Magnetic tweezers, a powerful single-molecule manipulation technique, can be adapted for the direct measurements of the twist (using a configuration called freely-orbiting magnetic tweezers) and torque (using a configuration termed magnetic torque tweezers) in biological macromolecules. Guidelines for performing such measurements are given, including applications to the study of DNA and associated nucleo-protein filaments.
Single-molecule techniques make it possible to investigate the behavior of individual biological molecules in solution in real time. These techniques include so-called force spectroscopy approaches such as atomic force microscopy, optical tweezers, flow stretching, and magnetic tweezers. Amongst these approaches, magnetic tweezers have distinguished themselves by their ability to apply torque while maintaining a constant stretching force. Here, it is illustrated how such a “conventional” magnetic tweezers experimental configuration can, through a straightforward modification of its field configuration to minimize the magnitude of the transverse field, be adapted to measure the degree of twist in a biological molecule. The resulting configuration is termed the freely-orbiting magnetic tweezers. Additionally, it is shown how further modification of the field configuration can yield a transverse field with a magnitude intermediate between that of the “conventional” magnetic tweezers and the freely-orbiting magnetic tweezers, which makes it possible to directly measure the torque stored in a biological molecule. This configuration is termed the magnetic torque tweezers. The accompanying video explains in detail how the conversion of conventional magnetic tweezers into freely-orbiting magnetic tweezers and magnetic torque tweezers can be accomplished, and demonstrates the use of these techniques. These adaptations maintain all the strengths of conventional magnetic tweezers while greatly expanding the versatility of this powerful instrument.
In recent years, single-molecule techniques have proven their wide applicability in the study of processive motor proteins and other enzymes, yielding insight into their kinetics and the underlying mechanochemistry. In the context of force spectroscopy, important contributions have been made by atomic force microscopy flow stretching, and optical and magnetic tweezers. Optical and magnetic tweezers (MT) have notably succeeded in combining great flexibility in terms of molecular manipulation with high spatial and temporal resolution. Here, we focus on MT, which can apply both stretching forces and torques to biological molecules tethered between a surface and superparamagnetic beads1-3.
Magnetic tweezers (MT, Figure 1a) are a very versatile single-molecule technique that has been used to monitor both the mechanical properties of nucleic acids as well as their interactions with proteins. MT have many strengths, including overall simplicity and robustness of the experimental implementation, facile application of torque, natural operation and straightforward calibration in constant force mode4, extension to parallel measurements5,6, and absence of sample heating and photodamage. Compared to other single-molecule approaches, MT provide a way to perform force-dependence measurements at forces as low as ≈ 10 fN and have the ability to straightforwardly control the degree of supercoiling. While MTs have predominantly been used as an experimental tool to investigate biological processes involving nucleic acids7,8, they have also found application in studies of the mechanical properties of proteins9-13 or cells10,14-17. Numerous useful references are available that describe how to build and run a MT4,18-20.
However, conventional MT do not track rotational motion directly, and, while they apply torque, they do not measure torque directly. In addition, they constrain the free rotation of the nucleic acid tether. Here, we present two extensions of magnet tweezers. The first, termed freely-orbiting magnetic tweezers (FOMT, Figure 1b)21, allows the measurements of equilibrium angle fluctuations and changes in the twist of tethered nucleic acid molecules, without constraining the rotational motion around the tether axis. The second, termed magnetic torque tweezers (MTT, Figure 1c), which has the capability to apply and directly measure both forces and torques to single biomolecules22-27.
In the following protocol, we presume that the reader has at his/her disposition a ‘conventional’ MT instrument. We refer the reader to the Discussion for references on how to build and run a MT set up, as well as considerations that must be taken into account in the selection of magnetic beads, magnets, and tracking routines. In addition, sections 1 and 2 of the Protocol Text describe how we typically prepare and incubate a DNA sample for use in the MT as well as the preliminary measurements that can be performed on a single DNA in the conventional MT. Sections 3 and 4 of the Protocol Text illustrate how an MT instrument can be readily adapted and used for FOMT and MTT measurements.
1. Preparation and Incubation of a DNA Sample
2. Measurements on a Single DNA Molecule in the Conventional Magnetic Tweezers
3. Measurements of DNA Twist Using the Freely-orbiting Magnetic Tweezers
4. Measurements of DNA Torque Using the Magnetic Torque Tweezers
Representative results from the MT (Figure 1a) are shown in Figure 2. Figure 2a shows rotation-extension curves for a 7.9 kb DNA taken at F = 0.25, 0.5, and 2.0 pN. The response of a single DNA to rotation should be symmetric at the lowest forces (0.25 pN), with the extension of the DNA decreasing as a result of the formation of positive or negative plectonemic supercoils. Qualitative knowledge of this response is useful when initially searching for a rotationally constrained DNA tether (step 2.1). Note that additional inspection of the tether is required to verify that it consists of a single DNA molecule: here, the asymmetric response of a single DNA to rotation at forces exceeding 0.5 pN helps to distinguish it from multiple DNAs (step 2.1.1). Once this has been verified, one returns to the rotational response at 0.25 pN in order to determine the exact number of magnet turns at which the single DNA is torsionally relaxed, where one takes a force-extension curve, which should resemble Figure 2b. For this particular measurement, a fit of the data to the worm-like chain model (solid line) yielded a fitted contour length LC = 2.71 µm and bending persistence length LP = 45 nm. For dsDNA, the fitted values of the persistence length should lie in the range 40-55 nm, depending on the buffer conditions33, and the fitted contour length should be close (typically within 10%) to the value expected for the DNA construct that is used in the measurements, using the relationship LDNA = 0.34 nm/bp · number of base pairs.
Figure 3 shows the procedures and results of alignment in the FOMT (Figure 1b). The initial (x,y) excursions recorded in step 3.2 can be compared to the overall view of fluctuations as a function of the transverse magnet position shown in Figure 3a, which shows a “vortex” pattern that can be used to guide subsequent relative displacement between the magnet and DNA-tethered bead held in the FOMT. When subsequent coarse alignment is complete, the bead’s (x,y)-fluctuations trace out a circular trajectory, as is also shown by the black trace in Figure 3b. At this point, the torque from the magnets about the z-axis is reduced to the point that thermal fluctuations suffice to rotate the bead around its attachment point. The radius Rcircle of the resulting circular annulus (fitted circle is shown in red) represents the radial distance between the DNA attachment point and the bead’s center (Figure 1b). As shown in Figure 3c, however, a histogram of the data in Figure 3b shows that coarse alignment does not guarantee uniform coverage of all possible positions along the circular annulus. Even though thermal fluctuations are sufficient to explore all rotations angle on the circle, there remains a small energy barrier (of the order of the thermal energy kBT) to free rotation.
When finer alignment is carried out in the FOMT (step 3.4), the instrument can be used to determine the torsional modulus of DNA (Figure 4). First, fine alignment of the sample is used to obtain circular motion (Figure 4a) whose two-dimensional histogram should now show uniform coverage (Figure 4b). The corresponding time trace q(t) of angular fluctuations (obtained from conversion of the (x,y)-positions, see below) shows no periodicity corresponding to 360˚ (Figure 4c) and reveals large excursions corresponding to several full turns (Figure 4d). The implied energy landscape is harmonic over a range of > 1,000˚ (Figure 4e). The standard deviation of the fluctuations is σθ = 223°, corresponding to an angular trap stiffness of kθ=kBT/σθ 2= 0.27 pN·nm/rad, which in turn gives an estimate of the effective torsional persistence length of DNA equal to C = LC / σθ 2 ~ 76 nm (LC = 1,150 nm for the 3.4 kbp DNA used in this measurement) at the measured force.
An example of how FOMT can be used to measure the change in twist induced into the tethered DNA molecule through the binding of proteins31,34 is shown in Figure 5. Here, we have monitored the binding of RAD51 protein to double-stranded DNA; RAD51 is known to both lengthen and unwind DNA as it forms a nucleoprotein filament31. Upon flushing RAD51 into the flow-cell, we observe that the bead undergoes a spiraling trajectory in the FOMT (Figure 5a). By converting trace of (x,y) motion as a function of time to q(t) as described above, we can co-plot the effect that RAD51 has on the DNA tether length and its degree of unwinding (Figure 5b,c).
An alternative approach to measuring the torsional properties of DNA are the MTT (Figure 1c, Figure 6). The schematic in Figure 6a illustrates the principle of the measurement: after overwinding (or underwinding) the DNA tether by N turns, the DNA exerts a restoring torque on the bead that leads to a shift in the equilibrium angular position from θ0 到 θN. In the MTT the transverse component of magnetic field is reduced compared to the MT, which facilitates measurement of such angular shifts while still permitting bead rotation (Figure 1). The magnitude of the angular shift measured after applying N = 45 turns to a 7.9 kbp DNA is shown in Figure 6b. The complete sequence of the MTT measurement protocol and the resulting outcome of a torque versus rotation curve for DNA are shown in Figure 6c-f. Here, measurements of the standard deviation (Figure 6c) and the mean (Figure 6d) of the angular coordinate are shown as a function of over- and underwinding, with the standard deviation being inversely proportional to the angular trap stiffness (Equation 1). Taken together, these quantities allow one to construct a torque versus rotation curve for DNA (Figure 6f), which should show a linear response region centered about 0 turns and two plateaus at which the torque saturates, at positive and negative rotations, respectively. Such a torque versus rotation curve complements the information in an extension versus rotation curve (Figure 6e), thereby quantifying the transitions that accompany the buckling and denaturation of DNA.
Figure 1. Schematics of conventional magnetic tweezers (MT), freely-orbiting magnetic tweezers (FOMT), magnetic torque tweezers (MTT), and two strategies for tracking rotation angle. (a) In all three implementations of magnetic tweezers, magnetic beads are tethered to a flow cell surface by functionalized macromolecules, e.g., the double-stranded DNA molecules shown schematically. Reference beads are attached to the flow cell surface and tracked for drift correction. All three MT set-ups employ magnets to apply an upward stretching force on the magnetic bead and, therefore, DNA tether. In conventional MT, a pair of magnets exerts a magnetic field that is oriented transversely relative to the tether axis, tightly constraining rotation of the bead around the DNA-tether axis. In FOMT, a cylindrically-shaped magnet provides a magnetic field that oriented along the tether direction. When the tether is aligned to the center of the cylindrically-shaped magnet, any remaining transverse fields are minimized, allowing free rotation about the tether axis In MTT, a side magnet is added to the cylindrically-shaped magnet used in FOMT in order to provide a small transverse field (reduced in magnitude compared to MT). This small transverse field enables the application of torque as well as its measurement. (b) Two strategies to measure the rotation angle of a magnetic bead about the DNA-tether axis are shown. 1): a marker bead (green) attached to the magnetic bead (brown) gives an asymmetric image that enables angle tracking by imagine analysis. Two CCD images of a 1.4-μm-radius magnetic bead with a 0.5-μm-radius fiducial marker are shown, in focus and out-of-focus. 2): when the DNA is tethered to the magnetic bead at a position away from the bead’s south pole, the center of the bead fluctuates along an arc whose center defines an angular position. Either strategy can be used to track rotation angle and to monitor shifts in the angle position as the tether is torsionally strained (traces on the right), thus enabling measurements of single molecule torque.
Figure 2. DNA calibration measurements in the conventional MT. (a) Rotation-extension curves for a 7.9 kb DNA taken at F = 0.25, 0.5, and 2.0 pN. The asymmetric response under rotation to positive and negative turns of single double-stranded DNA tethers can be used as a convenient test of the tether attachment. (b) Force-extension curve for a 7.9 kb DNA, together with a fit to the worm-like chain model (solid line), yielding a fitted contour length LC = 2.71 µm and bending persistence length LP = 45 nm. All measurements were performed in PBS buffer.
Figure 3. Alignment in the FOMT. (a) (x,y) fluctuations of DNA-tethered bead held in the FOMT as a function of magnet position. The position of the cylindrical magnet was scanned at a constant height of 3 mm across the flow cell surface in steps of 250 μm in x and y and is indicated on the outer plot axes. At each (x,y)-position of the magnet, fluctuations of the same DNA-tethered bead were recorded and are plotted in the small coordinate systems (the scale bar on the bottom right applies to all sub-coordinate systems). Systematic variations of the bead’s (x,y)-fluctuation pattern with magnet position resembling a cyclone or vortex are apparent. This “vortex” pattern can be used to guide the displacement of the magnet (or alternatively the tether while keeping the magnet fixed) in x and y (indicated by the large arrows) to achieve alignment. When coarse alignment is complete, the bead’s (x,y)-fluctuations trace out a circular trajectory (blue trace in the center of the plot). This trace was recorded in a separate experiment after aligning the magnets in smaller steps about the center and is shown for illustration in this plot. (b) (x,y)-fluctuations of a DNA-tethered bead held in the FOMT after successful coarse-alignment of the magnet (black trace). The fluctuations lie on a circular annulus and thermal fluctuations are sufficient to explore all rotations angles on the circle. A fitted circle is shown in red. (c) A histogram corresponding to the data in (b), showing that coarse alignment does not guarantee uniform coverage of all possible positions along the circular annulus. Even though thermal fluctuations are sufficient to explore all rotations angle on the circle, there remains an energy barrier (on of the order of the thermal energy kBT) to free rotation.
Figure 4. Measurement of DNA torsional stiffness using FOMT. (x,y)-trajectory (a) and histogram (b) of a DNA-tethered bead’s fluctuations after fine alignment of the relative magnet-tether position in the FOMT. Under these circumstances, the histogram reveals essentially uniform coverage of the positions on the circle. (c) Rotational fluctuations of the bead determined from the (x,y)-positions. (d) Histogram of the rotational fluctuations. The red line is a Gaussian fit with σθ = 223°. (e) The energy landscape implied by the rotational fluctuation density from (c) and (d). The difference between the energy landscape implied by the rotational fluctuations and an harmonic approximation (with kθ = kBT / σθ2 = 0.27 pN-nm/rad) is much smaller than the thermal energy kBT over several turns. Data are offset for clarity such that θ0 = 0. The width of the fluctuations can be used to determine the torsional stiffness of DNA, see main text. The measurement was carried out in PBS buffer at a stretching force of ~1 pN. Data are adapted from Lipfert et al21.
Figure 5. The binding of RAD51 protein to DNA measured using FOMT. (a) Assembly of RAD51 protein onto a tethered 7.9 kbp dsDNA monitored at 3.5 pN. The (x,y,z)-trajectory executed by the magnetic bead (diameter 1.0 mm) during the first 200 sec of the assembly is shown, with time color-coded from blue to red. (b) The extension of the dsDNA deduced from the z-component of the bead trajectory in (a) as a function of time. (c) The rotational angle about the dsDNA tether axis deduced from the x,y components of the bead trajectory in (a) as a function of time.
Figure 6. Torque measurements on a single DNA tether in the MTT. (a) Schematic showing the principle of the torque measurement. After over- (or under-) winding the DNA tether by N turns, the DNA exerts a restoring torque on the bead that leads to a shift in the equilibrium angular position from θ0 到 θN. (b) Example of angle traces used to measure torque: angular fluctuations of a bead tethered to a torsionally relaxed 7.9 kbp DNA molecule before (blue) and after introducing 40 turns (dark red). (c-f) Torque measurement on a 7.9 kbp DNA molecule in PBS buffer held at a stretching force of ~3 pN using the fiducial marker bead based angular tracking protocol. Angular fluctuations as shown in (b) were recorded as a function of the number of applied turns. (c) The standard deviation of the angular fluctuations as a function of applied turns. The width of the fluctuations is approximately constant, indicating constant angular trap stiffness. (d) The shift in the mean rotation angle as a function of applied turns. Systematic shifts of the mean angle upon over- and underwinding are apparent. (e) The simultaneously monitored DNA tether extension as a function of applied turns. (f) The torque exerted by the DNA tether determined from the mean angle shown in (d), see main text. Over- and underwinding around zero turns gives rise to a linear torque vs. turns response of the DNA-tether (fitted grey slopes ion (d) and (f)) that can be used to determine the effective torsional persistence length (~77 nm for this data set). Further overwinding leads to buckling and formation of plectonemic supercoils (schematically shown in the insets), corresponding to a torque plateau (black line at positive turns in (f) at ~26 pN·nm) and a linear decrease of the tether extension with number of turns (black slope in (e)). Unwinding beyond the linear regime causes the DNA to locally melt (shown in the insets on the left), marked by a torque plateau equal to the melting torque (black line at negative turns in (f) at ~ -11 pN·nm).
When running experiments using the MTT or the FOMT, a number of choices need to be made regarding beads, magnets, tracking protocols, etc. The best choices to be made will depend on the experiment of interest. Below, we describe the trade-offs that accompany different choices, which should facilitate selection for a particular experiment. Next, we describe several critical steps that accompany the alignment and running of MTT and FOMT experiments. Lastly, we discuss the significance of the MTT and FOMT with respect to existing methods as well as future applications.
Considerations Prior to the Start of MTT and FOMT Experiments
Any experiment requires one to select a type of magnetic bead for use. One can select between several commercially-available streptavidin-coated superparamagnetic beads, e.g., 0.25 μm radius beads, 0.5 μm radius beads, or 1.4 μm radius beads (see the Materials table). Larger beads will have an increased magnetic moment compared to smaller beads (roughly scaling as the volume) and therefore their use will facilitate the application of higher forces (for typical forces achieved in our instruments, see Table 1). When angular tracking using marker beads is desired, we typically work with 1.4 μm radius and use 0.5 μm radius non-magnetic biotinylated beads as marker beads (see paragraph 1.9 for the corresponding attachment protocol). The use of smaller beads is particularly recommended for the FOMT, as the characteristic timescale for bead rotation τC equals the ratio of the system’s drag over its spring constant γ/kθ; importantly, the rotational drag coefficient relevant for the angular measurement time scale scales as ~ Rbead3, i.e. with the third power of the radius (see Table 2 for the characteristic time scales for several bead-DNA combinations in FOMT and MTT measurements). Accompanying reductions in the maximum force that can be applied can be addressed by using a flipped stack of cylindrical magnets27. Nonetheless, in FOMT measurements it may sometimes be necessary to compromise between the best achievable temporal resolution and the maximum applied force.
Additionally, an experiment requires the selection of a magnet configuration. In the conventional magnetic tweezers configuration (Figure 1a), we typically use a pair of 5x5x5 mm cubic magnets in vertical orientation with a 0.5 or 1 mm gap between the magnets4. When the magnets are spaced along the x (y) axis, this yields a magnetic field that is primarily directed along the x (y) axis. For FOMT experiments, a cylindrically-shaped magnet is selected at whose center the magnetic field is primarily directed along the z axis (Figure 1b). In practice, we use a stack of three such cylindrically-shaped magnets, each with a diameter of 6 mm and a 2 mm diameter central hole, for a total thickness of 6 mm. When higher pulling forces are desired, a “flipped stack” magnet configuration in which the bottom magnet is stacked with opposite magnetization is preferred. To achieve the MTT configuration (Figure 1c), we add an additional magnet to the side of the main magnet stack of the FOMT configuration, typically a solid cylinder with 4 mm diameter and a height of 7 mm. To see how the maximal forces achieved in our instruments depend on the magnet configuration, see Table 1.
The Alignment of MTT and FOMT Experiments
Since magnetic beads have an (approximately) uniformly functionalized surface (typically streptavidin) and since the attachment of both the functionalized nucleic acid tethers and marker beads (in case the marker bead-based angular tracking is employed) occurs via simple incubation in solution, one does not control where the tether and/or marker bead attach to the magnetic bead. The magnetic beads have a preferred magnetization axis that tends to align along the direction of the external field. If we denote the points where the preferred magnetization axis intersects the bead’s surface as the north and south poles, then beads where the DNA-tether is attached close to the equator will trace out a circular annulus with a radius close to or slightly larger than the bead radius in the FOMT; in contrast, beads that are attached close to the south pole will fluctuate on a circular annulus with very small radius in the FOMT, which can preclude fitting of the circle using Equations 3-5. We note that by simple spherical geometry, the probability of attaching near the equator is much larger than an attachment exactly at the poles; therefore, most beads will be tethered such that the (x,y)-based angular tracking can be carried out successfully.
A similar argument holds for the attachment of the marker beads for the fiducial marker based angular tracking. The marker bead is used to create an asymmetry in the image of the magnetic bead that enables angle tracking. If the marker bead is attached exactly at the north or south pole of the bead (i.e. directly on top or on the bottom), the resulting image is still rotationally symmetric and the angular tracking protocol fails. However, by the same spherical geometry argument, the chance for a marker bead to attach directly at one of the poles is relatively small; we find that in practice most marker beads give a sufficient asymmetry to enable angular tracking. Finally, we note that in the conventional magnetic tweezers the field direction is in the (x,y)-plane; therefore, the preferred magnetization axis of the bead will align in the (x,y)-plane and the north and south poles, as defined above, are going to be at the sides of the bead, unlikely the situation in the FOMT or MTT, where the poles are at the top and bottom.
In FOMT experiments, a critical step is the alignment of the cylindrical magnet such that the radial magnetic field is negligible in proximity to the bead. This alignment is performed for a single bead at a time. To judge whether bead motion in the FOMT is evenly distributed over a circular annulus, the measurement time should exceed 20·τC. As τC equals ~45 sec for 8 kbp DNA and a 0.5 mm radius bead, the measurement time is ~900 sec in the final stages of alignment. For comparison, use of 1.9 kbp DNA and 0.25 mm radius beads reduces τC twenty-fold to ~2 sec (see also Table 2).
Critical Steps and Considerations for Tracking During FOMT and MTT Experiments
To track the bead’s in-plane fluctuations, i.e. its (x,y)-position, we employ a cross-correlation analysis of the intensity profiles displayed by a bead at subsequent time intervals35,36. This can be carried out at sub-pixel resolution to an accuracy of a few nanometers20. To track the bead’s motion in z, we typically use a method first designed by Gosse and Croquette, in which the objective’s focal plane (OFP) is accurately shifted in the vertical direction while imaging the diffraction rings of the bead attached to the nucleic acid20. In this manner, a calibration profile is generated correlating the diffraction pattern of the bead to the distance between the bead and the OFP19. When this calibration profile is interpolated, the vertical displacements of the bead can be also measured with an accuracy of up to a few nm20. We refer the reader to additional references that describe more refined tracking algorithms37,38 as well as their application to parallel tracking of multiple beads5,6,37.
When using angular tracking that relies on the conversion of (x,y)-positions into angular coordinates, we advise to proceed as follows. From a time trace in which the bead traces out a circular annulus, use the (xi ,yi) positions (where the index i denotes subsequent measurement points) to fit the circle center (x0, y0) and radius Rcircle (Figure 2a) by minimizing:
(3)
where the sum runs over all data points. After fitting x0, y0, and Rcircle, determine the polar coordinates (ri ,θi) of each data point in the time trace using:
(4)
(5)
Note that one should take care to “unwrap” the angle θ, i.e. to add phase jumps of ±π where appropriate. Custom-written code for the fitting and conversion from (x,y) to (r, θ) coordinates is available from the authors upon request. In the FOMT, a time trace in which the bead traces out a circular annulus can be obtained by achieving coarse alignment (cf. step 3.3) and recording the thermal fluctuations of the bead. In the MTT, thermal fluctuations are insufficient to trace out the circular annulus; instead, use a time trace where the magnets are slowly (typically at 0.1 Hz) rotated by several turns to fit the circle using Equations 3-5.
We note that for the MTT, it is important to choose the proper angular tracking approach, i.e. via an angular tracking marker (Figure 1c, Figure 1d, Figure 3a) or via the conversion of (x,y)-positions into angular coordinates (Figure 1d, Figure 2b). While typically the accuracies of the angular tracking from (x,y)-positions and the use of marker beads are comparable, it is important to realize that crosstalk occurs between a bead’s fluctuations in (x,y) and in angle, as described in Janssen et al32: thus, angular tracking from (x,y)-positions is only valid provided that the Brownian fluctuations in (x,y) contribute only negligibly to the uncertainty in the angular coordinate, and its proper use of (x,y)-tracking may require tuning of the rotational trap stiffness via adjustment of the position of the side magnet. Typically, the use of higher trap stiffness requires the use of angular tracking using marker beads. The use of marker beads requires an additional attachment step, which may reduce the number of usable tethers (see the attachment protocol in step 1.9). When using the marker bead-based tracking, it is important to select magnetic beads which have a marker bead is attached near the equator for best results.
Significance of the FOMT and MTT Approaches Compared to Existing Methods and Applications
In the above, we have shown how one can, starting from conventional MT, easily modify the magnet configurations to convert the instrument into MTT or FOMT. This straightforward modification, which may be accompanied by the introduction of angular tracking when the use of an angular tracking marker is desired, is an immediate strong point of both configurations, as it permits the user to apply torque, measure torque, or measure twist depending on the experiment at hand. As mentioned in the Introduction, both FOMT and MTT benefit from many of the existing strengths of MT, notably their simplicity, with the MTT in particular also benefiting from the capability of parallel measurements5,6 (these are not as easily achieved in FOMT given the requirement of alignment of the tether with respect to the center of the cylindrical magnet). Notably, MTT and FOMT do not require, in contrast with other techniques, specially nano-fabricated particles22,39,40, complex optical design41, or the introduction of additional beads within the tethered (DNA) molecule42. Such other techniques may nonetheless provide other advantages such as higher temporal resolution27,43,44. Both FOMT and MTT should find future applications in the study of genome processing, as the behavior of molecular motors on DNA is both influenced by and has consequences for the local twist and torque. Additional applications can be found in the emerging field of DNA nanotechnology27 or in the wider field of rotary motors active in biological processing7,45.
M270 (Rbead = 1.4 µm) | MyOne (Rbead = 0.5 µm) | Ademtech (Rbead = 0.25 µm) | |
Conventional MT (pair of cubic 5 x 5 x 5 mm3 magnets, 1 mm gap, vertical alignment) | 70 pN | 8 pN | 1.6 pN |
FOMT or MTT* (stack of three cylindrical magnets, 6 mm diameter, 2 mm diameter gap) | 9 pN | 1 pN | 0.2 pN |
FOMT or MTT* (stack of three cylindrical magnets, 6 mm diameter, 1 mm diameter gap) | 18 pN | 2 pN | 0.4 pN |
FOMT or MTT* (stack of three cylindrical magnets with last one flipped, 1 mm diameter gap) | ~50 pN | 9 pN | 1.8 pN |
*The presence of the small side magnet in the MTT has a negligible effect on the stretching force
Table 1. Maximum forces typically achieved for different magnet configurations and bead types.
Rbead = 1.4 µm | Rbead = 0.5 µm | Rbead = 0.25 µm | |
Friction coefficient* | 120 pN·nm·sec | 5.5 pN·nm·sec | 0.7 pN·nm·sec |
Characteristic time scale: FOMT, 10 kbp DNA** | 1200 sec | 55 sec | 7 sec |
Characteristic time scale: FOMT, 1 kbp DNA | 120 sec | 5.5 sec | 0.7 sec |
Characteristic time scale: MTT, kq = 100 pN·nm/rad | 1.2 sec | 0.06 sec | 0.007 sec |
Characteristic time scale: MTT, kq = 1000 pN·nm/rad | 0.12 sec | 0.006 sec = 6 msec | 0.0007 s = 0.7 msec |
*Friction coefficient for rotation about an axis through the “equator” (i.e. the situation shown in Figure 1b), given by 14·p·h·Rbead3, where h is the viscosity of the buffer.
**In the FOMT, the rotational trap stiffness is given by the torsional stiffness of the DNA, kq,DNA = C·kBT/LC, where C is the effective torsional persistence length, assumed to be 80 nm here (which is characteristic of an intermediate force regime, F ~ 1 pN) and LC is the contour length of DNA, 0.34 nm per base pair.
Table 2. Friction coefficients and characteristic time scales for FOMT and MTT.
The authors have nothing to disclose.
This work was supported by the TU Delft, the Netherlands Organisation for Scientific Research (NWO), the Foundation for Fundamental Research on Matter, and by the European Science Foundation.
Sandblaster | Great Lake Orthodontics | 190-070 Microetcher II | |
Nitrocellulose | Life Technologies | LC2001 | |
Magnetic particle concentrator | Life Technologies | 12002D | |
Non-magnetic latex beads (0.5 μm radius) | Polysciences | 17010 | |
Non-magnetic latex beads (1.5 μm radius) | Sanbio | PV05N/2179 | |
Antidigoxigenin | Roche | 11 214 667 001 | |
Streptavidin-coated superparamagnetic beads (0.25 μm radius) | Ademtech | 3150 | |
Streptavidin-coated superparamagnetic beads (0.5 μm radius, “MyOne”) | Life Technologies | 650.01 | |
Streptavidin-coated superparamagnetic beads (1.4 μm radius, “M270”) | Life Technologies | 653.05 | |
Biotin-coated latex beads (0.5 μm radius) | Life Technologies | F-8768 | |
Cubic magnets for conventional tweezers | Supermagnete | W-05-N50-G | |
Cylindrical magnet for MTT and FOMT | Supermagnete | R-06-02-02G | |
Side magnet for MTT | Supermagnete | S-04-07-N | |
Linear stage | Physik Instrumente | M-126.PD | |
Rotary stage | Physik Instrumente | C-150 | |
High-resolution automated sample stage | Physik Instrumente | P-733.2D | |
Software for coding analysis routines | The Mathworks | Matlab | custom-written routines are available from the authors |