High-precision micro-displacement measurement is significant in the field of aerospace engineering, ultra-precision machining, and micro-assembly. The present protocol describes measuring micro displacements based on the shadow technique.
The precision measurement of micro displacement is important in scientific and industrial fields. However, it is a tough challenge due to the complex design and the high cost of measuring instruments. Inspired by the shadow formed from water striders walking on a water surface under sunlight, a micro-displacement measurement method was proposed. Water strider legs with superhydrophobic properties bend the water surface. The curved surface of the water refracts sunlight, creating a shadow with a bright edge at the bottom of the pond. The shadow size is generally larger than the indentation depth of the legs from the water surface. In the micro-displacement measurement system, the applied displacement is proportional to the change in the diameter of the shadow. The presented study proposes a micro-displacement measurement procedure based on this shadow technique. The displacement sensitivity can reach 10.0 nm/pixel in the range of 5 µm. This system is simple to construct, low cost, and has high precision with good linear performance. The method provides a convenient additional option to measure micro-displacement.
Precise displacement measurements play a vital role in the fields of aerospace engineering1, ultra-precision machining2, and micro-assembly3. Structural deformation must be measured precisely for structural health monitoring4. However, micro-displacement measurements with high precision remain a tough challenge due to the complex design and high cost of the measuring instruments5.
The micro-displacement measurement technique can be divided into conventional and non-conventional methods. Conventional methods, such as magnetic, capacitive, inductive, and electric sensors, are susceptible to electromagnetic interference6. Non-conventional methods are mainly optical methods, such as the optical fiber-based method and the laser method.
Ke Tian et al. designed a balloon-shaped bent multimode fiber structure to measure micro displacement, whose displacement sensitivity could achieve 0.51 dB/µm with a measuring range of 0-100 µm experimentally7. However, the size and the cost of the fiber optic demodulator must be considered first; and it was not easy to eliminate the thermal effect. Qianbo Lu et al. proposed a sub-nanometer resolution displacement sensor based on a grating interferometric cavity, whose sensitivity could reach 44.75 mV/nm by the intensity compensation and phase modulation8. The laser interferometer is one of the commonly used micro-displacement instruments with nanoscale resolution. However, the reflector requires complicated signal processing, and the fringe resolution of the interferometry limits its applications9. Therefore, an alternative simply-constructed, low-cost, high-precision measurement system is needed.
This article proposes a micro-displacement measurement procedure based on the shadow technique10,11,12,13 that is simple, low cost, and highly precise with good linear performance. This method was inspired by water striders walking on the water surface. Water strider legs with superhydrophobic properties bend the water surface. The curved surface of the water refracts sunlight, creating a shadow with a bright edge at the bottom of the pond. The shadow size is generally much larger than the indentation depth of legs from the water surface14,15,16. In the system, the applied displacement and the change in the diameter of the shadow were proportional, which was verified by the calibration experiment. The research indicates that this method provides an alternative for measuring micro displacement precisely.
1. Preparation of the PDMS piece
2. Experimental preparation for measuring micro-displacement
3. Image processing
4. Displacement calibration
5. Measuring micro displacement
Following the protocol, the sensitivity of the micro-displacement measurement system can be calibrated, and the micro displacement can be measured. The shadow method in the micro-displacement measurement is presented as follows. Figure 3 shows the travel path of parallel light through the PDMS deformed surface due to the applied displacement. The refraction of parallel light forms a shadow having a bright edge. The explicit solution of displacements z(x) of the PDMS surface based on the Johnson-Kendall-Roberts (JKR) model is18:
(Equation 1)
where Z is the applied displacement, and r is the radius of the rigid cylindrical leg. The incident angle θ1 of the outermost light can be expressed as Equation 2:
(Equation 2)
where T is the radius of the parallel light beam. The refraction angle θ3 of outgoing rays can be expressed as Equation 3 based on Snell's Law2 because tan θ1 tends to be zero:
(Equation 3)
where n is the refractive index of PDMS. Finally, the relationship between the applied displacement and the shadow diameter change is12,17:
(Equation 4)
where P is the thickness of the PDMS piece, W is the working distance between the PDMS surface and the camera, and k is the sensitivity to be calibrated. According to Equation 4, the applied displacement is proportional to the shadow diameter change. Hence, the displacement can be measured by the shadow diameter change after the calibration.
Figure 4 illustrates the diameter fitting process of the shadow images. The original image represented by Figure 4A was converted to the grayscale image and preprocessed by Gaussian filtering and image enhancement, where Figure 4B was obtained. Figure 4C illustrates the sub-pixel edge contour set, which was acquired by the Canny operator. Figure 4D is the least squares circle fit for the circular shadow. Finally, the shadow diameter could be obtained with sub-pixel accuracy.
Figure 5 shows typical shadow images in the process of calibration. The shadow diameter increases as the applied displacement increases. It is observed that the shadow diameter changes symmetrically as the applied displacement increases and decreases.
Figure 6 shows the curve fit of mean values of the changes in shadow diameters due to micro-displacements. The slope of the fitted curve, which denotes the displacement sensitivity, was 10.0 nm/pixel with the fit goodness of 0.9995. Hence, according to Equation 4, the resolution of the measurement system was 10.0 nm when the circle fitting accuracy was 1 pixel. In the measurement range of 5 µm, the maximum deviation was 100.6 nm with a linearity error of 2.01%, which is shown in Figure 7.
Figure 1: Schematic representation of the micro-displacement measurement system. Please click here to view a larger version of this figure.
Figure 2: Appearance of the micro-displacement measurement system. Inset: Photo of a shadow for display. Please click here to view a larger version of this figure.
Figure 3: Travel path of parallel light beam through the bent surface of the PDMS. Z is the applied displacement. r is the radius of the rigid cylindrical leg. T is the radius of the parallel light beam. n is the refractive index of PDMS. P is the thickness of the PDMS piece. W is the working distance between the PDMS surface and the camera. θ1 and θ2 are the incident angle and the refraction angle of the outermost light. θ3 is the refraction angle of the outgoing rays. D is the shadow diameter. Please click here to view a larger version of this figure.
Figure 4: Typical processing of the shadow image. (A) RGB shadow image. (B) Preprocessing image. (C) Sub-pixel edge. (D) Least-squares circle fitting. Please click here to view a larger version of this figure.
Figure 5: Typical shadow images in the process of calibration. The applied displacements are 0 µm, 1 µm, 2 µm, 3 µm, 4 µm, and 5 µm, respectively. Please click here to view a larger version of this figure.
Figure 6: Results of micro-displacement. The displacement sensitivity was 10.0 nm/pixel. The goodness of the linear fitting was 0.9995, determined by the least squares method. Please click here to view a larger version of this figure.
Figure 7: Error distribution curve of the displacement calibration. Please click here to view a larger version of this figure.
This protocol proposed a micro-displacement measurement system based on the shadow technique. The displacement calibration is the critical step within the protocol to obtain the displacement sensitivity and the measuring range. The displacement sensitivity can be improved by reducing the diameters of the cylindrical legs and that of the parallel light beam and increasing the working distance based on Equation 4. Furthermore, the pixel size and the resolution of the camera, as well as the accuracy of image processing, are also key to improving displacement sensitivity. The pixel size must be as small as possible to get the smallest change of shadow diameter. The camera resolution should be as big as possible to increase the measuring range. The fitting accuracy of shadow images directly affects the measurement system’s performance.
The system based on the shadow method is not suitable for large displacement measurements. The PDMS is a sensitive component that will be damaged due to large deformation. The camera resolution also limits the measuring range. The camera resolution must be high enough to contain all the shadow images. Moreover, the PDMS has a relatively short life of only 2 years at room temperature.
The micro-displacement measurement system has the advantages of simple construction, low cost, high precision, and good linear fitting. Compared with the existing methods5,19, the shadow method can only determine a high-precision displacement measurement with a parallel light source and a camera (such as a mobile phone).
This system shows an advancement in micro-displacement measurements in scientific and industrial fields. For example, it has a great potential in the movement displacement measurement of parts during ultra-precision machining. Furthermore, many mechanical measurements can be first converted into displacement tests through appropriate conversion20, such as the measurement of force, torque, speed, acceleration, temperature, flow, and other parameters.
The authors have nothing to disclose.
We thank the National Key Research and Development Program of China (No 2021YFC2202702) for funding this work.
Aperture diaphragm | Processed by high precision grinding | The diameter of the aperture is 0.7 mm. | |
Camera | Canon | EOS80D | The pixel size and the resolution of the camera are about 3.72 μm and 4000 × 6000, respectively. |
HALCON | MVTec Software GmbH | 18.11 | MVTec HALCON is the comprehensive standard software for machine vision with an integrated development environment (HDevelop) that is used worldwide. |
Motorized linear stage | Zolix | TSA50-C | Resolution 0.625 μm |
Parallel light source | Oriental Technology (Shanghai) Co, Ltd. | BTPL-50G | The peak wavelength is 523 nm. |
Polydimethylsiloxane (PDMS) | Dow Corning | Sylgard 184 | PDMS is a transparent silicon-based crosslinked polymer. |
Vacuum pump | SHANGHAI LICHEN-BX INSTRUMENT TECHONOLOGY CO.,Ltd | 2XZ-6B | The pumping rate is 6 L/s.The ultimate vacuum is ≤1 Pa |
Vertical precision positioner | PI | P-620.ZCD | The resolution is 0.2 nm in the range of 50 μm. |
Workbench with three rigid cylindrical legs | Processed by high precision grinding | The diameter of legs is 0.5 mm. The legs are distributed on the trisection points of a circle with a radius of 14 mm |
.