in the depth direction, FF-OCM enables non-invasive 3D imaging of fresh or fixed ex vivo biological tissue samples. It has been used to image the cornea21,22,23. By providing both high-resolution en face and cross-sectional views, FF-OCM provides information on both the histological structure and cellular details of the cornea. In fact, FF-OCM has been shown to provide structural information superior to histology and was able to identify more disease indicators as was possible with the combination of spectral-domain OCT and confocal microscopy24,25.
Here we describe a protocol for the qualitative and quantitative assessment of corneal donor stroma using FF-OCM. The method is based on the histology-like analysis of macroscopic and microscopic features indicative of stromal condition, including three quantitative stromal parameters (i.e., Bowman's layer thickness and its variability, and stromal reflectivity). The described protocol is therefore applied to normal and abnormal cornea tissues and permits differentiation of diseased from normal human corneal tissues.
All methods described here were carried out according to the tenets of the Declaration of Helsinki and followed international ethical requirements for human tissues. This was a prospective observational case control study. Informed consent was obtained from patients. No modifications to French standards of treatment or follow-up were made. Institutional Review Board (IRB) approval was obtained from the Patient Protection Committee, Ile-de-France V (14944).
1. Tissue Selection and Preparation
2. FF-OCM Initialization, Setup, and Image Acquisition
3. Management of Acquired Images
4. Image Analysis: Qualitatively and Quantitatively Assessment of Stromal Morphology and Features
The FF-OCM device (see Table of Materials)28 and general setup used in this manuscript is shown in Figure 1. Figure 2 shows a swelled human donor cornea after storage in organ culture medium, giving a pathophysiological model of edematous cornea and preventing FF-OCM image acquisition through the entire corneal thickness due to limited penetration depth. Transfer to dextran-supplemented organ culture medium causes deswelling and results in donor corneas of normal thickness, as represented in Figure 3. Diseased corneas can be recognized by morphological changes and typical stromal features, including a decreased and variable thickness of the stroma (Figure 4) and/or Bowman's layer (Figure 5). Assessment of the keratocyte density (Figure 6) and stromal reflectivity (Figure 7) may further aid in the histology-like analysis and differentiation of normal from pathologic corneal tissues with FF-OCM, beyond the capabilities of clinical OCT systems (Figure 8).
Figure 1: Schematic and photographs of the general setup and FF-OCM device used in this work. (A) Photograph of the FF-OCM device (see Table of Materials)28. (B) Schematic and optical setup of the device. (C) Photograph of a human donor cornea in the sample holder. (D) Zoom on the immersion objective and the sample holder. Please click here to view a larger version of this figure.
Figure 2: Organ-cultured normal human cornea before deturgescence. Cross-sectional (A) and en face view (B, slice through the stroma) of swelled ("edematous") cornea caused by storage in organ culture medium, where lakes can be seen as dark areas (as indicated by arrows). Corneal thickness has doubled to over 1,100 µm, preventing image acquisition through the entire depth. Please click here to view a larger version of this figure.
Figure 3: Organ-cultured normal human cornea after deturgescence. Cross-sectional view through the entire cornea (A) and en face stromal view (B) of de-swelled cornea immersed in dextran-supplemented medium, showing regularly distributed hyper-reflective (white) keratocytes. Please click here to view a larger version of this figure.
Figure 4: Overall assessment of stromal morphology and features, including stromal thickness. Compared with normal human cornea (A), corneas with keratoconus (B) feature decreased and variable stromal thickness, along with numerous, parallel Vogt striae that are observable as dark vertical bands in cross-sectional views, and thick stromal nerves that may be found in mid-stromal en face slices (C). Please click here to view a larger version of this figure.
Figure 5: Assessment of Bowman's layer thickness, which is delineated anteriorly by the hyper-reflective epithelial basement membrane and posteriorly by the hyper-reflective keratocytes in the very anterior stroma. Compared with a normal human cornea (A), pathologic cornea (e.g., keratoconus) (B) features decreased and variable Bowman's layer thickness due to interruption and scarring. Please click here to view a larger version of this figure.
Figure 6: Assessment of keratocyte density. (A) Keratocyte nuclei are manually counted on enhanced and inverted en face images after selection of a 300 µm x 300 µm region of interest. (B) Counted nuclei are indicated by arrows. Please click here to view a larger version of this figure.
Figure 7: Assessment of stromal reflectivity. The amount of backscattered light decreases (exponentially) with depth in the stroma of normal donor corneas (see Figure 3 and Figure 4A), resulting in linear logarithmic depth profiles, represented by R-square values close to 1 (green trace). This may not hold true for pathological corneas featuring stromal regions of increased light backscatter (see Figure 4B). The presence of such macroscopic features, as in the example of a cornea with stromal scar after infectious keratitis (depicted in the inset), will thus create non-linear logarithmic intensity depth profiles, represented by an R-square value below a certain threshold (e.g., lower than 0.9425; red trace). Note that logarithmic amplitude depth profiles are in fact shown as FF-OCM measures the amplitude of light backscattered by the sample rather than its intensity. Please click here to view a larger version of this figure.
Figure 8: Demonstration of benefits of FF-OCM over histology and spectral-domain OCT. (A) Histology and (B) corresponding spectral-domain-OCT image acquired on the same cornea in vivo prior to keratoplasty. (C) Corresponding FF-OCM cross section of the ex vivo corneal button post-keratoplasty, illustrating superior resolution compared to the clinical spectral-domain OCT images. The fibrosis is also more clearly visible in FF-OCM images than in the histology. The high density of keratocytes in the upper stroma can clearly be seen in both the (C) cross-sectional and (D) en face views. Please click here to view a larger version of this figure.
The protocol described here for the qualitative and quantitative assessment of corneal donor stroma using FF-OCM is based on the histology-like analysis of macroscopic and microscopic features indicative of stromal condition, beyond the capabilities of spectral-domain OCT and confocal microscopy21,24,25, and enables differentiation of diseased from normal human tissues.
Aside from an excellent endothelial quality assessment of human donor corneas by means of specular microscopy, assessment of stromal quality is challenging in eye banks, and generally limited to a gross observation with slit-lamp biomicroscopy and/or light microscopy in current protocols. Lack of fine resolution with existing methods not only means that corneas with some stromal disease may be selected that compromise the result of keratoplasty, but also that corneas may be rejected for stromal opacities that are in fact constraint to anterior stroma or epithelial regions and could still be used for endothelial keratoplasty procedures14.
The current eye-bank protocol could be supplemented by the addition of FF-OCM, which due to its superior resolution, constitutes a powerful and non-invasive tool to complete the quality assessment of the cornea, especially the stroma (including Bowman's layer). Unlike during slit-lamp examination, the graft remains immersed in a closed chamber filled with storage medium throughout FF-OCM image acquisition, decreasing any potential risk of contamination.
For successful image acquisition with FF-OCM (see Table of Materials), it is important for the microscope objective to be well immersed in the optical gel that is applied on top of the coverslip of the sample holder (step 2.2.3). It is further recommended to regularly check the calibration of the device, a procedure also to be performed after unsuccessful auto-adjustment (step 2.2.2) and accessed via "Tools and options" in the acquisition software (see Table of Materials). The procedure, which involves the utilization of a calibration mirror in the sample holder, is the same as the usual sample preparation (see step 1.2) except that the optical gel should be applied on the mirror before positioning of the coverslip.
A series of donor corneal grafts, considered to have normal stroma as per existing eye-bank procedures, were used to describe the protocol in this manuscript and specifically demonstrate the suitability of FF-OCM for precise and reliable assessment of donor stromal quality. These normal donor corneas were compared with pathological corneas immersed in storage medium, showing that the histology-like analysis made possible with FF-OCM of several stromal features (illustrated in Figure 2, Figure 3, Figure 4, Figure 5, Figure 6, Figure 7, and Figure 8) in corneal grafts allows distinguishing diseased from normal human corneal tissues.
Aside from morphological changes, such as the presence of scars (Figure 5 and Figure 7), fibrotic tissue (Figure 8), lakes (Figure 2), Vogt striae (Figure 4), or increased stromal nerve diameter (Figure 4), typical stromal features are present in diseased corneas. Stromal parameters particularly relevant in the stromal quality assessment appear to be Bowman's layer thickness and its variability, and stromal reflectivity. Critical steps within the protocol are thus steps 4.1 and 4.3.
While being secreted during human corneal development, Bowman's layer, in particular, becomes distinct by 19 weeks of gestation and never repairs after birth32. Damage to Bowman's layer is thus irreversible and serves as an ideal indicator of previous stromal damage in donor corneal tissue, including damage caused by refractive surgery, infectious keratitis, keratoconus. Such corneal diseases, which constitute contraindications for donor corneal usage, are associated with decreased and variable Bowman's layer thickness due to interruption and scarring (Figure 5), and are likely to be missed by current eye-bank protocols when the donor history is not precisely known.
Although corneal transparency is impaired after donor death due to post mortem corneal edema, the amount of backscattered light, or stromal reflectivity is expected to decrease exponentially with depth in the stroma (see Figure 3 and Figure 4A); as a result, the logarithm of normalized stromal reflectivity will be a linear function of stromal depth in normal donor corneas, represented by R-square values close to 1. Conversely, the presence of macroscopic features is associated with non-linear logarithmic depth profiles and indicative of stromal disease (Figure 4B and Figure 7)25.
Since keratocyte density is responsible for stromal collagen fibril and extracellular matrix synthesis and renewal, it appears reasonable to assume that keratocyte density is another relevant parameter for assessing donor stromal quality, and that tissues exhibiting very low keratocyte counts should not be transplanted. The protocol therefore includes a precise and reliable method to measure keratocyte density from FF-OCM images that can be easily used in eye banks25 and follows the convention of confocal microscopy. Note that with FF-OCM, keratocyte density may also be determined by counting keratocytes directly in the cross-sectional view33, a potential advantage over confocal microscopy, which requires keratocytes to be counted on multiple en face slices. However, unlike in living patients, where keratocyte densities have been demonstrated to be lower in disease patients than in normal controls34,35,36,37 and to correlate with disease severity34,38, this was not the case in human ex vivo tissue samples25, and further studies are necessary to determine whether a minimal number of keratocytes is required in donor corneas to result in good visual recovery after transplantation. Low keratocyte density in donor tissue as in pathological tissue could be explained by aging, post mortem loss of cells induced by ischemia, and/or storage of donor tissue27,39,40,41. It should also be pointed out that the normal donor corneas that were obtained and imaged in this protocol were either stored and edematous or de-swelled, or had been discarded by the eye bank before transplantation because of poor endothelial quality according to the standards of the EU Eye Bank Association. Were FF-OCM imaging along with the described protocol to be included in the eye bank setting, the corneas would typically be assessed in a fresher state than was possible here, which may affect the keratocyte densities.
The protocol described here for the stromal quality analysis could be extended for assessment of Descemet's membrane, which can also be resolved with FF-OCM in terms of thickness and structure21,24. This may prove useful for selection of tissues for Descemet's membrane endothelial keratoplasty, where thin Descemet's membranes may be more difficult to separate from the stroma.
In conclusion, FF-OCM enables precise and reliable assessment of human donor corneal stroma during storage. By improving the graft quality, the addition of this protocol to current eye-banking procedures has the potential to improve the screening and selecting of donor tissues, and hence the results of keratoplasty. Real-life integration of the FF-OCM device into the eye-bank routine should be facilitated by recent technological updates, including faster image acquisition and larger field of view thanks to the development of a custom CMOS camera, and the design of custom sterile disposable cassettes for cornea storage and handling during imaging.
The authors have nothing to disclose.
This work has received funding from the Agence Nationale de Recherche (ANR), under a PRTS (Projet de Recherche Translationelle en Santé) grant No ANR-13-PRTS-0009 (V.B.) and from the European Union's Horizon 2020 research and innovation programme under the Marie Skłodowska-Curie grant agreement No 709104 (K.I.). The authors thank Céline de Sousa for help with cell counting and histological processing.
Light-CT Scanner | LLTech, France | http://www.lltechimaging.com/products-applications/products/ | FF-OCM device used in this manuscript for imaging |
CorneaJet | EuroBio, France | http://www.eurobio-cornea.com/en/corneamax-10-100-ml-xml-352-822.html | Organ culture medium in which donor corneas are stored |
CorneaMax | EuroBio, France | http://www.eurobio-cornea.com/en/corneajet-10-50-ml-xml-352-823.html | Dextran-supplemented organ culture medium used for deturgescence |
Fiji (ImageJ) | National Institute of Health, Bethesda, MD, USA | https://fiji.sc/ | Open source image processing software |
Matlab | Mathworks, Inc., Natick, MA, USA | https://www.mathworks.com/products/matlab.html | Mathematical computing software |