This video describes high-performance fluorescence anisotropy that helps to monitor the interaction between transcription factors and DNA. The assay monitors the interactions of a fluorophore-labeled DNA molecule with its specific transcription factor by measuring the degree of polarization due to molecular rotation or anisotropy of the fluorophore-labeled DNA.
Protocol
1. Polarization Microscopy For widefield laser illumination, focus a 638 nm line of a continuous diode laser (40 mW) on the aperture of multimode optical fiber for beam cleanup. Mount a linear polarizer at the output of the fiber to set the polarization of laser light. Block the excitation component of the emitted light with a dichroic mirror (640 nm cut-off) and a bandpass filter (bandpass 700/75). Let the fluorescence signal pass through a polarizing beam splitter, whic…
Representative Results
Figure 1: Schematic depictions of the HIP-FA assay and experimental setup. (a) Gel delivery system for titrating competitor DNA in single wells. (b) HIP-FA microscopy setup.Customized automated widefield microscope with polarized fluorescence light detection on an EM-CCD camera. (c) Raw fluorescence image with the two regions of interest used to determine the parallel (red) and the perpendic…
Offenlegungen
The authors have nothing to disclose.
Materials
Cy5-labled 16- / 18-bp DNA-oligomers
Eurofins
Custom synthesis
16- / 18-bp DNA-oligomers
Eurofins
Custom synthesis
Nile Blue A
Sigma
N5632-25G
Sensoplate plus microplate 96- or 384-well, PS
Greiner
655891
384-Well Sensoplate, black
Greiner
788896
Agarose, low gelling temperature
Sigma
A9414-50G
Sodium Chloride
Merck
1.06404.1000
Tween-20
Sigma
P1379-1L
Di-Potassium hydrogen phosphate trihydrate
Merck
1.05099.1000
Potassium dihydrogen phosphate
Merck
1.04873.1000
Q-POD Element
Merck Millipore
ZMQSP0DE1
Millipak 40 Gamma Gold Filter
Merck Millipore
MPGL04GK2
Milli-Q Integral 3 Water Purification System
Merck Millipore
ZRXQ003WW
Quantum TIX
Merck Millipore
QTUMOTIX1
DL-Dithiothreitol
Sigma
43815-1G
Mastercycler gradient
Eppendorf
Z316083
SafeSeal tube 1.5 mL
Sarstedt
72.706.200
Tube 15 mL
Sarstedt
62.554.502
Multiply-Pro cup 0.2 mL PP
Sarstedt
72.737.002
MICROSCOPY SETUP:
Automated widefield microscope
LEICA
DMI6000
Long distance objective
LEICA
HCX PL FLUOAR L 60x/0.60 N.A. Dry
638 nm line continuous diode laser
Omicron
PHOxX 638-40, 40mW
Back-illuminated EM-CCD Camera
Andor
iXon DV897
Dichroic mirror
AHF
640nm cut-off
Bandpass filter
AHF
ET bandpass 700/75
Linear polarizer
Thorlabs
LPVISC050-MP2
Polarizing beam splitter
Thorlabs
BS010
Achromatic lens
Thorlabs
200 mm focal length
Multimode optical fiber
Optronis
FVP600660710
ROBOTIC SYSTEM:
Our robotic system includes a Biomek NXP workstation with a 96-channel head and Span-8 pipettors, connected with a servo-shuttle, which are used for all liquid transfer steps. In addition, the system is equipped with orbital shakers and a microplate reader (Paradigm, Molecular device) served by the Span-8 gripper