A multidimensional gas chromatography method for the analysis of dissolved hydrogen sulfide in liquid crude oil samples is presented. A Deans switch is used to heart-cut light sulfur gases for separation on a secondary column and detection on a sulfur chemiluminescence detector.
A method for the analysis of dissolved hydrogen sulfide in crude oil samples is demonstrated using gas chromatography. In order to effectively eliminate interferences, a two dimensional column configuration is used, with a Deans switch employed to transfer hydrogen sulfide from the first to the second column (heart-cutting). Liquid crude samples are first separated on a dimethylpolysiloxane column, and light gases are heart-cut and further separated on a bonded porous layer open tubular (PLOT) column that is able to separate hydrogen sulfide from other light sulfur species. Hydrogen sulfide is then detected with a sulfur chemiluminescence detector, adding an additional layer of selectivity. Following separation and detection of hydrogen sulfide, the system is backflushed to remove the high-boiling hydrocarbons present in the crude samples and to preserve chromatographic integrity. Dissolved hydrogen sulfide has been quantified in liquid samples from 1.1 to 500 ppm, demonstrating wide applicability to a range of samples. The method has also been successfully applied for the analysis of gas samples from crude oil headspace and process gas bags, with measurement from 0.7 to 9,700 ppm hydrogen sulfide.
Accurate analysis of crude oil is essential for the oil and gas industry, as health and safety regulations and economics are functions of oil quality. In order to protect transporters of crude samples, it is necessary to determine the properties of crude samples to develop safety regulations to be implemented in the event of a release or spill. In particular, quantification of hydrogen sulfide (H2S) is important, due to its high toxicity in the gas phase; exposures as low as 100 ppm can be fatal (http://www.cdc.gov/niosh/idlh/7783064.html)1,2. Dissolved H2S in crude samples is generally considered to be corrosive3,4, and can deactivate catalysts used to treat the oil5-7. Removal of H2S from crude oil streams is ideal, but without a method to measure dissolved H2S, it is difficult to assess the success of removal treatments. For these reasons, this protocol was developed to measure dissolved H2S in heavy crude oil samples such as Canadian oil sands crudes.
A number of standard methods exist for quantification of H2S in lighter petroleum or fuel based samples, but none have been validated for use with the heavier crudes commonly extracted from the Canadian oil sands. H2S and mercaptans are determined using a titration technique by Universal Oil Products (UOP) method 1638, but this method suffers from user-interpretation bias that results from manual reading of titration curves. Institute of Petroleum (IP) method 570 uses a specialty H2S analyzer that heats fuel oil samples9, and benefits from simplicity and portability, but lacks accuracy with heavier samples10. The American Society for Testing and Materials (ASTM) method D5623 uses gas chromatography (GC) with cryogenic cooling and sulfur selective detection to measure H2S in light petroleum liquids11,12. This standard could be improved to use an ambient separation and also be applied to heavier crude oils, therefore it was used as the basis for the protocol discussed herein.
GC is a heavily used technique for the analysis of petroleum samples. Samples are vaporized in a hot inlet, and separations occur in the gas phase. The gas phase separation makes GC ideal for the analysis of H2S, as it is easily liberated from the liquid sample during heating in the inlet. GC methods can be created and tailored for different samples, depending on the temperature programs used, columns implemented, and the use of multidimensional chromatography13-15. There have been a number of recent developments for the measurement of H2S using GC. Luong et al. demonstrated H2S and other light sulfur compound measurement in light and middle distillates using multidimensional GC and Deans switching, but the method has not yet been applied to heavier crudes16. Di Sanzo et al. also quantified H2S in gasoline using GC, however it also has not been used on heavier crudes, and requires sub-ambient cooling17. The method presented here demonstrates considerable time saving over these previous methods, with a completed analysis time of 5 min, compared to 10 min (Luong) and 40 min (Di Sanzo). Unfortunately, implementation of these methods in our lab to compare accuracy was not possible due to equipment and time restrictions.
Multidimensional GC allows the user to exploit the selectivity of two columns, rather than a single column. In conventional GC, separation occurs on one column. In the case of multidimensional GC, the sample is separated on two different columns, enhancing the separation and selectivity. The Deans switch is one device used to employ a two-dimensional column configuration. The switch uses an external valve to direct gas flow from an inlet on the switch to one of two outlet ports18-20. Effluent from the first column can be directed in either direction; in this case, light sulfur gases are “heart cut”21 from the first separation to a porous layer open tubular (PLOT) column for secondary separation, which has been shown to be excellent for the separation of H2S from other light sulfur gases (http://www.chem.agilent.com/cag/cabu/pdf/gaspro.pdf)22-24. A sulfur chemiluminescence detector is used for detection, providing selectivity for sulfur compounds and eliminating possible interference from any other light gases that may have been transferred to the PLOT column during the heart cut. Hydrocarbons from the crude oil sample are retained on the first dimension column and are removed during a backflush procedure; this protects the PLOT column from any contamination25-27. This approach has also been successfully implemented for the analysis of oxidation inhibitors in transformer oils28.
Herein, a two-dimensional GC method is employed for the analysis and quantification of dissolved H2S in heavy crude oil samples. The method is shown to be applicable over a wide range of H2S concentrations, and can also be used to measure H2S in gas phase samples.
Caution: Please consult all relevant material safety data sheets (MSDS) for materials before using. In particular, CS2 is highly flammable and should be stored and handled appropriately. H2S gas is highly toxic, and any containers or gas bags containing H2S should not be opened or handled outside of a properly vented fumehood. Work with crude oil samples should only be done with full personal protective equipment (gloves, safety glasses, lab coat, pants and closed-toe shoes), and all crude samples should be opened, transferred and handled in a fumehood. Certified gas standards will be delivered from the manufacturer with an expiry date, and for the most accurate results care should be taken to use standards that have not expired.
1. Preparation of Standards
2. Instrument Set-up
Figure 1. Deans switch calculator. Screenshot of the Deans switch calculator program. User-adjustable parameters are shown in white boxes, and output parameters are shown in blue boxes. Please click here to view a larger version of this figure.
Figure 2. Gas chromatograph oven. Configuration of the column arrangement in the GC oven. FID: flame ionization detector, SCD: sulfur chemiluminescence detector. Please click here to view a larger version of this figure.
3. Instrument Calibration
4. Sample Analysis
Figure 3. A gas chromatogram with an overloaded H2S peak. A gas injection from the headspace of a liquid crude sample held at 30 °C, demonstrating an overloading of the SCD. Please click here to view a larger version of this figure.
5. Data Analysis
Figure 4. Crude sample spiked with H 2 S. Two overlaid chromatograms illustrating the change expected when spiking a crude sample with H2S. Please click here to view a larger version of this figure.
Figure 5. Gas chromatogram data analysis. A screenshot of a data analysis program highlighting the location of an H2S peak in a sample and the peak area to be used to determine the concentration of H2S. Please click here to view a larger version of this figure. Please click here to view a larger version of this figure.
Figure 6. Sample spreadsheet for data analysis. A screenshot of a spreadsheet program showing an example of how to calculate the concentration of H2S using the area of the calibration standard peak and the area of the sample peak. Please click here to view a larger version of this figure.
In order to obtain reliable quantification of H2S for both liquid and gas samples, proper calibration is necessary. For calibration injections and sample injections, the H2S peak should not be overlapping with neighboring peaks and should have a reproducible peak area. Figure 3 shows an injection of a gas sample where the gas is too concentrated for this method. It was found that gas concentrations of greater than 500 ppm using a 250 µl syringe overloaded the detector. This issue was not encountered for liquid samples, as gas phase concentrations of H2S were generally much higher than in the liquid. The overloading issue was addressed by injecting a smaller volume of gas. It was found that adjusting other parameters such as split ratio degraded the chromatographic performance, whereas smaller injection volumes were the most reproducible. For both liquid and gas injections the first injection often had a different peak area than the three subsequent injections, and was regularly discarded. The SCD was also calibrated at the beginning of each day of analysis.
Figures 7 and 8 illustrate typical chromatograms achieved using this method. The H2S peak is close to, but does not coelute with, neighboring peaks. Other peaks in the chromatograms were not identified, as the focus of the protocol was H2S. Proper timing and balancing of the Deans switch is essential for achieving and maintaining good separation and chromatography of H2S. An improperly timed switch will be indicated by small, variable peak areas, or intermittent loss of peaks. If pressures are not balanced properly, the H2S gas will be split between both detectors, or will not be heart cut properly to the PLOT column, resulting in an absence of peaks. Backflushing occurs after the separation, and should not interfere with H2S measurement. Regular blank injections of toluene should indicate no carryover or system contamination.
Figure 7. Representative liquid crude chromatogram. A chromatogram of a liquid crude sample that contains 26.3 ppm of dissolved H2S. The H2S peak is identified with an arrow. Please click here to view a larger version of this figure.
Figure 8. Representative gas chromatogram. A chromatogram of a gas sample taken from the headspace of a liquid crude sample held at 30 °C. The arrow identifies the H2S peak; this gas sample contains 9.03 ppm of H2S. Please click here to view a larger version of this figure.
Carrier gas | H2 | ||
Oven | |||
Oven program | 50 °C for 2 min, then 100 °C/min to 250 °C for 1 min | ||
Run time | 5 min | ||
Post run* | 250 °C for 16 min | ||
Split-Splitless Inlet | |||
Liner | Deactivated glass wool | ||
Mode | Split | ||
Temperature | 250 °C | ||
Pressure | 40 psi | ||
Total flow§ | 30.778 ml/min | ||
Septum purge flow | 1 ml/min | ||
Split ratio# | 10:1 | ||
HP-PONA Column | |||
Initial pressure | 40 psi | ||
Flow | 2.7071 ml/min | ||
Pressure program | 40 psi for 5 min | ||
Post run * | 1 psi for 16 min | ||
GasPro column | |||
Initial pressure | 6.89 psi | ||
Flow | 2.9859 ml/min | ||
Pressure program | 6.89 psi for 5 min | ||
Post run* | 39.405 psi for 16 min | ||
Fused silica transfer line | |||
Initial pressure | 6.89 psi | ||
Flow | 5.1837 ml/min | ||
Pressure program | 6.89 psi for 5 min | ||
Post run* | 39.405 psi for 16 min | ||
FID | |||
Temperature | 250 °C | ||
H2 Flow | 40 ml/min | ||
Air flow | 450 ml/min | ||
Makeup flow | 20 ml/min | ||
Deans switch | |||
Off | 0.7 min | ||
On | 2.3 min | ||
Liquid autosampler* | |||
Syringe size | 10 µl | ||
Injection volume | 1 µl | ||
Pre-injection washes | 1 | ||
Post-injection washes | 2 | ||
Wash volume/sample wash volume | 8 µl | ||
Sample washes | 2 | ||
Sample pumps | 6 | ||
Solvent/sample wash draw speed | 300 µl/min | ||
Solvent/sample wash dispense speed | 6,000 µl/min | ||
Injection dispense speed | 6,000 µl/min | ||
Viscosity delay | 6 sec | ||
* Omitted for gas analysis | |||
§ 111.99 ml/min for gas analysis | |||
# 40:1 for gas analysis |
Table 1. Gas chromatograph method parameters for both liquid and gas analysis.
In order to achieve optimum measurement of H2S, this method employs a Deans switch, backflushing and a sulfur chemiluminescence detector (SCD). A dimethylpolysiloxane column is used as the first dimension GC column, and serves to retard the movement of heavier hydrocarbons present in the sample so that they do not contaminate the PLOT column. This effect is enhanced by a cool (50 °C) initial separation. Light gases pass through the first dimension column and are captured by the PLOT column during the heart-cut for further separation. The SCD only responds to compounds containing sulfur, adding an additional layer of selectivity, and preventing interference by any hydrocarbons or other light gases29,30. The column configuration used in this method is shown in Figure 2. The use of the PLOT column makes backflushing essential when injecting liquid crude samples. During the backflush, the columns are heated and gas flow is reversed out the inlet, removing hydrocarbons from the column and preventing their transfer to the PLOT column during subsequent injections25-27. The process of backflushing will result in a buildup of material in the inlet liner of the GC, and the liner will require cleaning and/or replacement approximately every 50 injections. Regular blank injections indicated that no sample carryover occurred between injections, and monitoring of chromatographic performance showed that hydrocarbon contamination was not an issue for the PLOT column. The limits of detection and quantification for this method were calculated using the signal/noise relationship of blank samples31. For gas samples, the limits of detection and quantification were calculated to be 0.2 ppm and 0.6 ppm, and 0.5 ppm and 1.6 ppm for liquid samples, respectively. The liquid values are comparable to the limits of quantification listed for standard methods ASTM D562311 and UOP 1638 (1.0 ppm), and somewhat greater than IP 5709 (0.5 ppm).
H2S is a light gas that will easily escape to the ambient air. When working with gas bags, they need to be monitored for leaks, and emptied and refilled when the area of the calibration peaks begins to change between day-to-day analyses. For this same reason, vials of crude oil for analysis were prepared on the day of, and not reused for a second day to mitigate evaporative losses. Obtaining the lowest relative standard deviation (%RSD) for manual injection also depends on user technique. Consistent practice using a gas tight syringe to manually inject samples improved %RSD for samples to consistently achieve <10% variation for samples, and <5% variation for standard calibration. Retention time variation was less than 1% for manual injection. When generating response factors for quantitation, a new response factor should be calculated on each day of analysis. While this limits the number of analyses that can be completed in a day, it was found to be optimal for the best accuracy, as instrument response varied by up to 10% over extended periods of use. Liquid samples that are diluted may require optimization; in our sample set, a 1:1 dilution with toluene was sufficient to preserve the H2S, but any greater dilution resulted in a loss of the H2S peak. The CS2 stock solution used for liquid calibration was stored at ambient temperature in a flammable storage cabinet, and was found to produce a consistent response over 6 months of use. The use of CS2 as a calibration standard is possible because the SCD provides a uniform response toward sulfur, and any stable sulfur-containing compound can be used.
Programming and balancing the Deans switch can present a challenge. The use of available software for determining inlet and PCM pressures greatly reduces the time required to implement switching (Figure 1). Prior to optimizing the heart-cut window, it was useful to inject the gaseous H2S calibration standard directly through the columns with no heart-cutting. This gave a baseline to which performance could be compared, and the H2S peak area after heart-cut optimization was compared to the peak area without heart cutting to ensure the peak was fully captured. This process should be done with a pure gas standard, and not with a spiked liquid crude, as contamination of the PLOT column with hydrocarbons will degrade chromatographic performance24. The system can also be modified from that recommended in this study. Other hydrocarbon columns have been successfully used in place of the 100% polydimethylsiloxane column, and helium as a carrier gas has been implemented as well. It is also possible to install short (<60 cm) fused silica connectors between the columns and the detectors if so desired; using 0.250 mm inner diameter fused silica reduces any additional backpressure, and does not require modification of the method.
The method described herein demonstrates the applicability of Deans switching for the analysis of targeted compounds in heavy crude oil. It is expected that the principle of this experiment could be applied to the analysis of other light gases present in crude oil, especially when the use of a selective detector is practical. To the best of our knowledge, this method is the only available technique that is capable of accurately measuring dissolved H2S in heavy crudes, and that does not employ the use of sub-ambient cooling. Samples ranging in density from 0.74 to 0.94 g/ml were analyzed without difficulty. Dissolved H2S was successfully quantified from 1.1 – 500 ppm in liquid samples, and gas phase H2S was quantified from 0.7 – 9,700 ppm. It is hoped that this work will serve as an excellent complement to previously established methods whose focus is on lighter crude oil streams and fuels.
The authors have nothing to disclose.
The authors would like to acknowledge support from the Government of Canada’s interdepartmental Program of Energy Research and Development, PERD 113, Petroleum Conversion for Cleaner Air. N.E.H would like to acknowledge her Natural Sciences and Engineering Research Council of Canada Visiting Fellowship.
Deans switch | Agilent | G2855A | Or equivalent flow switching device |
Restrictor tubing | Agilent | 160-2615-10 | Fused silica, deactivated, 180 µm |
HP-PONA column | Agilent | 19091S-001 | |
GasPro column | Agilent | 113-4332 | |
Sulfur chemiluminescence detector, 355 | Agilent/Sievers | G6603A | |
H2S calibration standard, in He | Air Liquide | Custom order | 211 ppm H2S |
CS2 | Fisher Scientific | C184-500 | |
Toluene, HPLC grade | Fisher Scientific | T290-4 | |
Gas bag, 2 L | Calibrated Instruments, Inc. | GSB-P/2 | Twist on/off nozzle |
250 µL gas tight syringe | Hamilton | 81130 | |
500 mL amber glass bottle | Scientific Specialties | N73616 | |
Open top screw caps | Scientific Specialties | 169628 | |
Tegrabond disc for screw caps | Chromatographic Specialties | C889125C | 25 mm, 10/90 MIL |
1 mL gas tight syringe | Hamilton | 81330 | |
2.5% H2S in He gas standard | Air Liquide | Custom order |